80-0078-00-MO-RevD 2018-08-02 SunTech CT50 User Manual | 49
6.3.3 Alarm Messages
The system uses different background colors to distinguish alarm level messages. The background color for different
alarm message levels is as follows:
High level alarms: red
Medium level alarms: yellow
Low level alarms: yellow (physiologic alarm)or cyan (technical alarm)
The number of * will indicate the alarm level in the message area as follows:
High level alarms: ***
Medium level alarms: **
Low level alarms: *
Caution: If several alarms occur, the highest-level alarm message will be displayed first. When two alarm
messages are the same alarm level, the latest alarm message will be displayed first. The display message
may be manually changed by the user to display the other alarm messages occurring.
6.4 Alarm Icons
The alarm is off.
The alarm is active.
The alarm sound is off.
The alarm is paused
6.5 Setting Alarm Volume
1. Select【Alarm】→【General】.
2. Select【Alarm Volume】and choose a desired value from 【Low】,【Medium】,【High】;
At the same time , you can select【SETTINGS】→【ADANCED】→【General】→【Alarm】to set the minimum alarm
volume to be 【Low】,【Medium】,【High】.
Warning:
• Auditory alarm signal sound pressure levels, when set to less than ambient levels, can impede operator
recognition of alarm conditions and the alarm system.
• When setting the alarm volume, the volume should be louder than the noise of the environment.