Pairing the Blood Pressure Monitor with your Smartphone
To begin using Bluetooth® for the rst time, please visit the web-
site at http://www.rossmax.com for the initial set-up instructions.
1. Download and install the applicable APP onto your smart-
phone.
2. To bind this device with your smartphone, turn on the device,
Bluetooth® and the App of smartphone, and follow set-up and
binding instructions.
3. If the binding is successful, the Bluetooth® symbol ( ) will
appear on the display and keep ashing during data transfer.
The current measured value will automatically be transferred
to the App when the measurement is completed.
4. If the binding has failed, the Bluetooth® symbol ( ) will not appear on the display and the current
measured value will not automatically be transferred to the App. In this case, the value is saved in
the selected user memory zone. Please re-bind this device with your smartphone and follow App
instructions for Bluetooth® transfer.
APP
APP
120mmHg
120mmHg
1. Adjust the date/time/Bluetooth in the monitor by holding down the ON/OFF/START key for ap-
proximately 5 seconds under power o mode. The display will show a blinking number show-
ing the year.
2. Change the year by pressing the Memory key, each press will increase the number. Press the ON/
OFF/START key to conrm the entry and the screen will show a blinking number representing
the date.
3. Change the date, the hour and the minute as described in Step 2 above, using the Memory key
to change and the ON/OFF/START key to conrm the entries.
4. After adjusting the date/time, the Bluetooth® symbol ( ) and the blinking icon “ ” will
be shown on the display simultaneously. Use the Memory key to choose whether automatic
Bluetooth® data transfer is activated (Bluetooth® symbol ( )+ ) or deactivated (Bluetooth®
symbol ( )+ ) and conrm with the ON/OFF/START key.
5. Press the ON/OFF/START key again, “0” will reappear as the Blood Pressure Monitor is ready for
measurement.
If any abnormality will arise during use, please check the following points.
Pr oblems Possible causes Solution
Nothing appears on the
display when operating the
unit.
Battery at. F ully charge the battery.
Malfunction. Press reset butt on on the back
side of the unit with a paper clip or
similar thin object.
LED indicator does not
illuminate when charging
the battery.
The AC Adapt er has not been cor-
rectly inserted into the por t of the
device or the power socket.
Correctly insert the AC Adapter into
the port/sock et.
The monitor or the AC
Adapter is abnormally hot
when charging the battery.
The monitor or the AC adapter
may be damaged.
Unplug the AC adapter from the
socket immediately and contact
your local distributor .
The unit does not work
from its battery.
Battery at. F ully charge the battery.
Battery ages. If the unit does not operate after
fully charging , replace the battery
with a new one.
Battery broken.
If the batter y cannot be fully
charged and the GREEN charg-
ing indicator keeps illuminating,
replace the battery with a new one.
EE mark shown on display
or the blood pressure value
is displayed excessiv ely low
(high).
Is the cu placed correctly? Wrap the cu properly so that it is
positioned correctly.
Did you talk or move during
measurement? Measure again. Keep arm steady
during measurement.
Did you vigorously shake the cu
during measurement.
Any abnormality occurs
during the measurement. Malfunction. Press reset butt on on the back
side of the unit with a paper clip or
similar thin object.
Note: If the device is still not working properly, return it to your dealer. Under no circumstance
should you disassemble and repair the unit by yourself.
1. The unit contains high-precision assemblies. Therefore, avoid extreme temperatures, hu-
midity, and direct sunlight. Avoid dropping or strongly shocking the main unit, and pro-
tect it from dust.
2. Clean the blood pressure monitor body and the cu carefully with a slightly damp, soft
cloth. Do not press. Do not wash the cu or use chemical cleaner on it. Never use thinner,
alcohol or petrol (gasoline) as cleaner.
3. The unit should not be operated by children so to avoid hazardous situations.
4. If the unit is stored near freezing, allow it to acclimate at room temperature before use.
5. This unit is not eld serviceable. You should not use any tool to open the device nor should
you attempt to adjust anything inside the device. If you have any problems, please contact
the store or the doctor from whom you purchased this unit or please contact Rossmax
International Ltd.
6. As a common issue for all blood pressure monitors using the oscillometric measurement
function, the device may have diculty in determining the proper blood pressure for us-
ers diagnosed with diabetes, poor circulation of blood, kidney problems, or for users suf-
fered from stroke, or for unconscious users.
7. This unit is able to detect common arrhythmia (atrial or ventricular premature beats or
atrial brillation). The ARR, AFib and PC icons are displayed after the measurement if Atrial
Fibrillation and Premature Contraction was detected during the measurement. If ARR,
AFib or PC icons are displayed, you are advised to wait for a while and take another meas-
urement. It is strongly recommended that you consult your physician if the ARR, AFib or
PC icons appear often.
8. While the given device is able to detect specic pulse arrhythmia, the measurement ac-
curacy of the blood pressure meter may be impaired with the occurrence of pulse ar-
rhythmia.
9. To stop operation at any time, press the ON/OFF/START key, and the air in the cu will be
rapidly exhausted.
10. Once the ination reaches 300 mmHg, the unit will start deating rapidly for safety rea-
sons.
11. Please note that this unit can be a home healthcare product, but it is not intended to
serve as a substitute for the advice of a physician or medical professional.
12. Do not use this device for diagnosis or treatment of any health problem or disease. Meas-
urement results are for reference only. Consult a healthcare professional for interpreta-
tion of pressure measurements. Contact your physician if you have or suspect any medi-
cal problem. Do not change your medications without the advice of your physician or
healthcare professional.
13. Electromagnetic interference: The device contains sensitive electronic components.
Avoid strong electrical or electromagnetic elds in the direct vicinity of the device (e.g.
mobile telephones, microwave ovens) or less than 1.5 km from AM, FM or TV broadcast
antennas. These may lead to temporary impairment of measurement accuracy.
14. Dispose of device, batteries, components and accessories according to local regulations.
15. This monitor may not meet its performance specication if stored or used outside tem-
perature and humidity ranges specied in Specications.
16. Please note that when inating, the functions of the limb in question may be impaired.
17. During the blood pressure measurement, blood circulation must not be stopped for an
unnecessarily long time. If the device malfunctions, remove the cu from the arm.
18. Avoid any mechanical restriction, compression or bending of the cu line.
19. Do not allow sustained pressure in the cu or frequent measurements. The resulting re-
striction of the blood ow may cause injury.
20. Ensure that the cu is not placed on an arm in which the arteries or veins are undergoing
medical treatment, e.g. intravascular access or therapy, or an arteriovenous (AV) shunt.
21. Do not apply the cu on the side, where a mastectomy has been performed in your
patient history.
22. Do not place the cu over wounds as this may cause further injury.
23. Only ever use the cus provided with the monitor or original replacement cus. Other-
wise erroneous results will be recorded.
24. Batteries can be fatal if swallowed. You should therefore store the batteries and products
where they are inaccessible to small children. If a battery has been swallowed, call a doc-
tor immediately.
25. Do not use the tubing and/or AC Adapter for any other purpose than those specied, as
they can cause risk of strangulation.
26. Do not service or maintain device and cu while in use.
27. This unit should not be used adjacent to or stacked with other equipment.
28. Please do not use any other cables or accessories not approved by the manufacturer in
this manual to avoid negative inuence on electromagnetic compatibility.
Measurement Method Oscillometric
Measurement Range Pressure: 30~260 mmHg; Pulse: 40~199 beats/
minute
Pressure Sensor Semi conductor
Accuracy Pressure: ± 3 mmHg; Pulse: ± 5% of reading
Inflation Pump Driven
Deflation Automatic Air Release Valve
Memory capacity 60 memories for each zone x 2 zones
Auto-shut-off 1 minute after last key operation
P ermissible Operating T emperature
and Humidity 10°C~40°C (50°F~104°F); 15%~85% RH; 700~1060
hPa
P ermissible T ranspor t and Storage
T emperature and Humidity -10°C~50°C (14°F~122°F); 10%~90% RH; 700~1060
hPa
Adapter Input 100-240V, 50/60 Hz
Adapter Output / USB Input 5V 1A (Type C)
Li-lon Battery DC 3.7V
Dimensions 120 (L) X 80 (W) X 57 (H) mm
W eight 276.0g (with battery, w/o cuff)
Arm circumference Adult: 24~40 cm (9.4”~15.7”)
Limited Users Adult users
Type BF: Device and cuff are designed to provide
special protection against electrical shocks.
IP Classification IP21: Protection against harmful ingress of water
and particulate matter
* Specifications are subject to change without notice.
T ime and Bluetooth® Adjustment
Data T ransf er via Bluetooth®
T roubleshooting
Cautionary Notes Specications
APP
APP
APP
120 mmHg
120 mmHg
USB PC
BP Manager
Electromagnetic Compatibility Information
1. This device needs to be installed and put into service in accordance with the information provided
in the user manual.
2. W ARNING: Portable RF communications equipment (including peripherals such as antenna cables
and external antennas) should be used no closer than 30 cm (12 inches) to any part of the Z5,
including cables specied by the manufacturer . O therwise, degradation of the per formance of this
device could result.
If higher IMMUNIT Y TEST LEVELS than those specied in T able 9 are used, the minimum separation
distance may be lower ed. Lo wer minimum separation distances shall be calculated using the
equation specied in 8.10.
Manufacturer ’ s declaration- electromagnetic immunity
The Z5 is intended for use in the electromag netic environment specied below . The customer or
the user of the Z5 should assure that is used in such and envir onment.
Immunity
test IEC 60601 test level Compliance level Electromagnetic envir onment-guidance
Conducted
RF IEC
61000-4-6
3 Vrms:
0,15 MHz – 80 MH z
6 Vrms: in ISM and
amateur radio bands
between
0,15 MHz and 80 MH z
80 % AM at 1 kHz
3 Vrms:
0,15 MHz – 80 MH z
6 Vrms: in ISM and
amateur radio bands
between
0,15 MHz and 80 MH z
80 % AM at 1 kHz
Portable and mobile RF communications
equipment should be used no closer to
any part of the Z5 including cables, than
the recommended separation distance
calculated from the equation applicable
to the frequency of the transmitter .
Recommended separation distance:
d = 1,2 √P , d = 1,2 √P 80MH z to 800 MHz,
d = 2,3 √P 800MHz to 2,7 GHz
Where P is the maximum output power
rating of the transmitter in watts ( W ) ac-
cording to the transmitter manufacturer
and d is the recommended separation
distance in metres (m).
Inter ference may occur in the vicinity of
equipment marked with the following
symbol:
Radiated RF
IEC 61000-
4-3
10 V/m
80 MHz – 2,7 GH z
80 % AM at 1 kHz
10 V/m
80 MHz – 2,7 GH z
80 % AM at 1 kHz
NOTE1: A t 80 MHz and 800 MH z, the higher frequency range applies.
NOTE2: These guidelines may not apply in all situations. E lectromagnetic propagation is aected
by absorption and reection from structures, objects and people.
More information on EMC compliance of the device can be obtained fr om Rossmax website: www.
rossmax.com.
Notes: 1. Unbinding your device will not delete the data from the App.
2. If you re-bind your smartphone with your blood pressure monitor, all prior reading history stored
on the App will be retained.
3. Bluetooth® data transfer will reduce the battery capacity.
1. Press the User-Switching key to select memory zone 1 or memory zone 2.
2. Press and hold the Memory key for approximately 5 seconds, then the data in the memory zone
can be erased automatically.
Clearing V alues from Memory
IN0Z50000000000XX
OBM_IB_Z5(1)NA19BT_
EN_SW_ver2023
W ARNING: The symbol on this product means that it's an electronic product and
following the Eur opean directive 2012/19/EU the electronic products have to be dis-
posed on your local recycling centre f or safe treatment.
Rossmax Swiss GmbH,
W idnauerstrasse 1, CH-9435 Heerbrugg , Switzerland
Model: Z5
w w w .rossmax.c om
W arranty Card
This instrument is cover ed by a 5 year guarantee fr om the date of purchase. The guar-
antee is valid only on presentation of the warranty card completed or stamped by the
seller/dealer confirming date of purchase or the receipt. Batt eries, cuff and accessories
are not included. Opening or alt ering the instrument invalidates the guarantee. The
guarantee does not cover damage , accidents or non-compliance with the instruc tion
manual. Please contact your local seller/dealer or www .rossmax.com.
Cust omer Name: ____________________________________________________
Address: ___________________________________________________________
T elephone: __________________________________________________________
E-mail address: ______________________________________________________
Product Information
Date of purchase: ___________________________________________________
Store wher e purchased:
____________________________________________________________________
Blood Pressure Monit or
- T he Bluetooth ® wor d mark and logos are r egistered tr ademarks owned by the Bluetooth SIG, Inc. and an y use of such marks by Rossmax Interna tional Ltd . is under
license. Other trademarks and tr ade names are those of their r espective owners.
- T he blood pressur e monitor uses Bluetooth ®(Bluetooth ® low ener gy technology)
- Apple and the Apple logo are tr ademarks of Apple Inc., r egister ed in the U.S. and other c ountries. App S tore is a service mark of Apple Inc.
- Google Play and the Google Play logo are tr ademarks of Google Inc.
Data T ransfer via Bluetooth®
Please download and install the
free Healthstyle APP onto your
smartphone
Healthstyle APP