70% to 100%:±2%(adult/pediatric, non-motion conditions)
70% to 100%:±3%(neonate, non-motion conditions)
70% to 100%:±3%(motion conditions)
Note : The Masimo SPO
2
sensor for adult/pediatric/ neonate has been validated for
motion accuracy in human blood studies on healthy adult volunteers in induced
hypoxia studies while performing rubbing and tapping motions at 2 to 4 Hz at an
amplitude of 1 to 2 cm and a non-repetitive motion between 1 to 5 Hz at an
amplitude of 2 to 3 cm in induced hypoxia studies in the range of 70-100%
SPO2
against a laboratory co-oximeter and ECG monitor. 1% has been added to the results
to account for the effect of fetal hemoglobin. This variation equals plus or minus one
standard deviation. Plus or minus one standard deviation encompasses 68% of the
population.
0% to 69%,unspecified