DUSPOL
®
digital
9
safety switch is done by testing the external conductor
(phase indication) to PE (earth). (figure F)
6. External conductor test (phase indication) (gure E)
- Fully grasp the handles L1
8
and L2
9
, in order to en-
sure a capacitive coupling to earth.
- Switch the voltage tester on by briefly actuating the push-
button
7
of the display handle L2
9
(remains switched
on for approx. 10 seconds!). When the device is switched
on, the display shows „0.0“.
- Apply the probe tip L2/+
3
to the system part to be tested.
During the single-pole external conductor test (phase in-
dication), make absolutely sure not to touch the probe tip
L1/-
2
and that it remains contactless.
- If the red LED K and the symbol O light up on the
LC display
6
, the external conductor (phase) of an AC
voltage is applied to this system part.
Note:
The single-pole external conductor test (phase indication)
can be carried out in an earthed mains from 230 V, 50/60 Hz
(phase to earth) on. Protective clothing and insulating condi-
tions on site might impair the function.
Attention!
The absence of voltage can only be determined by means of
a two-pole test.
7. Phase sequence test (gure G/H)
- Fully grasp both handles L1
8
and L2
9
, in order to en-
sure a capacitive coupling to earth.
- Apply the probe tips L1/-
2
and L2/+
3
to two external
conductors (phases) and check whether the external con-
ductor voltage of e.g. 400 V is applied.
- A clockwise phase sequence (phase L1 before phase L2)
is given, if the green LED „►“ of the phase sequence in-
dication
L
and the symbol of the phase sequence
indication
P
light up on the LC display
6
.
- A counter-clockwise phase sequence (phase L2 before
phase L1) is given, if the green LED „◄“ of the phase
sequence indication
L
and the symbol of the phase
sequence indication
P
light up on the LC display
6
.
- The phase sequence test always requires a countercheck
with the probe tips L1/-
2
and L2/+
3
being inverted dur-
ing which the phase sequence must change.
Note:
The phase sequence test can be carried out in an earthed
three-phase mains from 230 V - 900 V, 50/60 Hz (phase to
phase) on. Protective clothing and insulating conditions on site
might impair the function.
8. Continuity test (gure I)
- The continuity test must be carried out on system parts
which are free of voltage. If necessary, capacitors have to
be discharged.
- Apply the two probe tips L1/-
2
and L2/+
3
to the system
parts to be tested.
- In case of continuity (R < 100 kΩ), an acoustic signal is
emitted and the yellow LED Ω
M
for continuity lights up.
- This function also can be used to determine the conduct-
ing and non-conducting direction of semiconductor com-
ponents.
- If a voltage is applied to the test point, the voltage tester
automatically switches over to the voltage test function
and indicates this function.
9. Resistance measurement (gure J)
- The resistance measurement must be carried out on sys-
tem parts which are free of voltage. If necessary, capaci-
tors have to be discharged.
- Short-circuit the probe tips L1/-
2
and L2/+
3
and press
the push-button
7
of the display handle L2
9
once until
the symbol and „Ohm“ are shown on the LC display
6
. Display: „OL“ stands for a measured value which is
outside the measuring range.
- The resistance measurement function is active for approx.
10 seconds.
- Apply the probe tips L1/-
2
and L2/+
3
to the system
parts to be tested in order to measure resistances from
0.1 kΩ to 300 kΩ.
Note:
If required, it is possible to carry out a null balance with the
resistance measurement function being activated. For this pur-
pose, short-circuit the probe tips L1/-
2
and L2/+
3
and press
and hold the push-button
7
of the display handle L2
9
for
approx. 2 seconds until „0.0“ kΩ is shown on the LC display.
10. Diode test (gures K/L)
- The diode test must be carried out on system parts which
are free of voltage. If necessary, capacitors have to be dis-
charged.
- Short-circuit the probe tips L1/-
2
and L2/+
3
and press
the push-button of the display handle L2 twice until the
diode symbol and „diod“ are shown on the LC dis-
play
6
. Display: „OL“ VDC
- The diode test is active for approx. 10 seconds.
- Apply the probe tip L1/-
2
to the cathode and the probe
tip L2/+
3
to the anode of the diode to determine the con-
ducting-state voltage between 0.3 V and 2 V. In case of a
defective (broken down) diode, a voltage value of approx.
0.0 V will be displayed.
- In case of a diode tested in non-conducting direction, the
LC display shows „OL“.
11. Cable break detector (gure M)
- The cable break detector is intended for the non-contact
localization of cable breaks on exposed live lines.
- Switch the voltage tester on by briefly actuating the push-
button
7
of the display handle L2
9
(remains switched
on for approx. 10 seconds!). When the device is switched
on, the display shows „0.0“.
9

Hulp nodig? Stel uw vraag in het forum

Spelregels
1

Forum

Benning-Duspol-Digital

Zoeken resetten

  • Duspol digital LC IP 64 werkt na sluiting op een van de pennen niet meer, zit er in de meter een zekering die vervangen kan worden?

    met vriendelijke groet,

    Bonno Ruigrok Gesteld op 9-1-2016 om 21:35

    Reageer op deze vraag Misbruik melden

Misbruik melden

Gebruikershandleiding.com neemt misbruik van zijn services uitermate serieus. U kunt hieronder aangeven waarom deze vraag ongepast is. Wij controleren de vraag en zonodig wordt deze verwijderd.

Product:

Bijvoorbeeld antisemitische inhoud, racistische inhoud, of materiaal dat gewelddadige fysieke handelingen tot gevolg kan hebben.

Bijvoorbeeld een creditcardnummer, een persoonlijk identificatienummer, of een geheim adres. E-mailadressen en volledige namen worden niet als privégegevens beschouwd.

Spelregels forum

Om tot zinvolle vragen te komen hanteren wij de volgende spelregels:

Belangrijk! Als er een antwoord wordt gegeven op uw vraag, dan is het voor de gever van het antwoord nuttig om te weten als u er wel (of niet) mee geholpen bent! Wij vragen u dus ook te reageren op een antwoord.

Belangrijk! Antwoorden worden ook per e-mail naar abonnees gestuurd. Laat uw emailadres achter op deze site, zodat u op de hoogte blijft. U krijgt dan ook andere vragen en antwoorden te zien.

Abonneren

Abonneer u voor het ontvangen van emails voor uw Benning Duspol Digital bij:


U ontvangt een email met instructies om u voor één of beide opties in te schrijven.


Ontvang uw handleiding per email

Vul uw emailadres in en ontvang de handleiding van Benning Duspol Digital in de taal/talen: Nederlands, Duits, Engels, Frans, Italiaans, Spaans, Pools, Turks, Deens, Zweeds, Noors, Fins als bijlage per email.

De handleiding is 14,29 mb groot.

 

U ontvangt de handleiding per email binnen enkele minuten. Als u geen email heeft ontvangen, dan heeft u waarschijnlijk een verkeerd emailadres ingevuld of is uw mailbox te vol. Daarnaast kan het zijn dat uw internetprovider een maximum heeft aan de grootte per email. Omdat hier een handleiding wordt meegestuurd, kan het voorkomen dat de email groter is dan toegestaan bij uw provider.

Stel vragen via chat aan uw handleiding

Stel uw vraag over deze PDF

loading

Uw handleiding is per email verstuurd. Controleer uw email

Als u niet binnen een kwartier uw email met handleiding ontvangen heeft, kan het zijn dat u een verkeerd emailadres heeft ingevuld of dat uw emailprovider een maximum grootte per email heeft ingesteld die kleiner is dan de grootte van de handleiding.

Er is een email naar u verstuurd om uw inschrijving definitief te maken.

Controleer uw email en volg de aanwijzingen op om uw inschrijving definitief te maken

U heeft geen emailadres opgegeven

Als u de handleiding per email wilt ontvangen, vul dan een geldig emailadres in.

Uw vraag is op deze pagina toegevoegd

Wilt u een email ontvangen bij een antwoord en/of nieuwe vragen? Vul dan hier uw emailadres in.



Info