TERMOMETRO A DISTANZA A INFRAROSSI
NON-CONTACT INFRARED BODY THERMOMETER
THERMOMÈTRE À INFRAROUGES
SANS CONTACT
TERMÓMETRO INFRARROJO
SIN CONTACTO
M25583-M-Rev.0-04.20
PROFESSIONAL MEDICAL PRODUCTS
25583
Hetaida Technology Co., Ltd.
4F, BaiShiDa High-Tech Park, XiangDong
Industrial Area, DaLingShan Town, DongGuan City,
Guangdong, China
Made in China
WellKang Ltd,
16 Castle St, Dover, CT16 1PW Dover,
CT16 1PW, UK
0598
IP22
Importato da / Imported by / Importé de / Importado de:
Gima S.p.A.
Via Marconi, 1
20060 Gessate (MI) Italy
Modello / Model / Modèle / Modelo: HTD8816C
18
Safety Information
This device may only be used for the purposes described in these instructions. The
manufacturer cannot be held liable for damage caused by incorrect application.
The non contact infrared body thermometer is designed to minimize the possibility of
hazards from errors in the software program by following sound and light engineering
design processes, Risk Analysis and Software Validation.
Warning
The non contact infrared body thermometer is a medical device, and can be used
repeatedly, which using life is 5 years.
The Non Contact Infrared Body Thermometer is to be operated by consumers in the
home setting and primary care setting as screening tool. This manual, accessories,
Directions for Use, all precautionary information, and specications should be read
before use.
This product is designed to measure human body temperature on the forehead. Do
not use it for any other purpose.
This product is intended in the home setting and primary care setting as screening
tool.
Do not use the thermometer if it malfunctions or has been damaged in any matter.
When the ambient temperature of the thermometer changes too much, such as
moving the Thermometer from one place of lower temperature to another place
of higher temperature, Allow the thermometer to remain in a room for 30 minutes
where the temperature is between15℃to 40℃ (59℉- 104℉).
Remove primary batteries if equipment is not likely to be used for long time.
This product is not waterproof, do not be immersed in water or other liquid; If clean-
ing and disinfection, please follow the “Care and Storage” section requirements.
Do not touch the sensor of infrared detection with your ngers.
If a cold compress on the forehead fever patients, or take other measures to cool
down the temperature data will low, should be avoided in this case to measure body
temperature.
If measure human forehead temperature , please select “body” mode; for measure
other objects, liquids, food and other temperature please select “surface” mode.
This product must be operated in a stable environment, if the ambient environment
was mutations, please should be note whether there is fog on the sensor, if any,
before using accordance with the “Care and Storage” section to removing the fog.
Do not near strong electrostatic eld or strong magnetic elds, thus avoiding the
impact on the accuracy of the measurement data.
Do not mix the old and new batteries to avoid damage to the product.
It may affect the accuracy of measurements when the forehead is covered by hair,
ENGLISH
19
perspiration, cap or scarf.
The measuring result of this product is only for your reference. If you have any
doubt, please measure the temperature in other methods.
- The device should be kept out of the reach of children/pets. When not in use, store
the device in a dry room and protect it against extreme moisture, heat, lint, dust and
direct sunlight. Never place any heavy objects on the storage case.
- Do not throw batteries into re.
- Only use recommended batteries. Do not use rechargeable batteries.
- This thermometer will irreplaceable the diagnostic in hospitals.
- Do not fall, disassemble or modify the device.
- Do not use this device if you think it is damaged or notice anything unusual.
- This device comprises sensitive components and must be treated with caution.
Observe the storage and operating conditions described in the ‘Technical Speci-
cations’ section.
- Not servicing/maintenance while the thermometer is in use.
- When using, shall not touch battery and the patient simultaneously.
- Do not use the device if it is damaged/ degraded/loosened in any way. The contin-
uous use of a damaged unit may cause injury, improper results, or serious danger.
- Based on the current science and technology, other potential allergic reactions are
unknown.
- This equipment needs to be installed and put into service in accordance with the
information provided in the ACCOMPANYING DOCUMENTS.
1. Overview
Intended Use
The Non Contact Infrared Body Thermometers are designed to be used for intermit-
tent measurement and monitoring of human body temperature by consumers in the
home setting and primary care setting as screening tool.
Description of Non Contact Infrared Body Thermometer
Device principle and introduction
The Non contact infrared body thermometer are hand-held, reusable, battery oper-
ated devices, which can measure human body temperature on forehead, the skin
temperature on one’s forehead.
The operation principle is based on Infrared Sensor technology. The IR sensor can
output different signal when measuring different object temperature or in different
ambient temperature, and the ASIC can turn the signal from IR Sensor to a digital
ENGLISH
20
value and display it on the LCD.
Description on Controls and Indicators.
Thermometer Applications
Thermometer Style Adult Pediatric
Ear Forehead Ear Forehead
Non Contact Infrared Body
Thermometers
1. IR sensor 9. Data indicator
2. Liquid crystal display(LCD) 10. Indicator of measurement result
3. MODE button 11. Low Battery indicator
4. Memo button 12. Voice on/off indicator
5. On/Scan button 13. Memory Number
6. Battery Cover 14. Fahrenheit
7. Surface mode 15. Celsius
8. Body mode 16. Memory indicator
ENGLISH
21
Technical Specications
Measurement Unit ℃/℉
Operating mode Adjusted mode (Body mode)
Direct mode (surface mode)
Reference Body Site Axillary
Rated output range Body mode: 34.0℃~43.0℃ / 93.2°F -109.4°F
Surface mode: 0℃~100℃ / 32℉~212℉
Output Range Body mode: 34.0℃~43.0℃/ 93.2°F -109.4°F
Surface mode: 0℃~100.0℃/ 32°F -212°F
Laboratory Accuracy Body mode:
34.0℃~34.9℃:±0.3℃ / 93.2°F-94.8°F: ±0.5°F;
35.0℃~42.0℃:±0.2℃ / 95.0°F-107.6°F: ±0.4°F;
42.1℃~43.0℃:±0.3℃ / 107.8°F-109.4°F: ±0.5°F;
Surface mode: ±2℃/±3.6°F.
Display Resolution 0.1℃/0.1℉
Three-color Backlight
(Color Alarm)
35.5-37.3°C / 95.9-99.1°F:
Green (Normal Temperature);
37.4-38.0°C (Alarm point) / 99.3-100.4°F:
Yellow (Slight Fever)
38.1-43.0°C / 100.6-109.4°F:
Red (High Fever)
Note:
1.Surface mode is always with Green backlight.
2. In body mode 34.0-35.4°C is with green backlight.
Auto Power Off Time ≤18s
Measuring Time ≤2s
Measuring Distance 1 cm -5cm (0.4 in -2in)
Memories 50
Power Supply
Requirements
Batteries 1.5V (AAA) Alkaline battery x2 (IEC Type LR03)
Adaptable Range 2.6V~3.6V
ENGLISH
22
Environmental
Operating Condition Operating Temperature: 15℃~40℃
(59°F ~104°F),
Relative Humidity ≤85%,
atmospheric pressure:70 Kpa -106 Kpa
Transport and Storage
Condition
Storage Temperature:-20°C -55°C /
(-4°F - 131°F),
Relative Humidity ≤93%,
Atmospheric pressure:70 Kpa -106 Kpa
Dimension and Weighting
Weight (without batteries) 80g
Size L:122mm X W:96mm X H:42mm
Compliance
Item Compliant with
Equipment classication Safety Standards: EN 60601-1: 2006+A1:2013,
EN 60601-1-2: 2015
Type of protection Internally powered equipment (on battery power)
Degree of protection Non Applied part
Front panel and case
labeling
EN ISO15223-1:2016
Temperature EN ISO80601-2-56:2017
Home healthcare
environment
EN 60601-1-11:2015
Protection against electric shock Internally powered ME equipment
Applied part Non Applied part
Protection against harmful ingress of
water or particulate matter
IP22
Mode of operation Continuous operation
Safety classication of ME EQUIPMENT
Note: Not intended to be sterilized. Not for use in an OXYGEN RICH ENVIRONMENT
ENGLISH
23
2. Operation
2.1 Battery installation
Caution: The Non Contact Infrared Body Thermometer does not operate with dead
batteries and does not input outer power. Install new batteries.
1. Pull the battery downward, toward the bottom of the Non Contact Infrared Body
Thermometer, and remove the battery access door;
2. Insert two pieces AAA size batteries according to the “+” and “-”;
3. Close the battery cover.
2.2 How to Operate
Before Applying the Thermometer
Be sure to read and understand all warnings listed of the instructions before use.
The thermometer is aligned with the middle of the forehead to measure body tem-
perature (between the eyebrows above) and keep the vertical distance, press the
On/Scan button, temperature display immediately, see gure 2.
Figure 2-Measuring position and distance
When the ambient temperature of the thermometer changes too much, such as
moving the Thermometer from one place of lower temperature to another place
of higher temperature, Allow the thermometer to remain in a room for 30 minutes
where the temperature is between15℃ to 40℃.
The ambient temperature around the test person should be stable, should keep
away from the larger ow fan, air-conditioning vents and so on.
When people moving from one place of lower temperature to another place of high-
er temperature, should at least remain in the test environment more than 5 minutes,
to be consistent with the ambient temperature after the re-measurement.
Wait at least 1 second for the next measurement. If the continuous measurement
of ve times, it is recommended to wait at least 30 seconds and then continue
measurement.
ENGLISH
24
You cannot use the thermometer in place where the sun is strong.
If for some reason the low forehead temperature measurement can try to align
behind the ears. See gure 3.
Figure 3- Align behind the ears
to measurement
General Setup and Use
Start measuring
1. Turn on the thermometer by pressing the On/Scan button. The thermometer will
perform self-test with all segments displayed for 2 seconds.
2. Align staff forehead to keep the distance, and then press the On/Scan button to
start the measurement, read the data.
Note: 1) After full display over, you will hear a rattle or “beep beep beep” three times,
which means that the measurements have been completed, while the target value of
the measured temperature is displayed on the LCD, while backlit display according
to the appropriate setting among the three colors red, green, yellow one of. And the
Green means ready for next measurement. When 37.4℃-38.0℃, it’s yellow ,means
slight fever warning. Please pay attention to body temperature. When the body tem-
perature is above 38.1℃, it’s red, means high fever. Please take action to cool down
or go for a doctor.
2) To ensure the accuracy of the measurement, wait at least 30 seconds after 5 con-
secutive measurements.
Mode conversion
When the device is running, pressing the MODE button to cycle conversion between
“body” mode and “surface” mode.
“body” mode is used for measuring human body temperature, the “surface “ mode
is used to measure the surface temperature. (The factory default is “body” mode).
Recalling and Erasing Memory Data
The last temperature taken before the thermometer powers off is stored in memory,
up to 50.
ENGLISH
25
As below step:
1. In the boot or shutdown state, short press the MEMO button to view the history of
measured values.
2. An empty memory cell shows “---℃” or “---℉”.
3. Temperature readings can be stored in memory. Up to 50 temperature readings can
be stored into the memory cells and automatically overwrite historical data.
4. In boot mode, press the MEMO button until the LCD display “CLR”, which means
that all stored data is cleared completely.
Parameter settings
This product can be set according to the subjects of different colors and different
environments data to meet the different characteristics of populations or individuals.
As below step:
1. Unit Set-F1
Under the boot mode, long press the MODE button to enter F1, press the “MODE”
or “MEMO”button to switch Celsius and Fahrenheit temperature units, press the
On/Scan button to conrm the unit settings (factory default is Celsius).
2. Fever alert set-F2
Under F1 state ,press On/Scan button to enter the F2, press the “MODE” button to
plus 0.1℃, press the “MEMO” button decrease 0.1 ℃, long press to accelerate the
speed of temperature regulation, and nally press the On/Scan button to save. (The
factory default is 38.1℃).
3. Prompt sound settings-F3
Under F2 state, press On/Scan button to enter F3, press MODE button or MEMO
button to set voice switch, and press the On/Scan button to conrm the settings.
(The factory default is the voice Prompt to open).
4. Exit setting mode
In the F3 mode, press the On/Scan button will automatically turn off the screen, exit
setting.
Prompt sound settings ON/OFF function
Under the boot mode .Short press power on/off button ,at the same time press the
MODE button to take on or take off the Prompt sound function.
Restore to factory setting function
Under the boot mode .Short press power on/off button ,at the same time press the
MEMO button. Until LCD display “rst”. Two seconds later ,former F1-F3 parameter
back to factory setting.
ENGLISH
26
3. Troubleshooting
MESSAGE SITUATION SOLUTION
Temperature taken in not within
Typical human temperature
range.
(34.0~43.0℃ or
93.2℉~109.4℉).
Make sure the forehead
thermometer is for forehead
measurement, not other human
body site.
Measured over the distance 1-5
cm (0.4-2in). Optimum measurement distance
is 1cm.
Incorrect test position. See gure 2 Measuring position
and distance.
Subjects forehead hair,
Antipyretic stickers, head with
sweat, etc.
Subjects sit quietly 5-10 minutes
before the test.
Some people’s body tempera-
ture is lower than the general
population.
The main concern fever
temperature,
Operating temperature
exceeds the range of specied
temperature.
Move to a room within the
operating range wait 30 minutes
before taking temperature.
The screen icker, automatic
turn off. Replace battery or the product
has been damaged, needs
repairs.
Battery capacity is too low.
Taking temperature is not
allowed.
Install a new battery
Ambient temperature changes
too fast. Wait until the ambient
temperature is stably.
ENGLISH
27
1) Power is off.
2) Improper battery installation.
3) The battery is exhausted.
4) Display remains blank.
1) Press ON button again.
2) Check the battery polarity.
3) Replace with a new battery.
4) Contact the retailer or service
center.
4. Replacing the Battery
1. Open and release battery cover following indicator on the surface of battery cover.
Before changing the battery be sure the system is already power off.
2. Remove the battery and replace with 2 new one, type AAA, Make sure align prop-
erly as indicated inside the battery cover.
3. Slide the battery cover back in until it snaps in place.
4. Do not dispose of used batteries in household waste. Take them to special local
collection sites.
5. In case, if system is latched up after changing battery. You may not follow up the
process of rule one. Just take off battery, waiting 30 sec, then load battery again.
Warning
Do not recharge, disassemble or dispose of in re.
1. The typical service life of the new and unused batteries is 2000 measurements for
the operation time is 18s.
2. Only use the recommended batteries, do not recharge non-rechargeable batteries
and do not burn them.
3. Remove the batteries if the thermometer is not to be used for a long period.
5. Cleaning, Care and Storage
The lens is very delicate.
It is very important to protect the lens from dirt and damage.
Use a clean, soft cloth to clean the surface of the device and LCD. Do not use
solvents or immerse the device into water or other liquids.
Always keep the thermometer a within the storage temperature and humidity range
as specied.
It is recommended to store the thermometer in a dry location free from dust.
It is recommended to store the thermometer in a dry location free from dust. Do not
expose the thermometer to direct sunlight, high temperature/ humidity or any extreme
environment, otherwise the function will be reduced.
When the ambient temperature of the thermometer changes too much, such as mov-
ing the thermometer from one place of lower temperature to another place of higher
ENGLISH
28
temperature, allow the thermometer to remain in a room for 30 minutes where the
temperature is between 15℃ to 40℃.
EMC Declaration
1. This equipment needs to be installed and put into service in accordance with the
information provided in the ACCOMPANYING DOCUMENTS;
This product needs special precautions regarding EMC and needs to be installed
and put into service according to the EMC information provided, and this unit can
be affected by portable and mobile RF communications equipment.
2. * Caution: Do not use a mobile phone or other devices that emit electromagnetic
elds, near the unit. This may result in incorrect operation of the unit.
3. * Caution: This unit has been thoroughly tested and inspected to assure proper
performance and operation!
4. * Caution: this machine should not be used adjacent to or stacked with other equip-
ment and that if adjacent or stacked use is necessary, this machine should be
observed to verify normal operation in the conguration in which it will be used.
Guidance and manufacturer’s declaration – electromagnetic emission
The Non Contact Infrared Body Thermometer is intended for use in the electromagnetic
environment specied below. The customer of the user of the Non Contact Infrared Body
Thermometer should assure that it is used in such an environment.
Emission test Compliance Electromagnetic environment – guidance
RF emissions CISPR 11 Group 1 The Non Contact Infrared Body Thermome-
ter use RF energy only for its internal func-
tion. Therefore, its RF emissions are very
low and are not likely to cause any interfer-
ence in nearby electronic equipment.
RF emission CISPR 11 Class B The Non Contact Infrared Body Thermome-
ter is suitable for use in all establishments,
other than domestic and those directly con-
nected to the public low-voltage power sup-
ply network that supplies buildings used for
domestic purposes.
Harmonic emissions
IEC 61000-3-2 Not
applicable
Voltage uctuations/ icker
emissions IEC 61000-3-3 Not
applicable
ENGLISH
29
Guidance and manufacturer’s declaration – electromagnetic immunity
The Non Contact Infrared Body Thermometer is intended for use in the electromagnetic
environment specied below. The customer or the user of Non Contact Infrared Body Ther-
mometer should assure that it is used in such an environment.
Immunity test IEC 60601 test
level Compliance
level Electromagnetic
environment - guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
±6 kV contact
±15 kV air ±6 kV contact
±15 kV air Floors should be wood, con-
crete or ceramic tile. If oor are
covered with synthetic material,
the relative humidity should be
at least 30%.
Electrical fast
transient/burst
IEC 61000-4-4
±2 kV for power
supply lines
±1 kV for input/
output lines
Not
applicable Mains power quality should be
that of a typical commercial or
hospital environment.
Surge
IEC 61000-4-5 ± 1 kV line(s)
to line(s)
± 2 kV line(s)
to earth
Not
applicable Mains power quality should be
that of a typical commercial or
hospital environment.
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
5% UT
(>95% dip in UT)
for 0.5 cycle
40% UT
(60% dip in UT)
for 5 cycles
70% UT
(30% dip in UT)
for 25 cycles
<5% UT
(>95% dip in UT)
for 5 sec
Not
applicable Mains power quality should
be that of a typical commer-
cial or hospital environment.
If the user of the Non Contact
Infrared Body Thermometer
requires continued operation
during power mains interrup-
tions, it is recommended that
the Non Contact Infrared Body
Thermometer be powered from
an uninterruptible power supply
or a battery.
Power frequency
(50Hz/60Hz)
magnetic eld
IEC 61000-4-8
30A/m 30A/m Power frequency magnetic
elds should be at levels char-
acteristic of a typical location in
a typical commercial or hospital
environment.
NOTE: UT is the a.c. mains voltage prior to application of the test level.
ENGLISH
30
ENGLISH
Guidance and manufacturer’s declaration – electromagnetic immunity
The Non Contact Infrared Body Thermometer is intended for use in the electromagnetic
environment specied below. The customer or the user of the Non Contact Infrared Body
Thermometer should assure that it is used in such an environment.
Immunity test IEC 60601 test
level Compliance
level Electromagnetic
environment - guidance
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3 Vrms
150 kHz to 80
MHz
10V/m
80 MHz
to 2.5 GHz
Not
applicable
10 V/m
Portable and mobile RF com-
munications equipment should
be used no closer to any part of
the Non Contact Infrared Body
Thermometer, including cables,
than the recommended sepa-
ration distance calculated from
the equation applicable to the
frequency of the transmitter.
Recommended separation
distance
80MHz to 800MHz
80MHz to 800MHz
Where P is the maximum out-
put power rating of the trans-
mitter in watts (W) according
to the transmitter manufacturer
and d is the recommended sep-
aration distance in metres (m).
Field strengths from xed RF
transmitters, as determined
by an electromagnetic site
survey,a should be less than
the compliance level in each
frequency range.b
Interference may occur in the
vicinity of equipment
marked with the
following symbol:
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reection from structures, objects and people.
d= 1.2 √P
d= 1.2 √P
d= 2.3 √P
31 ENGLISH
a Field strengths from xed transmitters, such as base stations for radio (cellular/cordless)
telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV
broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic
environment due to xed RF transmitters, an electromagnetic site survey should be consid-
ered. If the measured eld strength in the location in which the Non Contact Infrared Body
Thermometer is used exceeds the applicable RF compliance level above, the Non Contact
Infrared Body Thermometer should be observed to verify normal operation. If abnormal
performance is observed, additional measures may be necessary, such as re-orienting or
relocating the Non Contact Infrared Body Thermometer.
b Over the frequency range 150 kHz to 80 MHz, eld strengths should be less than 10 V/m.
Recommended separation distances between portable and mobile RF communi-
cations equipment and the Non Contact Infrared Body Thermometer.
The Non Contact Infrared Body Thermometer is intended for use in an electromagnetic en-
vironment in which radiated RF disturbances are controlled. The customer or the user of the
Non Contact Infrared Body Thermometer can help prevent electromagnetic interference by
maintaining a minimum distance between portable and mobile RF communications equip-
ment (transmitters) and the Non Contact Infrared Body Thermometer as recommended be-
low, according to the maximum output power of the communications equipment.
Rated maximum
output power of
transmitter (W)
Separation distance according to frequency of transmitter (m)
150KHz to 80MHz 80MHz to 800MHz 800MHz to 2.5GHz
0.01 0.12 0.12 0.23
0.1 0.38 0.38 0.73
1 1.2 1.2 2.3
10 3.8 3.8 7.3
100 12 12 23
For transmitters rated at a maximum output power not listed above, the recommended
separation distance d in metres (m) can be estimated using the equation applicable to the
frequency of the transmitter, where P is the maximum output power rating of the transmitter
in watts (W) according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range
applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reection from structures, objects and people.
d= 1.2 √P d= 1.2 √P d= 2.3 √P
32
ENGLISH
Caution: read instructions
(warnings) carefully Follow instructions for use
Manufacturer Date of manufacture
Product code Lot number
IP22 Covering Protection rate Authorized representative
in the European community
WEEE disposal Medical Device complies
with Directive 93/42/EEC
Recyclable PVC Recyclable
This device compiles with Part 15
of FCC( Federal Communications
Commission) Rules
Restriction of Hazardous
Substances
Temperature limitation Moisture limitation
Atmospheric pressure limitation Serial number
Disposal: The product must not be disposed of along with other domestic
waste. The users must dispose of this equipment by bringing it to a specic
recycling point for electric and electronic equipment.
GIMA WARRANTY TERMS
The Gima 12-month standard B2B warranty applies.
1

Hulp nodig? Stel uw vraag in het forum

Spelregels

Misbruik melden

Gebruikershandleiding.com neemt misbruik van zijn services uitermate serieus. U kunt hieronder aangeven waarom deze vraag ongepast is. Wij controleren de vraag en zonodig wordt deze verwijderd.

Product:

Bijvoorbeeld antisemitische inhoud, racistische inhoud, of materiaal dat gewelddadige fysieke handelingen tot gevolg kan hebben.

Bijvoorbeeld een creditcardnummer, een persoonlijk identificatienummer, of een geheim adres. E-mailadressen en volledige namen worden niet als privégegevens beschouwd.

Spelregels forum

Om tot zinvolle vragen te komen hanteren wij de volgende spelregels:

Belangrijk! Als er een antwoord wordt gegeven op uw vraag, dan is het voor de gever van het antwoord nuttig om te weten als u er wel (of niet) mee geholpen bent! Wij vragen u dus ook te reageren op een antwoord.

Belangrijk! Antwoorden worden ook per e-mail naar abonnees gestuurd. Laat uw emailadres achter op deze site, zodat u op de hoogte blijft. U krijgt dan ook andere vragen en antwoorden te zien.

Abonneren

Abonneer u voor het ontvangen van emails voor uw Gima HTD8816C bij:


U ontvangt een email met instructies om u voor één of beide opties in te schrijven.


Ontvang uw handleiding per email

Vul uw emailadres in en ontvang de handleiding van Gima HTD8816C in de taal/talen: Engels als bijlage per email.

De handleiding is 0.49 mb groot.

 

U ontvangt de handleiding per email binnen enkele minuten. Als u geen email heeft ontvangen, dan heeft u waarschijnlijk een verkeerd emailadres ingevuld of is uw mailbox te vol. Daarnaast kan het zijn dat uw internetprovider een maximum heeft aan de grootte per email. Omdat hier een handleiding wordt meegestuurd, kan het voorkomen dat de email groter is dan toegestaan bij uw provider.

Stel vragen via chat aan uw handleiding

Stel uw vraag over deze PDF

Andere handleiding(en) van Gima HTD8816C

Gima HTD8816C Gebruiksaanwijzing - Français - 17 pagina's

Gima HTD8816C Gebruiksaanwijzing - Espanõl - 17 pagina's


Uw handleiding is per email verstuurd. Controleer uw email

Als u niet binnen een kwartier uw email met handleiding ontvangen heeft, kan het zijn dat u een verkeerd emailadres heeft ingevuld of dat uw emailprovider een maximum grootte per email heeft ingesteld die kleiner is dan de grootte van de handleiding.

Er is een email naar u verstuurd om uw inschrijving definitief te maken.

Controleer uw email en volg de aanwijzingen op om uw inschrijving definitief te maken

U heeft geen emailadres opgegeven

Als u de handleiding per email wilt ontvangen, vul dan een geldig emailadres in.

Uw vraag is op deze pagina toegevoegd

Wilt u een email ontvangen bij een antwoord en/of nieuwe vragen? Vul dan hier uw emailadres in.



Info