783666
1
Verklein
Vergroot
Pagina terug
1/2
Pagina verder
Cleaning
Maintenance
Troubleshooting
Specification
1. Please clean the surface of the device before using.
Wipe the device with medical alcohol (70% isopro-
pyl alcohol)first, and then let it dry in air or clean it
by dry clean fabric. When cleaning the device with
water, the water temperature should be lower than
60°C
2. Using the medical alcohol to disinfect the product
after use, prevent from cross infection for next time
use.
3. The best storage environment of the device is - 25°C
to 70°C ambient temperature and not higher than
90% relative humidity.
Note: 1. Do not sterilize, autoclave or immerse this de-
vice in liquid. Do not pour or spray any liquids
onto the device.
2. Do not use caustic or abrasive cleaning
agents, or any cleaning agent containing am-
monium chloride or isopropyl alcohol.
Recommends user to return this device to the man-
ufacturer perform the following checks every 24
months.
- Inspect the equipment for mechanical and function-
al damage or deterioration.
- Ensure all user interface keys and accessories func-
tion normally.
Note: Manufacturer use simulator of model Index 2
to verify operation of the pulse oximeter equip-
ment.
Symptoms Check points Corrections
SpO2 or
pulse rate
cannot
displayed
Applied nger
improperly.
Place the nger prop-
erly and try again.
SpO2 is too low to
detect
Try again; go to con-
sult with your physi-
cian if you are sure the
device works well.
SpO2 or
pulse rate are
not displayed
stably
Applied nger
improperly.
Place the nger prop-
erly and try again.
Finger is shaking
or body is moving. keep body steady
No display
when button
is pressed
Batteries run
down
Replace with new
batteries
Batteries not
inserted correctly. Re-insert batteries
The display
disappears
suddenly
The device will
auto power o
when it gets no
signal.
Normal
Low battery Replace with new
batteries
Note: If the unit does not work, return it to your
dealer. Under no circumstance should you
disassemble and repair the unit by yourself.
SpO2
Measuring
range
35%~99%, (the resolution is 1%).
Accuracy 70%~99%: ±2%, Below 35~69%:
unspecied.
Optical Sensor The wavelength of red LED is 660 nm
and Infrared LED is 905/880 nm with
maximum optical output power of 4
mW/sr.
Pulse
Measuring
range
30 bpm~250 bpm (the resolution is 1
bpm)
Accuracy ±3 bpm
Power source AAA × 2 (Alkaline)
Battery life Continually for 16 hours with two alkaline
batteries
Operating
Condition
Temperature: 5°C~40°C (41°F ~ 104°F),
Relative Humidity: 15-95% (non
condensing), Atmospheric pressure:
700hPa ~ 1060hPa, Attitude: -1,280 to
12,000 feet (-390m to 3,658m)
Storage /
Transportation
Condition
Temperature: -25°C~+70°C(-13°F ~
158°F), Relative humidity: 15-90%(non
condensing), Atmospheric pressure:
700hPa ~ 1060hPa, Attitude: -1,280 to
12,000 feet (-390m to 3,658m), The time
from 70°C or -25°C back to use : 3 hours
Dimensions 63.5(L) × 34W) × 35(H) mm
Weight About 37g (without the batteries)
Standards IEC60601-1-2, Class B, IEC60601-1, Type BF,
ISO80601-2-61, IEC/EN60601-1-11
Type BF applied parts
IP Classication IP22: Protection against harmful ingress of
water and particulate matter
Note:
- A description of the effect on displayed and transmit-
ted SpO2 and pulse rate:
- Data averaging: 4 seconds for SpO2; 8 seconds for
pulse rate.
- Data update delay: Less than 2 seconds.
be evaluated at the measurement site. Rather, they
are exclusively used to display the current signal vari-
ation at the measurement site and do not enable re-
liable diagnostics for the pulse.
- A warning that other cables and accessories may
negatively aect EMC performance.
- A warning regarding stacking and location close to
other equipment.
- A warning that use of other accessories results in
non-compliance.
- The maximum temperature of sensors which the
user will touch might reach 42°C when operating in
the 40°C environment.
- The safety way for all people use is measuring for 10
minutes, and turn it o for 20 minutes before meas-
ure again.
- The oximeter is calibrated in the factory before sale,
there is no need to calibrate it during its life cycle.
Model: SB100
www.rossmax.com
Fingertip Pulse Oximeter
Rossmax InnoTek Corp.
12F., No. 189, Kang Chien Rd., Taipei,
114, Taiwan.
CMC Medical Devices & Drugs S.L.
C/ Horacio Lengo Nº 18, CP 29006,
Málaga, Spain
WARNING: The symbol on this product means that it's an electronic product
and following the European directive 2012/19/EU the electronic products have
to be dispose on your local recycling centre for safe treatment.
EMC guidance and manufacturer’s declaration
Recommended separation distances between portable and mobile RF communications equipment and the ME equipment
The Finger-tip pulse oximeter is intended for use in an electromagnetic environment in which radiated RF disturbances are con-
trolled. The customer or the user of the Finger-tip pulse oximeter can help prevent electromagnetic interference by maintaining a
minimum distance between portable and mobile RF communications equipment (transmitters) and the Finger-tip pulse oximeter
as recommended below, according to the maximum output power of the communications equipment.
Rated maximum output
power of transmitter / W
Separation distance according to frequency of transmitter / m
150 kHz to 80 MHz , d=[3.5/
V1]√P 80 MHz to 800 MHz , d=[3.5/
E1]√P 800 MHz to 2,5 GHz , d=[7/
E1]√P
0.01 0.1 0.1 0.2
0.1 0.4 0.4 0.7
11.2 1.2 2.3
10 3.7 3.7 7.4
100 11.7 11.7 23.3
Declaration – electromagnetic emissions
The Finger-tip pulse oximeter is intended for use in the electromagnetic environment specied below. The customer or the user of
the Finger-tip pulse oximeter should assure that it is used in such an environment.
Emissions test Compliance Electromagnetic environment - guidance
RF emissions CISPR 11 Group 1 Portable and mobile RF communications equipment should be used no
closer to any part of the EQUIPMENT or SYSTEM including cables, than the
recommended separation distance calculated from the equation applicable
to the frequency of the transmitter. Interference may occur in the vicinity of
equipment marked with the following symbol.
RF emissions CISPR 11 Class B The Finger-tip pulse oximeter is suitable for use in all establishments,
including domestic establishments and those directly connected to the
public low-voltage power supply network that supplies buildings used
for domestic purposes.
Harmonic emissions IEC 61000-3-2 N/A
Voltage uctuations/Flicker emissions
IEC 61000-3-3 N/A
Declaration – electromagnetic emissions and immunity – for EQUIPMENT and SYSTEMS that are use in the professional health-
care facility environment or in the home healthcare environment
The Finger-tip pulse oximeter declaration – electromagnetic immunity
The Finger-tip pulse oximeter system is intended for use in the electromagnetic environment specied below. The customer or
the user of the Finger-tip pulse oximeter system should assure that it is used in such an environment.
Immunity test IEC 60601 test level Compliance level Electromagnetic environment - guidance
Conducted RF IEC
61000-4-6 3 Vrms ; 6 Vrms ; 150
kHz to 80 MHz N/A Portable and mobile RF communications equipment
should be used no closer to any part of the EQUIPMENT
or SYSTEM including cables, than the recommended
separation distance calculated from the equation ap-
plicable to the frequency of the transmitter. Interference
may occur in the vicinity of equipment marked with the
following symbol.
Radiated RF IEC
61000-4-3 3 V/m ; 10V/m ; 80
MHz – 2.7 GHz l 80% 3 V/m ; 10V/m ; 80
MHz – 2.7 GHz ; 80%
Proximity elds
from RF wireless
Communications
equipment IEC
61000-4-3
27 V/m 385 MHz 27 V/m 385 MHz
28 V/m 450 MHz 28 V/m 450 MHz
9 V/m 710 MHz 9 V/m 710 MHz
745 MHz 745 MHz
780 MHz 780 MHz
28 V/m 810 MHz 28 V/m 810 MHz
870 MHz 870 MHz
930 MHz 930 MHz
28 V/m 1720 MHz 28 V/m 1720 MHz
1845 MHz 1845 MHz
1970 MHz 1970 MHz
28 V/m 2450 MHz 28 V/m 2450 MHz
9 V/m 5240 MHz 9 V/m 5240 MHz
5500 MHz 5500 MHz
5785 MHz 5785 MHz
Declaration – electromagnetic immunity
The Finger-tip pulse oximeter system is intended for use in the electromagnetic environment specied below. The customer or
the user of the Finger-tip pulse oximeter system should assure that it is used in such an environment.
Immunity test IEC 60601 test level Compliance level Electromagnetic environment - guidance
Electrostatic discharge
(ESD) IEC 61000-4-2 ±8 kV contact
±2 kV, ±4 kV, ±8 kV,
±15 kV air
±8 kV contact
±2 kV, ±4 kV, ±8
kV, ±15 kV air
Floors should be wood, concrete or ceramic tile.
If oors are covered with synthetic material, the
relative humidity should be at least 30 %.
Electrical fast transient/
burst IEC 61000-4-4 ±2 kV for power supply lines
±1 kV for input/output lines N/A Mains power quality should be that of a typical
commercial or hospital environment.
Surge IEC 61000-4-5 ±0.5 kV
±1 kV dierential mode
±2 kV common mode
N/A Mains power quality should be that of a typical
commercial or hospital environment.
Voltage dips, short
interruptions and voltage
variations on power
supply input lines IEC
61000-4-11
0 % UT; 0, 5 cycle At 0°, 45°,
90°, 135°, 180°, 225°, 270°
and 315°
0 % UT; 1 cycle And 70
% UT; 25/30 cycle Single
phase: at 0°
N/A Mains power quality should be that of a typical
commercial or hospital environment. If the user
of the EQUIPMENT or SYSTEM requires continued
operation during power mains interruptions, it is
recommended that the EQUIPMENT or SYSTEM
be powered from an uninterruptible power sup-
ply or a battery.
Power frequency (50/60
Hz) magnetic eld IEC
61000-4-8
30 A/m 30 A/m Power frequency magnetic elds should be at
levels characteristic of a typical location in a typi-
cal commercial or hospital environment.
IN0SB100000000252
Warranty Card
This instrument is covered by a 2 year guarantee from the date of purchase, batteries
and accessories are not included. The guarantee is valid only on presentation of the
guarantee card completed by the dealer confirming date of purchase or the receipt.
Opening or altering the instrument invalidates the guarantee. The guarantee does
not cover damage, accidents or non-compliance with the instruction manual. Please
contact your local seller/dealer or www.rossmax.com.
Customer Name: __________________________________
Address: _______________________________________
Telephone: _____________________________________
E-mail address: ___________________________________
Product Information: ______________________________
Date of purchase: _________________________________
Store where purchased:
______________________________________________
RI_IB_SB100_EN_
2Yr_TP_ver2004
1

Hulp nodig? Stel uw vraag in het forum

Spelregels

Misbruik melden

Gebruikershandleiding.com neemt misbruik van zijn services uitermate serieus. U kunt hieronder aangeven waarom deze vraag ongepast is. Wij controleren de vraag en zonodig wordt deze verwijderd.

Product:

Bijvoorbeeld antisemitische inhoud, racistische inhoud, of materiaal dat gewelddadige fysieke handelingen tot gevolg kan hebben.

Bijvoorbeeld een creditcardnummer, een persoonlijk identificatienummer, of een geheim adres. E-mailadressen en volledige namen worden niet als privégegevens beschouwd.

Spelregels forum

Om tot zinvolle vragen te komen hanteren wij de volgende spelregels:

Belangrijk! Als er een antwoord wordt gegeven op uw vraag, dan is het voor de gever van het antwoord nuttig om te weten als u er wel (of niet) mee geholpen bent! Wij vragen u dus ook te reageren op een antwoord.

Belangrijk! Antwoorden worden ook per e-mail naar abonnees gestuurd. Laat uw emailadres achter op deze site, zodat u op de hoogte blijft. U krijgt dan ook andere vragen en antwoorden te zien.

Abonneren

Abonneer u voor het ontvangen van emails voor uw Rossmax SB100 bij:


U ontvangt een email met instructies om u voor één of beide opties in te schrijven.


Ontvang uw handleiding per email

Vul uw emailadres in en ontvang de handleiding van Rossmax SB100 in de taal/talen: Engels als bijlage per email.

De handleiding is 1.96 mb groot.

 

U ontvangt de handleiding per email binnen enkele minuten. Als u geen email heeft ontvangen, dan heeft u waarschijnlijk een verkeerd emailadres ingevuld of is uw mailbox te vol. Daarnaast kan het zijn dat uw internetprovider een maximum heeft aan de grootte per email. Omdat hier een handleiding wordt meegestuurd, kan het voorkomen dat de email groter is dan toegestaan bij uw provider.

Stel vragen via chat aan uw handleiding

Stel uw vraag over deze PDF

loading

Andere handleiding(en) van Rossmax SB100

Rossmax SB100 Gebruiksaanwijzing - Deutsch - 2 pagina's


Uw handleiding is per email verstuurd. Controleer uw email

Als u niet binnen een kwartier uw email met handleiding ontvangen heeft, kan het zijn dat u een verkeerd emailadres heeft ingevuld of dat uw emailprovider een maximum grootte per email heeft ingesteld die kleiner is dan de grootte van de handleiding.

Er is een email naar u verstuurd om uw inschrijving definitief te maken.

Controleer uw email en volg de aanwijzingen op om uw inschrijving definitief te maken

U heeft geen emailadres opgegeven

Als u de handleiding per email wilt ontvangen, vul dan een geldig emailadres in.

Uw vraag is op deze pagina toegevoegd

Wilt u een email ontvangen bij een antwoord en/of nieuwe vragen? Vul dan hier uw emailadres in.



Info