• F inger is too large or t oo small to fit into the device .
• P oor pulse quality
• Venous pulsations
• Anemia or low hemoglobin concentrations.
• Cardiogr een and other intravascular dyes
• Carboxyhemoglobin
• Methemoglobin
• Dysfunc tional hemoglobin
• Ar tificial nails or fingernail polish
• On fingers with anatomical changes, oedemas, scars or burns.
- Using the device f or long periods may cause pain f or people
with circulator y disorders. Reposition the device at least once
ever y 4 hours to allow the patient’ s skin t o br eath and to check
patient ’ s condition r egularly .
- Do not use the device near flammable or explosive gas
mixtures.
- Do not use the device during an MRI or C T scan.
- The device may not work when circulation is reduced. W arm or
rub the finger , or r e -position the device.
- This device is a precision electronic instrument and must be
repaired by qualified t echnical professionals . F ield repair of the
device is not possible. Do no attempt to open the case or repair
the electronics. Opening the case ma y damage the device and
void the warranty .
- Do not overextend the device ’ s spring.
- A func tional tester cannot be used to access the accuracy of a
pulse oximeter monitor .
- Do not self- diagnose or self-medicate on the basis of the
measurements without consulting your doctor . I n par ticular , do
not start tak ing any new medication or change the t ype and/
or dosage of an y existing medication without prior approval .
- Do not look directly inside the housing during the
measurement. The red light and the invisible infra-red light in
the pulse oximet er are harmful t o your eyes .
- This device is not intended for use by people (including
children) with restric ted physical, sensor y or mental sk ills or a
lack of exper ience and/or a lack of k nowledge, unless they are
super vised by a person who has responsibility for their safety
or they r eceive instructions from this person on how to use the
device. Children should be super vised around the device to
ensure they do not play with it.
- Neither of the displa ys f or the pulse wave and pulse bar allows
the strength of the pulse or circulation to be evaluated at the
measurement site. R ather , they are exclusively used to display
the current signal variation at the measurement site and do
not enable r eliable diagnostics f or the pulse.
- The maximum temperature of sensors which the user will touch
might reach 43°C when operating in the 40°C envir onment.
Cleaning
T roubleshooting
Specification
1. Please clean the sur face of the device before using. Wipe
the device with medical alcohol (70% isopropyl alcohol)
first, and then let it dr y in air or clean it by dr y clean fabric.
When cleaning the device with water , the water temperature
should be lower than 60°C
2. Using the medical alcohol to disinfect the product af ter use,
prevent fr om cross infection f or next time use.
3. The best storage environment of the device is -25°C to 70°C
ambient temperature and not higher than 90% relative
humidity.
Symptoms Check points Corrections
SpO2 or pulse
rate cannot
displayed
Applied nger
improperly .
Place the nger
properly and tr y again.
SpO2 is too low to
detect
T r y again; go to
consult with your
physician if you ar e
sure the device works
well.
SpO2 or pulse
rate are not
displayed stably
Applied nger
improperly .
Place the nger
properly and tr y again.
F inger is shaking or
body is moving. keep body steady
No display
when button is
pressed
Batteries run down Replace with new
batteries
Batteries not
inser ted correctly . Re -inser t batteries
The display
disappears
suddenly
The device will auto
power o when it
gets no signal.
Normal
Low batt er y Replace with new
batteries
Note: I f the unit does not work, return it to your dealer .
Under no circumstance should you disassemble and
repair the unit by yourself .
SpO2
Measuring range 35%~99%, (the r esolution is 1%).
Accuracy 70%~99%: ±2%, Below 35~69%: unspecied .
Optical Sensor Red light (wavelength is 660nm), Infrared
(wavelength is 905/880nm)
Pulse
Measuring range 30bpm~250bpm (the r esolution is 1 bpm)
Accuracy ±3bpm
Po wer source AAA × 2 (Alkaline)
Battery life Continually f or 16 hours with two alkaline
batteries
Operating
Condition
T emperature: 5°C~40°C (41°F ~ 104°F),
Relative Humidity: 15-95% (non condensing),
Atmospheric pr essure: 700hPa ~ 1060hP a,
Attitude: -1,280 to 12,000 feet (-390m to
3,658m)
Storage /
T ranspor tation
Condition
T emperature: -25°C~+70°C (-13°F ~ 158°F),
Relative humidity: 15-90%(non condensing),
Atmospheric pr essure: 700hPa ~ 1060hP a,
Attitude: -1,280 to 12,000 feet (-390m to
3,658m)
Dimensions 63.5(L) × 34W ) × 35(H) mm
Weight About 37g (without the batt eries)
Standards IEC60601-1-2, Class B, IEC60601-1, Type BF ,
ISO80601-2-61, IEC60601-1-11
T ype BF applied par ts
IP Classication IP22: P rotection against harmful ingr ess of
water and par ticulate matter
Note: 1. Do not sterilize, autoclave or immerse this de vice in
liquid. Do not pour or spra y any liquids onto the de vice.
2. D o not use caustic or abrasive cleaning agents, or
any cleaning agent containing ammonium chloride or
isopropyl alcohol .
Maintenance
Recommends user to return this device to the manufacturer
per form the following checks ev er y 24 months.
- I nspect the equipment for mechanical and func tional damage
or deterioration.
- Ensure all user inter face keys and accessories func tion nor mally .
Model: SB210
w w w .rossmax.com
W ARNING: The symbol on this pr oduct means that it's an
electronic product and following the European directive
2012/19/EU the electronic products have to be dispose on
your local recycling centre f or safe treatment.
EMC guidance and manufacturer’s declaration
Recommended separation distances between portable and mobile RF communications equipment and the ME equipmen t
The Finger-tip pulse o ximeter is intended for use in an electromagnetic environmen t in which radiated RF disturbanc es are con-
trolled. The customer or the user of the F inger-tip pulse oximeter can help preven t electromagnetic interference by main taining
a minimum distance between portable and mobile RF communications equipment (transmitters) and the F inger-tip pulse
oximeter as r ecommended below, acc ording to the maximum output pow er of the communications equipment.
Rated maximum output
power of transmit-
ter / W
Separation distance acc ording to fr equenc y of transmitter / m
150 kHz to 80 MHz , d=[3.5/
V1]√P 80 MHz to 800 MHz , d=[3.5/
E1]√P 800 MHz to 2,5 GHz , d=[3.5/
E1]√P
0.01 0.12 0.12 0.23
0.1 0.37 0.37 0.74
1 1.17 1.17 2.33
10 3.7 3.7 7.37
100 11.67 11.67 23.33
Declaration – electromagnetic emissions and immunity – for EQUIPMENT and SYSTEMS that ar e not LIFE-SUPPORTING and
are specied for use only in a shielded loca tion
The Finger-tip pulse o ximeter declaration – electromagnetic immunity
The Finger-tip pulse o ximeter system is intended for use in the electromagnetic envir onment specied below . T he customer or
the user of the Finger-tip pulse oximet er system should assure that it is used in such an en vironment .
Immunity test IEC 60601 test level Compliance level Electromagnetic environment-guidance
Conducted RF
IEC 61000-4-6 3 V rms
150 kHz to 80 MHz 3V Portable and mobile RF communications equipmen t should be used
no closer to any part of the EQUIPMENT or SYSTEM including cables,
than the recommended separ ation distance calculat ed from the
equation applicable to the fr equency of the transmitter . Interference
may occur in the vicinity of equipment marked with the follo wing
symbol.
Radiated RF
IEC 61000-4-3 3 V/m
80 MHz to 2.5 GHz 3V/m
Declaration – electromagnetic immunity
The Finger-tip pulse o ximeter system is intended for use in the electromagnetic envir onment specied below . T he customer or
the user of the Finger-tip pulse oximet er system should assure that it is used in such an en vironment .
Immunity test IEC 60601 test level Compliance lev el Electromagnetic environment-guidance
Electrostatic
discharge (ESD) IEC
61000-4-2
±6 kV contact
±8 kV air ±6 kV contact
±8 kV air Floors should be wood , concret e or ceramic
tile. If oors are c overed with syn thetic
material , the relative humidity should be
at least 30 %.
Electrical fast
transient/burst IEC
61000-4-4
±2 kV for power sup-
ply lines
±1 kV for input/output
lines
±2 kV for power sup-
ply lines Mains power quality should be that of a
typical commercial or hospital envir onment.
Surge IEC 61000-4-5 ± 1kV dierential mode
± 2kV common mode ± 1kV dierential mode
± 2kV common mode Mains power quality should be that of a
typical commercial or hospital envir onment.
V oltage dips, shor t
interruptions and
voltage variations on
power supply input
lines IEC 61000-4-11
<5% UT(>95% dip in UT)
for 0,5 cycle
40% UT(60% dip in UT)
for 5 cycles
70% UT(30% dip in UT)
for 25 cycles
<5% UT(>95% dip in
UT) for 5 s
<5% UT(>95% dip in UT)
for 0,5 cycle
40% UT(60% dip in UT)
for 5 cycles
70% UT(30% dip in UT)
for 25 cycles
<5% UT(>95% dip in
UT) for 5 s
Mains power quality should be that of a
typical commercial or hospital envir onment.
If the user of the EQUIPMENT or SYSTEM
requir es continued operation during po wer
mains interruptions , it is recommended tha t
the EQUIPMENT or SYSTEM be power ed
from an unint erruptible power supply or
a battery .
Pow er frequency
(50/60 Hz) magnetic
eld IEC 61000-4-8
3 A/m 3 A/m Pow er frequency magnetic elds should
be at levels characteristic of a typical loca-
tion in a typical commercial or hospital
environmen t.
Declaration – electromagnetic emissions
The Finger-tip pulse o ximeter is intended for use in the electromagnetic environmen t specied below . T he customer or the user
of the Finger-tip pulse oximet er should assure that it is used in such an envir onment.
Emission test Compliance Electromagnetic envir onment- guidance
CE emissions CISPR11 Group 1 The Finger-tip pulse o ximeter uses RF energy only for its internal func-
tion. Therefore , its RF emissions are very low and are not lik ely to cause
any interference in nearb y electronic equipment.
RF emissions CISPR 11 Class B T he Finger-tip pulse oximeter is suitable for use in all establishments,
including domestic establishments and those directly connected t o the
public low-voltage pow er supply network that supplies buildings used
for domestic purposes.
Harmonic emissions IEC 61000-3-2 Class A
V oltage uc tuations/icker emissions
IEC 61000-3-3 Complies
Rossmax Inno T ek Corp .
12F ., No. 189, Kang C hien Rd., T aipei, 114, T aiwan.
CMC Medical Devices & Drugs S.L.
C/ Horacio L engo N° 18, CP 29006, Málaga, Spain
Fingertip Pulse Oximeter
- The Bluetooth ® word mark and logos are r egistered trademarks owned by the Bluet ooth SIG, Inc. and any use of
such marks by Rossmax Interna tional Ltd . is under license. Other trademarks and trade names ar e those of their
respective owners .
- The Pulse Oximet er uses Bluetooth ®(Bluetooth ® low energy t echnology)
- Apple and the Apple logo ar e trademarks of Apple Inc., r egistered in the U .S. and other countries. App St ore is a
service mark of Apple Inc.
- Google Play and the Google Play logo ar e trademarks of Google Inc.
IN0SB210000000021
RI_IB_SB210_EN_2Yr_
TP_ver2003
W arranty Card
This instrument is covered by a 2 year guarantee from the date of
purchase, batter ies and accessories are not included. The guarantee
is valid only on presentation of the guarantee card completed by
the dealer confirming date of purchase or the receipt. Opening or
altering the instrument invalidates the guarantee. The guarantee
does not cover damage , accidents or non- compliance with the
instruction manual. Please contact your local seller/dealer or ww w .
rossmax.com.
Customer Name: ____________________________________
Address: ___________________________________________
T elephone: _________________________________________