SpO2
Measuring range 35% – 100% (the resolution is 1%)
Accuracy 70% – 100% (No Motion. ±2%, Motion. ±3%, Low Perfusion. ±2%)
Pulse Rate
Measuring range 30 – 250 bpm (the resolution is 1 bpm)
Accuracy 30 – 250 ± 3 digits
Probe Type
Probe model Rossmax PA100, PB100, PC100, PD100(Single Use), PF100
Extension cord Rossmax PE100
Battery AA * 4 (Alkaline)
Battery Life Continually for 15 hours with 4 alkaline batteries
AC Adaptor Model: HK-X205-A06, HK-XW05-A06, (W=1,2,3,4), HKKS-13116, HKKS-13117
Input: AC100-240V, 50/60Hz, 0.2A max; Output: DC 6V, 0.8A
Environmental conditions
Operation Condition Temperature: 5°C – 40°C (41°F – 104°F), Relative Humidity:15% – 95%(non
condensing), Atmospheric pressure: 700hPa ~ 1013hPa
Storage /Transport
Condition Temperature: -20°C – 70°C (-4°F – 158°F), Relative Humidity:15% – 95%(non
condensing), Atmospheric pressure: 700hPa ~ 1013hPa
Note: The condition of -20°C or 70°C back to use should stand for 3 hours at
room temperature.
Dimension Size: 14.5(L) x 7.25(W) x 2.25cm(H)
Weight About 150g (without the batteries)
Standard IEC/EN60601-1, IEC/EN60601-1-2, IEC/EN60601-1-11, ISO80601-2-61
Symbol Descriptors
Manufacturer
Serial number
EU representative
Type BF (Body Floating
IP22: Protected against foreign objects and moisture
CE Mark
Warning: the symbol on this product means that its an electronic product
and following the European directive 2012/19/EU the electronic products
have to be dispose on your local recycling centre for safe treatment.
Contains of natural rubber latex
Symptoms Check points Corrections
SpO2 or Pulse rate
cannot displayed
The icon “- -“ shows on the
screen try again.
ACT PPG
This icon means probe
connect failure.
Be sure ”Rossmax probe is
connected to the device
correctly.
ACT PPG
This icon means probe
dysfunction Replace with new probe.
try again
SpO2 or Pulse rate are
not displayed stably
Finger is shaking or body is moving Keep body steady
try again
No display when the
bottom is pressed
Batteries run down Replace with new batteries
Batteries inserted incorrectly Re-insert batteries
The display disappears
suddenly gets no signal Normal
Low battery Replace with new batteries
Note: If the unit does not work, return it to your dealer. Under no circumstance should you
disassemble and repair the unit by yourself.
Note: After clearing data, the icon will show 0 data.
Caution: The deleted data could not be restored.
Warning
Cleaning
Memories mode- Save Cycle
Memories mode- Clear Data
Specication
Troubleshooting
Audio Signals
This device is not intended for use by people (including children) with restricted physical, sensory or
mental skills or a lack of experience and/or a lack of knowledge, unless they are supervised by a per-
son who has responsibility for their safety or they receive instructions from this person on how to
use the device. Children should be supervised around the device to ensure they do not play with it.
This device only for spot-checking, but not medical result evaluation.
This device is designed to determine the percentage of arterial oxygen saturation of functional
hemoglobin. Factors that may degrade pulse oximeter performance or affect the accuracy of the
measurement include the following:
- Do not apply the pulse oximeter on the same arm as a blood pressure cuff, arterial catheter or
infusion line (s)
- Excessive light, such as sunlight or direct home lighting.
- Not steady at the site of application (e.g. term-bling)
- Moisture in the device
- Improperly applied device
- Finger is too large or too small to fit into the device
- Poor pulse quality
- Venous pulsations
- Anemia or low hemoglobin concentrations
- Cardio green and other intravascular dyes
- Carboxyhemoglobin
- Methemoglobin
- Dysfunctional hemoglobin
- Artificial nail or fingernail polish
- On fingers with anatomical changes, oedemas, scars or burns.
- The conditional of probe. Use only the Rossmax approved pulse oximeter sensor, cable and
accessories. Use of other sensors, cable and accessories can result in inaccurate readings.
Using the device for long periods may cause pain for people with circulatory disorders. Reposition
the device (probe) at least once every 4 hours to allow the patient’s skin to breath and to check
patient’s condition regularly.
Do not use the device near flammable or explosive gas mixtures.
Do not use the device during an MRI or CT scan, be used no closer than 30 cm (12 inches) to any part
of the [ME EQUIPMENT or ME SYSTEM], including cables specified by the manufacturer.
The device will be affected by electromagnetic interference during operation.
A warning that other cables and accessories may negatively affect EMC performance.
The device may not work when circulation is reduced. Warm or rub the nger, or re-position the
device.
This device is a precision electronic instrument and must be repaired by qualified technical
professionals. Field repair of the device is not possible. Do not attempt to open the case or repair
the electronics. Opening the case may damage the device and void the warranty.
Do not overextend the device’s spring.
A functional tester cannot be used to access the accuracy of a pulse oximeter monitor.
Do not self-diagonse or self-medicate on the basis of the measurements without consulting your
doctor. In particular, do not start taking any new medication or change the type and/or dosage of
any existing medication without prior approval.
Do not look directly inside the housing during the measurement. The red light and the invisible
infra-red light in the probe are harmful to your eyes.
Please be aware that user with susceptible skin.
As with all medical equipment, carefully route patient cabling to reduce the possibility of patient
entanglement or strangulation.
1. Please clean the surface of the device before using. Wipe the device with medical alcohol (70%
isopropyl alcohol) first, and then let it dry in air or clean it by dry clean fabric.
2. Using the medical alcohol to clean the product after use, prevent from cross infection for next
time use.
3. The best storage environment of the device is -20°C ~ 70°C ambient temperature and not higher
than 95% relative humidity.
Note: 1. Do not sterilize, autoclave or immerse this device in liquid. Do not pour or spray any liquids
onto the device.
2. Do not use caustic or abrasive cleaning agents, or any cleaning agent containing ammonium
chloride or isopropyl alcohol.
Optional of memory interval: 60 / 30 / 10 seconds
Press the button and / button to highlight “Memory” option, and then press the
button to select the option.
Entry “Memory Mode”, press the / button to highlight Save Cycle” option, and then press
the button to select the option.
Entry “Save Cycle”, press the / button to highlight option, and then press the button
to select the option.
Press the / button to change the value; press the button to save the desired value.
Press the / button and the button to select “Back/Exit” to return.
Press the button and / button to highlight “Clear Data” option, and then press the
button to select the option.
Entry “Clear Data”, press the / button to highlight “Sure” option, and then press the
button to select the option.
Press the / button and the button to select “Back/Exit” to return.
No. Name LCD Display Sound Control
1 Power on
V0.00 à
Beep for 2 seconds Press button
Software
version àstart measure
2Pulse
search The icon X X
3
Max./ Min
SpO2 and
Pulse rate
warning
Show or in red icon.
limits.
Beep-beep sounded
repeatedly Default
Mute for temporary, will
Beep-beep sounded
after two minutes. mode
mode
Bar Chart
Entry “Data History, press the / button to highlight “Bar Chartoption for viewing statistics
of SpO2, and then press the button to select the option.
Entry “SpO2 Memory Data”, press the button to go back to the monitoring screen.
Memory Transfer to ROSSMAX App
Entry into “Data History, press the / button to highlight “Memory Transferoption, and
then press the button to start data transmission for a few seconds.
Note: Before entering the data history mode, the icon should collect more than 1 data.
4Probe
failure
alarm
The icon shows
on the screen Beep-beep sounded
repeatedly X
5
Probe
connect
failure
alarm
The icon shows
on the screen Beep-beep sounded
repeatedly x
6Low
battery
alarm
The icon shows
on the screen
Beep-beep sounded
repeatedly for 1 minute x
7Automatic XBeep-beep sounded After pulse is unde-
tectable for around
1 minute.
8Unable
measure
Blood saturation
& pulse rate
appears “ - -
Beep-beep sounded
repeatedly. x
9 AC Power icon X X
Model: SA310
www.rossmax.com
Handheld Pulse Oximeter
60/30/10
seconds
Electromagnetic Compatibility Information
1. This device needs to be installed and put into service in accordance with the information provided in the user
manual.
2. WARNING: Portable RF communications equipment (including peripherals such as antenna cables and external
by the manufacturer. Otherwise, degradation of the performance of this device could result.
Manufacturers declaration-electromagnetic immunity
of the SA310 should assure that is used in such and environment.
Immunity test IEC 60601 test level Compliance level Electromagnetic environment-guidance
Conducted RF
IEC 61000-4-6
3 Vrms:
0,15 MHz – 80 MHz
6 Vrms: in ISM and
amateur radio bands
between
0,15 MHz and 80 MHz
80 % AM at 1 kHz
3 Vrms:
0,15 MHz – 80 MHz
6 Vrms: in ISM and
amateur radio bands
between
0,15 MHz and 80 MHz
80 % AM at 1 kHz
Portable and mobile RF communications
equipment should be used no closer to any
part of the SA310 including cables, than the
recommended separation distance calculated
from the equation applicable to the frequency
of the transmitter.
Recommended separation distance:
d = 1,2 √P, d = 1,2 √P 80MHz to 800 MHz, d = 2,3
√P 800MHz to 2,7 GHz
Where P is the maximum output power rating
of the transmitter in watts (W) according to the
transmitter manufacturer and d is the recom-
mended separation distance in metres (m).
Interference may occur in the vicinity of equip-
ment marked with the following symbol:
Radiated RF
IEC 61000-4-3
10 V/m
80 MHz – 2,7 GHz
80 % AM at 1 kHz
10 V/m
80 MHz – 2,7 GHz
80 % AM at 1 kHz
NOTE1: At 80 MHz and 800 MHz, the higher frequency range applies.
More information on EMC compliance of the device may be obtained from Rossmax using the contacts shown
in this manual.
IN0SA311000000030
RI_IB_SA310_EN_2Yr_
TP_ver1838
Warranty Card
This instrument is covered by a 2 year guarantee from the date of purchase, batteries
and accessories are not included. The guarantee is valid only on presentation of the
guarantee card completed by the dealer confirming date of purchase or the receipt.
Opening or altering the instrument invalidates the guarantee. The guarantee does
not cover damage, accidents or non-compliance with the instruction manual. Please
contact your local seller/dealer or www.rossmax.com.
Customer Name: __________________________________________________
Address: _________________________________________________________
Telephone: _______________________________________________________
E-mail address: ____________________________________________________
Product Information:
Date of purchase: __________________________________________________
Store where purchased:
__________________________________________________________________
Rossmax InnoTek Corp.
12F., No. 189, Kang Chien Rd., Taipei, 114, Taiwan.
CMC Medical Devices & Drugs S.L.
C/ Horacio Lengo N° 18, CP 29006, Málaga, Spain
1

Hulp nodig? Stel uw vraag in het forum

Spelregels

Misbruik melden

Gebruikershandleiding.com neemt misbruik van zijn services uitermate serieus. U kunt hieronder aangeven waarom deze vraag ongepast is. Wij controleren de vraag en zonodig wordt deze verwijderd.

Product:

Bijvoorbeeld antisemitische inhoud, racistische inhoud, of materiaal dat gewelddadige fysieke handelingen tot gevolg kan hebben.

Bijvoorbeeld een creditcardnummer, een persoonlijk identificatienummer, of een geheim adres. E-mailadressen en volledige namen worden niet als privégegevens beschouwd.

Spelregels forum

Om tot zinvolle vragen te komen hanteren wij de volgende spelregels:

Belangrijk! Als er een antwoord wordt gegeven op uw vraag, dan is het voor de gever van het antwoord nuttig om te weten als u er wel (of niet) mee geholpen bent! Wij vragen u dus ook te reageren op een antwoord.

Belangrijk! Antwoorden worden ook per e-mail naar abonnees gestuurd. Laat uw emailadres achter op deze site, zodat u op de hoogte blijft. U krijgt dan ook andere vragen en antwoorden te zien.

Abonneren

Abonneer u voor het ontvangen van emails voor uw Rossmax SA310 bij:


U ontvangt een email met instructies om u voor één of beide opties in te schrijven.


Ontvang uw handleiding per email

Vul uw emailadres in en ontvang de handleiding van Rossmax SA310 in de taal/talen: Engels als bijlage per email.

De handleiding is 3.96 mb groot.

 

U ontvangt de handleiding per email binnen enkele minuten. Als u geen email heeft ontvangen, dan heeft u waarschijnlijk een verkeerd emailadres ingevuld of is uw mailbox te vol. Daarnaast kan het zijn dat uw internetprovider een maximum heeft aan de grootte per email. Omdat hier een handleiding wordt meegestuurd, kan het voorkomen dat de email groter is dan toegestaan bij uw provider.

Stel vragen via chat aan uw handleiding

Stel uw vraag over deze PDF

loading

Uw handleiding is per email verstuurd. Controleer uw email

Als u niet binnen een kwartier uw email met handleiding ontvangen heeft, kan het zijn dat u een verkeerd emailadres heeft ingevuld of dat uw emailprovider een maximum grootte per email heeft ingesteld die kleiner is dan de grootte van de handleiding.

Er is een email naar u verstuurd om uw inschrijving definitief te maken.

Controleer uw email en volg de aanwijzingen op om uw inschrijving definitief te maken

U heeft geen emailadres opgegeven

Als u de handleiding per email wilt ontvangen, vul dan een geldig emailadres in.

Uw vraag is op deze pagina toegevoegd

Wilt u een email ontvangen bij een antwoord en/of nieuwe vragen? Vul dan hier uw emailadres in.



Info