Instruction Manual
Model: CalPro
Mobile Calibration System
for Blood Pressure Monitor
2
Table of Contents:
Introduction ...................................................................................... 3
Preliminary Remarks ........................................................................... 3
Cautionary Notes ............................................................................... 3
Name/Function of Each Part ............................................................ 4,5
LCD Display Explanations ................................................................... 6
Using the AC Adapter ........................................................................ 7
Connect Methods ........................................................................... 8,9
Operation Procedures
1. Data link status operation ......................................................... 10
2. Readout Mode .......................................................................... 11
3. Auto Calibration Mode ............................................................. 19
4. Measurement Count Display Mode ........................................... 21
5. Measurement Count Clearance Mode ....................................... 22
6. Calibration Reminder Setting Mode .......................................... 23
7. Pressure Gauge Mode .......................................................... 24-26
Change "mmHg" to "kPa" ............................................................. 27
Error Messages ................................................................................ 28
Specications ................................................................................... 29
Safety Certications ......................................................................... 30
USA FCC Class B ............................................................................. 30
Maintenance of ROSSMAX Calibrator .............................................. 30
3
Introduction
The calibration system developed by Rossmax Reasearch and Develop-
ment is equipped with the following features :
Automatic device pressure checking
Automatic device pressure calibration
Measurement count check and reset
Digital manometer
Which helps our partners to offer on-site and prompt calibration service
to end users as well as signicantly enhance customer satisfaction!
Preliminary Remarks
The quality of the device has been veried and conforms to the provi-
sions of the Electromagnetic Compatibility EMC Directive 2014/30/EU;
Low Voltage Directive (LVD) 2014/35/EU and EN/IEC 61010-1 :2010 Safe-
ty requirements for electrical equipment for measurement, control, and
laboratory use - Part 1: General requirements.
Cautionary Notes
1. The unit contains high-precision assemblies. Therefore, avoid extreme
temperatures, humidity, and direct sunlight. Avoid dropping or strong-
ly shocking the main unit, and protect it from dust.
2. The unit should not be operated by children so to avoid hazardous
situations.
3. If the unit is stored near freezing, allow it to acclimate at room tem-
perature before use.
4. This unit is not eld serviceable. You should not use any tool to open
the device nor should you attempt to adjust anything inside the de-
vice. If you have any problems, please contact the store or the doctor
from whom you purchased this unit or please contact Rossmax Inter-
national Ltd.
5. Dispose of device, batteries, components and accessories according to
local regulations.
6. Data link can only be connected with Rossmax blood pressure monitor
and calibrator. It cannot be used for any other purpose.
7. Clean the device carefully with a slightly damp, soft cloth.
4
Name/Function of Each Part
kPa mmHg
ReadoutAuto calibrationPressure gauge
Check
count
Clear
count
Set
count
LCD Display
Number
Entry key
Clear key
ON/OFF key
Readout key
Auto Calibration
key
Zero key
Deate key
Inate key
Check Count key
Clear Count key
Set Count key
Air tube socket
Data Link Port
AC Adapter Jack
5
Data Link CableAir Tube and Connector
for Wrist Monitors
Air Tube and Connector
for Upper Arm Monitors
Carrying Bag
AC Adapter
Output: 600mA
(Ø3.8/Ø1.3)
Name/Function of Each Part
Air Outlet
Data Link
Socket
Data Link
Socket
Air Outlet
Data Link
Socket
Air Outlet
6
1
2
3
5
6
4
kPa mmHg
ReadoutAuto calibrationPressure gauge
Check
count
Clear
count
Set
count
LCD Display Explanations
1
2
3
5
6
4
Readout Mode: Automatic device pressure checking
1
2
3
5
6
4
Auto Calibration Mode: Automatic device pressure calibration
1
2
3
5
6
4
Pressure Gauge Mode: Digital Manometer feature
1
2
3
5
6
4
Measurement Count Display Mode: Check number of uses
1
2
3
5
6
4
Measurement Count Clearance Mode: Zero setting of uses
1
2
3
5
6
4
Calibration Reminder Setting Mode: Re-set number of use cali-
bration reminder
7
Using the AC Adapter
1. Connect the AC adapter with the AC adapter jack in the back of the
calibrator.
2. Plug the AC adapter into the socket. Use the compatible AC adapters.
(AC adapters with required voltage and current indicated near the AC
adapter jack.)
Do not plug or unplug the AC adapter into the electrical
outlet with wet hands.
Unplug the AC adapter from the electrical outlet after using
the device.
8
Connection Methods
Upper Arm Blood Pressure Monitor:
1. Connect the tube between the calibrator and the monitor to the air
tube socket.
2. Then connect the data link cable between the calibrator and the moni-
tor to the data link port.
Data Link Cable
Air Tube Date Link Port
Air Tube
Socket
Data Link Cable
Air Tube
Date Link Port
Air Tube
Socket
9
Cuff Screw
Connection Methods
Wrist Blood Pressure Monitor:
1. Loosen the screw that secures the cuff to the monitor.
2. Remove the cuff from the monitor.
a. Push cuff to the screw direction.
b. Slightly pull cuff to remove the cuff from the monitor.
3. Connect the tube between the calibrator and the monitor to the air
outlet.
4. Connect the data link cable between the calibrator and the monitors
data link port.
Data Link Cable
Air Tube
Date Link Port
Air Tube Socket
Wrist Cuff
Wrist Cuff
10
kPa mmHg
ReadoutAuto calibrationPressure gauge
Check
count
Clear
count
Set
count
Operation Procedures
1. Data link status operation
Before operation, make sure the air tube and the data link cable are
securely connected to the calibrator and the blood pressure monitor.
Press the ON/OFF key of the calibrator, then press the M key of the
blood pressure monitor.
Note: If there is no M key on the blood pressure monitor, please press
the ON/OFF key instead.
A loop will appear on the LCD display of the Blood Pressure Monitor,
indicating the data link status.
Calibrator Blood Pressure Monitor
Calibrator Screen Blood Pressure
Monitor Screen
11
2. Readout Mode
This feature checks monitor pressure, if you would like to ignore
pressure checking , skip to "Auto Calibration Mode"(page 18) directly.
Press “Readout” to start. “RUN” will appear on LCD and pressure is
checked at 0/295/270/240 mmHg points. If the pressure difference
at 0/295/270/240 mmHg is within ± 3mmHg, the calibrator will rapid
exhaust to 0 mmHg, then go on checking 2nd points. If "FAIL" appears
, the faulty pressure points will be shown (both the default pressure
check point and the actual pressure measured will appear alternatively.).
If "FAIL" appears, press the "Auto calibration" key for re-calibration.
After this process is complete, press the "Readout" key to verify
calibration was successful.
Calibrator
Screen
Blood
Pressure
Monitor
Screen
Check First Pressure Points (295/270/240 mmHg)
Link Code
(Reference Only)
30
Safety Certifications
Rossmax’s products comply with the standard of LVD and EMC
Directire and affix the CE marking, could be placed in the market or
put into service within the internal market of the European Union (EU).
The CE mark certifies that products have met EU health, safety and
environmental requirements, which ensure consumer and workplace
safety.
To ensure continued accuracy, it is recommended that the monitor be
re-calibrated every two years. If you prefer, please send the instruments
to the certified calibration laboratory to test and verify the accuracy.
Maintenance of ROSSMAX Calibrator
We highly recommend that ROSSMAX CALIBRATOR be calibrated every
two years by your local Metrology Center.
The calibrator at the Metrology Center used to calibrate our calibrator
must have accuracy deviation of no more than +/-0.2 mmHg.
USA FCC Class B
This device complies with part 15 of the FCC Rules. Operation is subject
to the following two conditions:
1. This device may not cause harmful interference, and
2. This device must accept any interference received, including interference
that may cause undesired operation.
CLASS II equipment
The symbol on this product means that it's an electronic
product and following the European directive 2012/19/EU the
electronic products have to be disposed on your local recycling
centre for safe treatment.
Alternating current
Direct current
30

Hulp nodig? Stel uw vraag in het forum

Spelregels

Misbruik melden

Gebruikershandleiding.com neemt misbruik van zijn services uitermate serieus. U kunt hieronder aangeven waarom deze vraag ongepast is. Wij controleren de vraag en zonodig wordt deze verwijderd.

Product:

Bijvoorbeeld antisemitische inhoud, racistische inhoud, of materiaal dat gewelddadige fysieke handelingen tot gevolg kan hebben.

Bijvoorbeeld een creditcardnummer, een persoonlijk identificatienummer, of een geheim adres. E-mailadressen en volledige namen worden niet als privégegevens beschouwd.

Spelregels forum

Om tot zinvolle vragen te komen hanteren wij de volgende spelregels:

Belangrijk! Als er een antwoord wordt gegeven op uw vraag, dan is het voor de gever van het antwoord nuttig om te weten als u er wel (of niet) mee geholpen bent! Wij vragen u dus ook te reageren op een antwoord.

Belangrijk! Antwoorden worden ook per e-mail naar abonnees gestuurd. Laat uw emailadres achter op deze site, zodat u op de hoogte blijft. U krijgt dan ook andere vragen en antwoorden te zien.

Abonneren

Abonneer u voor het ontvangen van emails voor uw Rossmax CalPro - Mobile Calibration System bij:


U ontvangt een email met instructies om u voor één of beide opties in te schrijven.


Ontvang uw handleiding per email

Vul uw emailadres in en ontvang de handleiding van Rossmax CalPro - Mobile Calibration System in de taal/talen: Engels als bijlage per email.

De handleiding is 6.27 mb groot.

 

U ontvangt de handleiding per email binnen enkele minuten. Als u geen email heeft ontvangen, dan heeft u waarschijnlijk een verkeerd emailadres ingevuld of is uw mailbox te vol. Daarnaast kan het zijn dat uw internetprovider een maximum heeft aan de grootte per email. Omdat hier een handleiding wordt meegestuurd, kan het voorkomen dat de email groter is dan toegestaan bij uw provider.

Stel vragen via chat aan uw handleiding

Stel uw vraag over deze PDF

Uw handleiding is per email verstuurd. Controleer uw email

Als u niet binnen een kwartier uw email met handleiding ontvangen heeft, kan het zijn dat u een verkeerd emailadres heeft ingevuld of dat uw emailprovider een maximum grootte per email heeft ingesteld die kleiner is dan de grootte van de handleiding.

Er is een email naar u verstuurd om uw inschrijving definitief te maken.

Controleer uw email en volg de aanwijzingen op om uw inschrijving definitief te maken

U heeft geen emailadres opgegeven

Als u de handleiding per email wilt ontvangen, vul dan een geldig emailadres in.

Uw vraag is op deze pagina toegevoegd

Wilt u een email ontvangen bij een antwoord en/of nieuwe vragen? Vul dan hier uw emailadres in.



Info