776259
13
Verklein
Vergroot
Pagina terug
1/62
Pagina verder
GMC-I Messtechnik GmbH 13
Measurement of Touch Voltage via Finger Contact
When a measurement is started and if you touch the ON/START key
with your finger, the test instrument detects whether or not dan-
gerous touch voltage Ub is present at the PE terminal relative to
ground.
Error in the U Switch Position:
PE appears and the LIMIT LED lights up red.
Error in All Switch Positions Other than U:
The test instrument disables the measurement and the following
message appears: U.PE > UL!
Prerequisites for reliable finger contact measurement:
1 Nothing is plugged into the interfaces and the charging cable
is not plugged in.
2 Based on his standing surface, the user has an earth resis-
tance of R.eb < 1 MΩ.
3 While starting the measurement, the user touches the “ON/
START” key with the full surface of an unprotected finger with
direct skin contact.
Insufficient Supply Voltage
If (rechargeable) battery voltage falls below the allowable limit value
the instrument cannot be switched on, or it is immediately
switched off.
Conditions Resulting in Disabling and Abortion of Measurements
The measurement is interrupted automatically or the measuring
sequence is blocked (except for voltage measuring ranges and
phase sequence testing) in the event of:
Impermissible line voltages (< 60 V, > 253 V / > 330 V /
> 440 V or > 550 V) for measurements which require line volt-
age
Interference voltage during insulation resistance or low resis-
tance measurements
Overheating at the instrument.
As a rule, excessive temperatures only occur after approxi-
mately 50 measurement sequences at intervals of 5 seconds,
when the rotary selector switch is set to the ZL-PE or ZL-N
position.
If an attempt is made to start a measuring sequence, an
appropriate message appears at the display panel.
Automatic Instrument Shutdown
The instrument only switches itself off automatically after comple-
tion of an automatic measuring sequence, and after the predeter-
mined on-time has expired (see section 4.2). On-time is reset to
its original value as defined in the setup menu as soon as any key
or the rotary selector switch is activated.
The instrument remains on for approximately 75 seconds in addi-
tion to the preset on-time for measurements with rising residual
current in systems with selective RCDs.
The instrument always shuts itself off automatically, unless the fol-
lowing setting has been selected in SETUP: “>>>>>” (continuous
on).
5.3 Measurement Value Display and Memory
The following items appear at the display panel:
Measured values with abbreviations and units of measure
Selected function
Nominal voltage
Nominal frequency
Error messages
Measurement values for automatic measuring sequences are
stored and displayed as digital values until the next measurement
sequence is started, or until automatic shut-off occurs.
If the upper range limit is exceeded, the upper limit value is dis-
played and is preceded by the “>” symbol (greater than), which
indicates measurement value overrun.
Note
The depiction of LEDs in these operating instructions
may vary from the LEDs on the actual instrument due to
product improvements.
5.4 Testing Earthing Contact Sockets for Correct Connection
The testing of earthing contact sockets for correct connection
prior to protective measures testing is simplified by means of the
instrument’s error detection system.
The instrument indicates improper connection as follows:
Non-allowable line voltage (< 60 V or > 253 V):
The MAINS/NETZ LED blinks red and the measuring sequence is
disabled.
Protective conductor not connected or potential to earth 50 V at
50 Hz (switch position U – single-phase measurement):
If the contact surface of the START key is touched (finger con-
tact) while PE is being contacted (via the country-specific
measuring adapter, e.g. Z503K PRO-Schuko measuring
adapter as well as via the test probe in the case of 2-pole
measurement with the Z503L KS-PROFITEST INTRO), PE
appears (only after starting e test sequence). The MAINS LED
blinks red as well.
Neutral conductor N not connected (during mains dependent
measurements):
The MAINS/NETZ LED blinks green
One of the two protective contacts is not connected:
This is checked automatically during testing for touch current
UIΔN. Poor contact resistance at one of the contacts leads to
one of the following displays depending upon poling of the
plug:
Display in the connection pictograph:
PE interrupted (x), or bottom protective
conductor tab interrupted with reference
to the keys at the test plug
Cause:
voltage measuring path interrupted
Consequence: measurement is disabled
Display in the connection pictograph:
Top protective conductor tab interrupted
with reference to the keys at the test plug
Cause: current measuring path interrupted
Result: no measured value display
Note
See also “LED Indications, Mains Connections and
Potential Differences” beginning on page 43.
Attention!
!
Reversal of N and PE in a system without RCCBs cannot
be detected and is not indicated by the instrument.
In a system including an RCCB, the RCCB is tripped
during “touch voltage measurement without RCCB trip-
ping” (automatic ZL-N measurement), insofar as N and
PE are reversed.
13

Hulp nodig? Stel uw vraag in het forum

Spelregels

Misbruik melden

Gebruikershandleiding.com neemt misbruik van zijn services uitermate serieus. U kunt hieronder aangeven waarom deze vraag ongepast is. Wij controleren de vraag en zonodig wordt deze verwijderd.

Product:

Bijvoorbeeld antisemitische inhoud, racistische inhoud, of materiaal dat gewelddadige fysieke handelingen tot gevolg kan hebben.

Bijvoorbeeld een creditcardnummer, een persoonlijk identificatienummer, of een geheim adres. E-mailadressen en volledige namen worden niet als privégegevens beschouwd.

Spelregels forum

Om tot zinvolle vragen te komen hanteren wij de volgende spelregels:

Belangrijk! Als er een antwoord wordt gegeven op uw vraag, dan is het voor de gever van het antwoord nuttig om te weten als u er wel (of niet) mee geholpen bent! Wij vragen u dus ook te reageren op een antwoord.

Belangrijk! Antwoorden worden ook per e-mail naar abonnees gestuurd. Laat uw emailadres achter op deze site, zodat u op de hoogte blijft. U krijgt dan ook andere vragen en antwoorden te zien.

Abonneren

Abonneer u voor het ontvangen van emails voor uw Gossen Metrawatt Profitest Intro bij:


U ontvangt een email met instructies om u voor één of beide opties in te schrijven.


Ontvang uw handleiding per email

Vul uw emailadres in en ontvang de handleiding van Gossen Metrawatt Profitest Intro in de taal/talen: Engels als bijlage per email.

De handleiding is 1.69 mb groot.

 

U ontvangt de handleiding per email binnen enkele minuten. Als u geen email heeft ontvangen, dan heeft u waarschijnlijk een verkeerd emailadres ingevuld of is uw mailbox te vol. Daarnaast kan het zijn dat uw internetprovider een maximum heeft aan de grootte per email. Omdat hier een handleiding wordt meegestuurd, kan het voorkomen dat de email groter is dan toegestaan bij uw provider.

Uw handleiding is per email verstuurd. Controleer uw email

Als u niet binnen een kwartier uw email met handleiding ontvangen heeft, kan het zijn dat u een verkeerd emailadres heeft ingevuld of dat uw emailprovider een maximum grootte per email heeft ingesteld die kleiner is dan de grootte van de handleiding.

Er is een email naar u verstuurd om uw inschrijving definitief te maken.

Controleer uw email en volg de aanwijzingen op om uw inschrijving definitief te maken

U heeft geen emailadres opgegeven

Als u de handleiding per email wilt ontvangen, vul dan een geldig emailadres in.

Uw vraag is op deze pagina toegevoegd

Wilt u een email ontvangen bij een antwoord en/of nieuwe vragen? Vul dan hier uw emailadres in.



Info