How to Use
26
How to Use
Press and hold the key for
about 2 seconds.
[CAL] is displayed.
2
Calibration
Calibrate the unit when it has been dropped or otherwise subject to a
strong impact.
Accurate measurements are not possible if the distance between the opposing distance sensors on the main
unit has changed due to a strong impact. This can be addressed by calibrating the unit, thus allowing
accurate measurements to be taken again.
To Calibrate Correctly
• Calibrate on a flat, stable surface.
• Ensure no objects are between the distance sensors.
• Calibrate at a temperature of 15-25°C.
• Check that remove the AC adapter from the main unit.
Press the key on the main unit to turn
on.
The initial screen is displayed.
1
*Blue print in display illustrations indicates parts that illuminate or flash.
No objects
here!
For Your Reference
Please use the tip of a pen or other fine
instrument to press the [over 130cm] (taking care not
to damage the unit).
Do not insert AC adapter!