Page 9/13
10. Input “preset tare”
Input preset tare information.
XX.X of command format is 4 bytes fixed length.
(2 integer bytes, 1 decimal point byte and 1 fraction part byte)
Please input "0" at the blank when integer part is less than 2 digits.
It can be input in the range of 0.0-10.0
Default value is 0.0
(1-2 integer bytes, 1 decimal point byte and 1 fraction part byte)
Command format D0XX.X 00.0-10.0
ACK format D0
Reaction equipment is set preset tare
Transmission example D001.0 preset tare 1.0kg
ACK D0
Parameter error D0!
11. Input “gender”
Input gender information.
X of command format is 1 byte fixed length. (Female=2, Male=1)
Command format D1X
ACK format D1
Reaction equipment is set gender
Transmission example D11 Male
ACK D1
Parameter error D1!
12. Input “body type”
Input body type information.
X of command format is 1 byte fixed length. (Standard=0, Athletic=2)
Command format D2X
ACK format D2
Reaction equipment is set body type
Transmission example D20 Standard
ACK D2
Parameter error D2!
13. Input “height”
Input height information.
XXX.X of command format is 5 bytes fixed length.
(3 integer bytes, 1 decimal point byte and 1 fraction part byte)
Please input "0" at the blank when integer part is less than 3 digits.
It can be input in the range of 90-249
Command format D3XXX
ACK format D3
Reaction equipment is set height
Transmission example D3175.4 height 175.4cm
ACK D3
Parameter error D3!