Instructions for Use—English
Onyx
®
Vantage 9590 Finger Pulse Oximeter
Indications for Use
The Nonin
®
Onyx Vantage 9590 Finger Pulse Oximeter is a small, lightweight, portable device indicated for use in measuring
and displaying functional oxygen saturation of arterial hemoglobin (%SpO
2
) and pulse rate of patients who are well or poorly
perfused. It is intended for spot-checking of adult and pediatric patients on digits (fingers, thumb, toes) that are between 0.3–
1.0 inch (0.8 – 2.5 cm) thick. The device’s intended use environments include hospitals, clinics, long-term care facilities,
skilled nursing facilities, emergency medical services, and home healthcare services.
CAUTION: Regulatory authorities outside the U.S. recognize the use of this device in motion conditions.
Warnings
Do not use the device in an MR environment, in an explosive atmosphere, or on neonatal patients.
This device is not defibrillation proof per IEC 60601-1.
Inspect the sensor application site at least every 4 hours to ensure correct sensor alignment and skin integrity. Patient sensitivity to
the sensor may vary due to medical status or skin condition.
Avoid excessive pressure to the sensor application site as this may cause damage to the skin beneath the sensor.
This device is intended only as an adjunct in patient assessment. It must be used in conjunction with other methods of assessing
clinical signs and symptoms.
The device must be able to measure the pulse properly to obtain an accurate SpO
2
measurement. Verify that nothing is hindering the
pulse measurement before relying on the SpO
2
measurement.
Operation of this device below the minimum amplitude of 0.3% modulation may cause inaccurate results.
General operation of the device may be affected by the use of an electrosurgical unit (ESU).
This device should not be used adjacent to or stacked with other equipment. If adjacent or stacked use is necessary, the device should
be observed carefully to verify normal operation.
Keep the oximeter away from young children. Small items such as the battery door, battery, and lanyard are choking hazards.
Certain activities may pose a risk of injury, including strangulation, if lanyard should become wrapped around your neck.
Before changing batteries, make sure the device is off and is not applied to a digit.
Portable RF communications equipment such as cell phones or radios (including peripherals such as antenna cables and external
antennas) should be used no closer than 30 cm (12 inches) to any part of the ME system, including cables specified by the
manufacturer. Otherwise, degradation of the performance of this equipment could result.
Cautions
This device has no audible alarms and is intended only for spot-checking.
This device is designed to determine the percentage of arterial oxygen saturation of functional hemoglobin. Factors that may degrade
pulse oximeter performance or affect the accuracy of the measurement include the following:
The device may not work when circulation is reduced. Warm or rub the finger, or re-position the device.
This device’s display will go blank after 30 seconds of no readings or poor readings.
In some circumstances, the device may interpret motion as good pulse quality. Minimize patient motion as much as possible.
Clean the device before applying it to a patient.
Do not sterilize, autoclave, or immerse this device in liquid. Do not pour or spray any liquids onto the device.
Do not use caustic or abrasive cleaning agents, or any cleaning agent containing ammonium chloride.
This device is a precision electronic instrument and must be repaired by Nonin Technical Service. Field repair of the device is not
possible. Do not attempt to open the case or repair the electronics. Opening the case may damage the device and void the warranty.
A flexible circuit connects the two halves. Do not twist or pull the flexible circuit or overextend the device’s spring. Do not hang the
lanyard from the device’s flexible circuit.
A functional tester cannot be used to assess the accuracy of a pulse oximeter monitor.
This equipment complies with IEC 60601-1-2 for electromagnetic compatibility for medical electrical equipment and/or systems. This
standard is designed to provide reasonable protection against harmful interference in a typical medical installation. However, because
of the proliferation of radio-frequency transmitting equipment and other sources of electrical noise in healthcare and other
environments, it is possible that high levels of such interference due to close proximity or strength of a source might disrupt the
performance of this device. Medical electrical equipment needs special precautions regarding EMC, and all equipment must be
installed and put into service according to the EMC information specified in this manual.
Batteries may leak or explode if used or disposed of improperly. Remove batteries if the device will be stored for more than 30 days.
Do not use different types of batteries at the same time. Do not mix fully charged and partially charged batteries at the same time.
These actions may cause the batteries to leak.
Follow local, state, and national governing ordinances and recycling instructions regarding disposal or recycling of the device and
device components, including batteries.
In compliance with the European Directive on Waste Electrical and Electronic Equipment (WEEE) 2002/96/EC, do not dispose of this
product as unsorted municipal waste. This device contains WEEE materials; please contact your distributor regarding take-back or
recycling of the device. If you are unsure how to reach your distributor, please call Nonin for your distributor’s contact information.
When using the device in the home, avoid exposing the device to lint and dust.
%SpO
2
!
do not apply the pulse oximeter on the
same arm as a blood pressure cuff,
arterial catheter or infusion line(s) (IVs)
excessive light, such as sunlight or direct
home lighting
excessive motion
moisture in the device
improperly applied device
finger is outside recommended size
range
poor pulse quality
venous pulsations
anemia or low hemoglobin concentrations
cardiogreen and other intravascular dyes
carboxyhemoglobin
methemoglobin
dysfunctional hemoglobin
artificial nails or fingernail polish
residue (e.g., dried blood, dirt, grease, oil)
in the light path
1

Hulp nodig? Stel uw vraag in het forum

Spelregels
1

Forum

Nonin-Onyx-Vantage-9590

Zoeken resetten

Misbruik melden

Gebruikershandleiding.com neemt misbruik van zijn services uitermate serieus. U kunt hieronder aangeven waarom deze vraag ongepast is. Wij controleren de vraag en zonodig wordt deze verwijderd.

Product:

Bijvoorbeeld antisemitische inhoud, racistische inhoud, of materiaal dat gewelddadige fysieke handelingen tot gevolg kan hebben.

Bijvoorbeeld een creditcardnummer, een persoonlijk identificatienummer, of een geheim adres. E-mailadressen en volledige namen worden niet als privégegevens beschouwd.

Spelregels forum

Om tot zinvolle vragen te komen hanteren wij de volgende spelregels:

Belangrijk! Als er een antwoord wordt gegeven op uw vraag, dan is het voor de gever van het antwoord nuttig om te weten als u er wel (of niet) mee geholpen bent! Wij vragen u dus ook te reageren op een antwoord.

Belangrijk! Antwoorden worden ook per e-mail naar abonnees gestuurd. Laat uw emailadres achter op deze site, zodat u op de hoogte blijft. U krijgt dan ook andere vragen en antwoorden te zien.

Abonneren

Abonneer u voor het ontvangen van emails voor uw Nonin Onyx Vantage 9590 bij:


U ontvangt een email met instructies om u voor één of beide opties in te schrijven.


Ontvang uw handleiding per email

Vul uw emailadres in en ontvang de handleiding van Nonin Onyx Vantage 9590 in de taal/talen: Engels als bijlage per email.

De handleiding is 2,48 mb groot.

 

U ontvangt de handleiding per email binnen enkele minuten. Als u geen email heeft ontvangen, dan heeft u waarschijnlijk een verkeerd emailadres ingevuld of is uw mailbox te vol. Daarnaast kan het zijn dat uw internetprovider een maximum heeft aan de grootte per email. Omdat hier een handleiding wordt meegestuurd, kan het voorkomen dat de email groter is dan toegestaan bij uw provider.

Stel vragen via chat aan uw handleiding

Stel uw vraag over deze PDF

loading

Andere handleiding(en) van Nonin Onyx Vantage 9590

Nonin Onyx Vantage 9590 Gebruiksaanwijzing - Nederlands - 9 pagina's

Nonin Onyx Vantage 9590 Gebruiksaanwijzing - Deutsch - 9 pagina's


Uw handleiding is per email verstuurd. Controleer uw email

Als u niet binnen een kwartier uw email met handleiding ontvangen heeft, kan het zijn dat u een verkeerd emailadres heeft ingevuld of dat uw emailprovider een maximum grootte per email heeft ingesteld die kleiner is dan de grootte van de handleiding.

Er is een email naar u verstuurd om uw inschrijving definitief te maken.

Controleer uw email en volg de aanwijzingen op om uw inschrijving definitief te maken

U heeft geen emailadres opgegeven

Als u de handleiding per email wilt ontvangen, vul dan een geldig emailadres in.

Uw vraag is op deze pagina toegevoegd

Wilt u een email ontvangen bij een antwoord en/of nieuwe vragen? Vul dan hier uw emailadres in.



Info