GMC-I Messtechnik GmbH 5
safety measures or any consequential damage is
granted by the manufacturer.
Meanings of Symbols on the Instrument
Warning concerning a point of danger
(Attention: observe documentation!)
Indicates EC conformity
This device may not be disposed of
with the trash. Further information
regarding the WEEE mark can be
accessed on the Internet at
www.gossenmetrawatt.com by
entering the search term ’WEEE’.
3 Connecting the Test Adapter to the Mains
The following connections must first be
established before connecting the test adapter to
the mains:
a) Insert the mains plug from the
SECUTEST.../
SECULIFE ST
into the earthing contact outlet (12)
at the AT3-III E.
b) Insert the earthing contact plug (13) from the
AT3-III E into the test socket at the
SECUTEST.../SECULIFE ST.
c) In order to measure protective conductor
current, connect the SECUTEST port (14) of
the AT3-III E into sockets 2 and 3 at the
SECUTEST.../SECULIFE ST or into sockets V
and COM at the SECUTEST PRO.
d) In order to test extension cables, insert the test
probe at the end of the SECUTEST... probe cable
into the socket (15) at the AT3-III E.
Connect the test adapter to 230/400 V mains
power. The test adapter now performs a self-test.
If the tester is intact, the red LED lights up briefly.
4 Connecting a Device Under Test
to the Test Adapter
After the device under test has passed a visual
inspection and before connecting it to the
appropriate plug or socket at the test adapter, as
well as before each new test, the SECUTEST.../
SECULIFE ST test instrument must be returned to
its initial setting depending upon the type of test.
Connect the device under test to the test adapter
and switch all of its functions on, making sure
that, for example, thermostat contacts are closed
etc. Always measure protective conductor
resistance first for safety class I devices, because
if the protective conductor is defective, insulation
resistance, equivalent leakage current and
protective conductor current cannot be
measured, and high-voltage testing cannot be
performed.
5 Testing Devices
Perform testing in accordance with the operating
instructions included with the SECUTEST.../SECULIFE ST
test instrument!
Observe the following instructions when using the
SECUTEST.../SECULIFE ST in combination with the AT3-
III E during automatic test sequence :
5.1 Measuring Protective Conductor Resistance
The clip or the test probe at the end of the probe
cable from the SECUTEST.../SECULIFE ST test
instrument must be connected to the housing of
the device under test such that good contact is
assured.
Note!
Connector cable (13) resistance is 0.07 Ω. In
order to compensate for this error, proceed
as described in the operating instructions
included with the
SECUTEST.../SECULIFE ST test instrument
under Individual Measurements Protective
Conductor Resistance Zero Balancing.
If the AT3-III E is out of use for a prolonged period
of time, the earthing contact faces of the plug-
and-socket connectors and the connecting plug
(13) may corrode, thus leading to slightly elevated
resistance measurement values. In this case, plug
and unplug the connections several times until the
expected values are reached again.
5.2 Measuring Insulation Resistance and Equivalent
Leakage Current
L1, L2, L3 and N (short-circuited) are measured
against PE during insulation testing.
5.3 Measuring Protective Conductor Resistance with
the Residual Current Method (except for
SECUTEST BASE(10))
The DUT is placed into operation during residual
current measurement. The L1, L2 and L3 mains
lamps light up at the AT3-III E during this test.
This measurement may not be performed on safety
class I devices until after the protective conductor test
has been passed in accordance with chapter 5.1.
ÐTurn off the device under test.
ÐSelect the IL-DI measurement setting at the
SECUTEST.../SECULIFE ST.
ÐThe line contactor switches mains power to
the test outlets (1 – 4) at the AT3-III E. Signal
lamps L1, L2 and L3 indicate the presence of
line voltage.
ÐNow switch on the DUT
(as described in chapter 2).
Contact current is measured by means of residual
current measurement for safety class II devices,
!
5

Hulp nodig? Stel uw vraag in het forum

Spelregels

Misbruik melden

Gebruikershandleiding.com neemt misbruik van zijn services uitermate serieus. U kunt hieronder aangeven waarom deze vraag ongepast is. Wij controleren de vraag en zonodig wordt deze verwijderd.

Product:

Bijvoorbeeld antisemitische inhoud, racistische inhoud, of materiaal dat gewelddadige fysieke handelingen tot gevolg kan hebben.

Bijvoorbeeld een creditcardnummer, een persoonlijk identificatienummer, of een geheim adres. E-mailadressen en volledige namen worden niet als privégegevens beschouwd.

Spelregels forum

Om tot zinvolle vragen te komen hanteren wij de volgende spelregels:

Belangrijk! Als er een antwoord wordt gegeven op uw vraag, dan is het voor de gever van het antwoord nuttig om te weten als u er wel (of niet) mee geholpen bent! Wij vragen u dus ook te reageren op een antwoord.

Belangrijk! Antwoorden worden ook per e-mail naar abonnees gestuurd. Laat uw emailadres achter op deze site, zodat u op de hoogte blijft. U krijgt dan ook andere vragen en antwoorden te zien.

Abonneren

Abonneer u voor het ontvangen van emails voor uw Gossen Metrawatt AT3-III E bij:


U ontvangt een email met instructies om u voor één of beide opties in te schrijven.


Ontvang uw handleiding per email

Vul uw emailadres in en ontvang de handleiding van Gossen Metrawatt AT3-III E in de taal/talen: Engels als bijlage per email.

De handleiding is 0.59 mb groot.

 

U ontvangt de handleiding per email binnen enkele minuten. Als u geen email heeft ontvangen, dan heeft u waarschijnlijk een verkeerd emailadres ingevuld of is uw mailbox te vol. Daarnaast kan het zijn dat uw internetprovider een maximum heeft aan de grootte per email. Omdat hier een handleiding wordt meegestuurd, kan het voorkomen dat de email groter is dan toegestaan bij uw provider.

Stel vragen via chat aan uw handleiding

Stel uw vraag over deze PDF

loading

Uw handleiding is per email verstuurd. Controleer uw email

Als u niet binnen een kwartier uw email met handleiding ontvangen heeft, kan het zijn dat u een verkeerd emailadres heeft ingevuld of dat uw emailprovider een maximum grootte per email heeft ingesteld die kleiner is dan de grootte van de handleiding.

Er is een email naar u verstuurd om uw inschrijving definitief te maken.

Controleer uw email en volg de aanwijzingen op om uw inschrijving definitief te maken

U heeft geen emailadres opgegeven

Als u de handleiding per email wilt ontvangen, vul dan een geldig emailadres in.

Uw vraag is op deze pagina toegevoegd

Wilt u een email ontvangen bij een antwoord en/of nieuwe vragen? Vul dan hier uw emailadres in.



Info