Warnings: A potential hazard can exist if different alarm pre-sets are used for the same or similar equipment
in any single area. If different settings are used, the operator may be confused by the different device pre-set
alarms in the area and thus could harm the patient.
6.13 Alarm History
Select【ALARM】on the main measurement display and then select the【HISTORY】tab to see all the alarm times,
alarm levels, alarm messages, and alarm duration times as the following picture shows:
Note:
• The record number of the alarm log depends on the storage space.
• The alarm system generates a technical alarm condition when the storage space is insufficient. When the
storage space is less than 10M, a low level technical alarm occurs and a prompt message will pop up with the
message: “Insufficient storage space". When the storage space is less than 5M, another low level technical
alarm occurs and a prompt message will pop up with the message: “Critical shortage of storage space".
• When the alarm system is powered down, the log is maintained, but the time that the alarm system was powered
down will not be captured in the log.