Model number: RLV
Users Manual
Neuromuscular Electrical Stimulation Device
PLEASE READ THE
USERS MANUAL CAREFULLY
BEFORE USING THIS PRODUCT
USE ONLY AS DIRECTED
If symptoms persist, consult
your healthcare professional
2452_IFU06_18063164.indd 12452_IFU06_18063164.indd 1 19/05/2021 10:2219/05/2021 10:22
23
What is inside the box? 4
Parts and Controls 5
Introduction to Revitive 6
Indications for use 6
How Revitive works 6
Important Safeguards 7-15
Contraindications 7
Warnings and Cautions 8-13
Safety Precautions 14
Adverse Reactions 15
Instructions for use 16-25
How long should I use Revitive? 16
Step 1: Setting up Revitive for the first time 16
Step 2: Using the foot pads 17-19
Targeting other areas of the body 20-24
Replacing the Electrode Body Pads 25
Cleaning and storing 25
Troubleshooting 26-27
Technical Specifications 28-29
Symbols 29-30
Warranty 32
Table of contents
FOR BEST RESULTS
FOR 60 DAYS
INTENSITY +
30
30-60 MIN
USE DAILY
R
R
2452_IFU06_18063164.indd 2-3
2452_IFU06_18063164.indd 2-3
19/05/2021 10:22
19/05/2021 10:22
45
What is inside the box?
Upon opening the carton, please check that the following components are provided.
If you think anything is missing, please contact us using the helpline numbers on the
back of this booklet.
Control panel/display
Power ON/OFF
Foot-pads
Power adapter socket
Body pad socket
Intensity control – /+
Revitive Essential
Controls of Revitive Essential
Control panel
Parts and Controls
Parts of Revitive Essential
Time remaining
Intensity Level
1AC/DC power adapter
34
Electrode Body Pads
Cable length 180cm
Electrode pad cord
Cable length 150cm
2
2452_IFU06_18063164.indd 4-5
2452_IFU06_18063164.indd 4-5
19/05/2021 10:22
19/05/2021 10:22
67
Introduction to Revitive
Intended use:
Revitive is intended to be used for electrical stimulation of the muscles of the lower
leg for the purposes of supporting muscle function, and supporting and improving
blood circulation. In addition Revitive is intended to be used for electrical stimulation
of sore and aching muscles to relieve pain. Revitive is intended for home use for
30 minutes per treatment with a maximum of 3 hours treatment time per day as
required.
Indications for use
To temporarily increase local blood circulation in healthy leg muscles.
To stimulate healthy muscles in order to improve and facilitate
muscle performance
For temporary relief of pain associated with sore and aching muscles in
the shoulder, waist, back, upper extremities (arms) and lower extremities
(legs) due to strain from exercise or normal household duties
How Revitive works
Revitive delivers electrical stimulation to your feet via a pair of cushioned foot pads,
while you are seated. The electrical stimulation causes muscles in your calf and foot
to contract, which helps to increase circulation in your lower legs.
Revitive has been developed so that it is comfortable for you to use.
Revitive can also deliver electrical stimulation to the knees, shoulders, back or arms,
by using the Revitive Electrode Body Pads (sold separately).
Visit www.revitive.com for more information
Important Safeguards
Read all instructions before use
Who should not use Revitive
Revitive should not be used by some people.
Contraindications, do not use if:
You have a cardiac pacemaker, implanted defibrillator, or other
implanted electronic device because this may cause electric shock,
burns, electrical interference, or death
You are pregnant
You are being treated for, or have the symptoms of, an existing
Deep Vein Thrombosis (“DVT”)
Long periods of inactivity can put you at greater risk of developing Deep
Vein Thrombosis (DVT). DVT is a blood clot and usually occurs in a deep
leg vein. If part of the DVT breaks o it may lead to potentially life-
threatening complications such as pulmonary embolism.
If you have been inactive for prolonged periods and suspect you may
have a DVT, consult your doctor immediately. To prevent dislodging the
clot do not use Revitive.
In some cases of DVT there may be no symptoms. However it is important
to be aware of the symptoms that may include:
pain, swelling and tenderness in one of your legs (usually your calf)
a heavy ache in the aected area
warm skin in the area of the clot
redness of your skin, particularly at the back of your leg, below
the knee
Consult your doctor as soon as possible if you show any signs of the
above symptoms.
2452_IFU06_18063164.indd 7
2452_IFU06_18063164.indd 7
19/05/2021 10:22
19/05/2021 10:22
89
Important Safeguards
Warnings
Revitive is not intended to manage or reverse any underlying condition or provide
long-term results, it is intended to provide relief of symptoms
Consult with your doctor BEFORE using this device if:
you are in the care of a doctor
you have a history of heart problems because the device may cause lethal rhythm
disturbances to the heart in susceptible individuals
you are experiencing cramping leg pain that occurs during walking
you have slow healing wound or wounds and ulcers (open sores) on your feet
and legs that don’t heal
you have loss of feeling in the hands or feet and/ or loss of muscle mass in your legs
you have had medical or physical treatment for your pain, or suer from severe
pain, such as burning pain, in your legs and feet that continues even when you’re
resting.
you have suspected or diagnosed heart disease
you have suspected or diagnosed epilepsy
you are unsure about the suitability of the Revitive for you
you are unsure about the cause of your symptoms.
If your conditions worsen:
If your condition worsen during or aer use of the Revitive devices, it is up
to you to seek the necessary medical attention.
If you have a metallic implant:
If you have a metallic implant, you may experience pain or discomfort near the
implant when applying electrical stimulation, if this should occur discontinue use
and seek advice from your doctor
The electrical stimulation may feel more intense close to a metallic implant. It
is safe to continue use provided no pain is experienced. You may need to adjust the
intensity to a comfortable level.
Warnings
Do not use the Body Pads on your head, face, neck or chest:
There is a chance that serious health problems, and in rare cases death, may be
caused by using the Body Pads on your head, face, neck or chest
There is a chance that using the Body Pads on your head, face or neck may cause
closure of the airway, diiculty in breathing, or adverse eects on heart rhythm or
blood pressure
The introduction of electrical current into the chest may cause rhythm disturbances
to your heart, which could be lethal
Application of the electrodes near the thorax (chest) may increase the risk of cardiac
fibrillation
Use Body Pads only on parts of your body as directed in this guide.
Do not apply foot-pads or Body Pads directly on these areas:
On open wounds or rashes, swollen, red, infected, or inflamed areas or skin eruptions
(such as phlebitis, thrombophlebitis, varicose veins, cellulitis)
If the areas are cold and numb, turning red and then black, and/or beginning to
swell and produce foul-smelling pus with severe pain (gangrene)
On or close to, malignant tumours
On areas treated with radiotherapy (within the past 6 months)
On reproductive organs
Avoid areas that are pale, shiny, smooth and dry
On opposing thighs and refer to pad placement diagrams in this User’s Manual for
correct placement
2452_IFU06_18063164.indd 8-9
2452_IFU06_18063164.indd 8-9
19/05/2021 10:22
19/05/2021 10:22
10 11
Important Safeguards
Warnings
When applying foot-pads or body pads:
Apply stimulation only to normal, intact, healthy skin, and avoid areas that are pale,
shiny, smooth and dry
Electrode Body Pads are for single person use only. Do not share Body Pads with
others
• Do not use Body Pads aer their use by date. There is a chance you may experience
skin irritation or burns beneath the Body Pads
Some people may experience skin irritation or hypersensitivity due to electrical
stimulation or electrical conductive medium. The irritation can usually be reduced by
using an alternative conductive medium (gel) or alternate electrode placement
Ensure that balm/gel/moisturiser is evenly applied and thoroughly absorbed into the
skin before applying stimulation. There is a chance that uneven application of
a balm/ gel/ moisturiser could increase the risk of skin irritation or burn, when using
the electrical stimulation.
Do not use when the body pads are open circuited (e.g., when the pads are held
in your hands).
Do not remove electrodes when stimulation is being applied (to avoid unintended,
possibly hazardous, stimulation pathways.
Begin at the lowest intensity setting. Place feet on the device and gradually
begin to increase the intensity. Never start at a high level of intensity.
Keep feet on the device at all times, start at lowest intensity and slowly increase
the intensity. Do not remove feet during treatment or place feet on the device
aer the intensity has been increased.
Symptoms may worsen during the initial treatment phase before getting better.
This may occur if the body has not yet adjusted to increased muscular activity and
blood circulation. If this occurs, reduce the intensity and the duration of treatment
which will reduce the initial symptoms. If symptoms persist consult with your doctor
Warnings
When not to use Revitive:
• There are times you should not use Revitive. Do not use Revitive:
with electronic monitoring equipment (such as cardiac monitors, ECG
alarms), which may not operate properly when the electrical stimulation
device is in use
when you are in the bath or shower
while you are sleeping
while you are driving, operating machinery, or during any activity in which
electrical stimulation can put you at risk of injury
2452_IFU06_18063164.indd 10-11
2452_IFU06_18063164.indd 10-11
19/05/2021 10:22
19/05/2021 10:22
12 13
Important Safeguards
Cautions
When applying stimulation:
Be careful when applying stimulation:
over areas of skin that lack normal sensation - be careful not to overstimulate the
area. Check for signs of skin irritation (redness), bruising or pain. If in doubt
consult your doctor
over the abdomen during menstruation, or pregnant uterus
aer recent surgical procedures as stimulation may disrupt your healing process
if you are likely to bleed internally, such as following an injury or fracture
– use a low intensity and/or shorter time to avoid over-stimulation
aer a long period of immobility or inactivity – use a low intensity and shorter
time to avoid over-stimulation or muscle fatigue
This product is not intended for use by persons with reduced physical, sensory or
mental capabilities, unless they are supervised by a person responsible for
their safety
Use this device only with the leads, electrodes and accessories recommended by the
manufacturer.
It is possible Revitive may reduce the sensation of pain that would normally
act as a warning of an underlying problem
If your pain does not improve, gets worse or continues for more than 5 days, stop
using the device and seek advice from your doctor
The long-term eects of electrical stimulation are unknown.
Cautions
Additional Cautions
Electrical nerve stimulators are not a substitute for pain medications and other
pain management therapies
Electrical nerve stimulators are not eective for pain of central origin, including
headache
Electrical nerve stimulator devices have no curative value
Electrical nerve stimulators are a symptomatic treatment and, as such, suppresses
the sensation of pain that would otherwise serve as a protective mechanism
Eectiveness is highly dependent upon patient selection by a practitioner
qualified in the management of pain patients
The long-term eects of electrical stimulation are unknown
Since the eects of stimulation of the brain are unknown, stimulation should not be
applied across the head, and electrodes should not be applied on opposite sides of
the head
The safety of electrical stimulation during pregnancy has not been established
Use caution following recent surgical procedures when stimulation may disrupt the
patient’s healing process
Patients with suspected or diagnosed heart disease should follow precautions
recommended by their physicians
Patients with suspected or diagnosed epilepsy should follow precautions
recommended by their physicians.
2452_IFU06_18063164.indd 12-13
2452_IFU06_18063164.indd 12-13
19/05/2021 10:22
19/05/2021 10:22
14 15
Adverse reactions
Skin irritation and burns beneath the electrodes have been reported with the use
of powered muscle stimulators
You should stop using the device and should consult with your physician
immediately if you experience adverse reactions from the device.
Save these instructions
Safety Precautions
Do not stand on the machine. Use only when sitting down
Use the device only for its intended purpose
Do not expose the device to extreme heat
Do not spill liquid on the device or its accessories
Do not overload the electrical outlet
Keep device out of the reach of children
Keep power cords and cables out of the reach of children to prevent risk of
strangulation
Revitive may be used by multiple persons, ensure device is cleaned aer each use
Use Revitive only with the leads, electrodes, and accessories supplied by, or
purchased from the manufacturer
Check body pads, cords and cables periodically for damage
Do not open Revitive or repair it yourself. This will invalidate your warranty and may
serious harm
In the unlikely event your Revitive malfunctions, disconnect it from the power source
and contact your nearest authorized agent
Revitive has passed the required tests for EMI, it may still be aected by excessive
emissions and/or may interfere with more sensitive equipment
Aer any exposure to hot or cold temperatures outside the specified operating range
of 50 - 104°F allow the product to re-adjust to the recommended operating
temperatures to ensure continued product performance.
Important Safeguards
2452_IFU06_18063164.indd 14-15
2452_IFU06_18063164.indd 14-15
19/05/2021 10:22
19/05/2021 10:22
16 17
Step 2:
Using the Foot-Pads
a Get into a comfortable seated position.
For best results sit with your knees and ankles at right
angles to your body
b Place Revitive on the floor in front of you.
Never stand on the Revitive unit.
Revitive is designed to be used while seated.
c Place both bare feet on the foot pads.
Make sure you remove all footwear, including socks/
stockings.
d If desired, apply a moisturising lotion to the soles of
your feet.
Hint: Applying a moisturising lotion to the soles of the
feet before using Revitive can help improve the
delivery of electrical stimulation.
e Press the power button on the Revitive.
You will hear a beep and the display lights will light up.
The time will read 30 minutes and Intensity will
read 00.
To turn o the device at any time during operation, press
the power button.
Sit with
both bare feet
on the foot pads
R
R
Instructions for use
R
R
Instructions for use
Step 1: Setting up Revitive for the rst time
a Remove all parts from the carton:
Revitive Essential Power adapter
• Electrode Body Pads Body pad cords
See What is inside the Box? pages 4-5 to help identify the parts
b Locate the power adapter.
c Plug the small end of the power cord into the opening
on the side of the Revitive.
d Plug the power adapter into the nearest electrical
socket.
R
R
How do I get the most from Revitive?
For best results:
Use Revitive foot-pads for at least 30 to 60 minutes each day, 7 days a week.
It is important to use Revitive at a high enough intensity to give you a strong
comfortable muscle contraction. Most people achieve a strong muscle
contraction over intensity level 30 (intensity range 1-50).
It can take up to 8 weeks to reduce symptoms.
Do not use Revitive for more than 6 sessions of 30 minutes (or the equivalent)
per day.
2452_IFU06_18063164.indd 17
2452_IFU06_18063164.indd 17
19/05/2021 10:22
19/05/2021 10:22
18 19
i When Revitive times out, the Time display reads 00 and you will hear three
beeps, and the device will switch itself o automatically.
Important: the foot pads will not produce a stimulation if the body pad cord is
plugged in. To resolve this, remove the body pad cords.
FOR BEST RESULTS
FOR 60 DAYS
INTENSITY +
30
30-60 MIN
USE DAILY
f Press the (+) Intensity Control one time to
start stimulation.
Press the (+) button to increase the level.
Press the (–) button to decrease the level.
g Ensure you familiarise yourself with the feeling of
Electrical Muscle Stimulation (EMS). Use on a low
intensity for 30-60 minutes once per day for a few days.
Set the intensity at a level where you can feel the
stimulation in your feet and which creates a calf
contraction.
It is normal to feel tingling or varying sensations in
your calf muscles.
Revitive does not vibrate.
h Aer a few days of familiarisation, use a higher intensity
that provides you with strong but comfortable calf
muscle contractions.
Most people experience a strong contraction at intensity
level 40 or over. You can set the intensity between
1-50.
Please note that the intensity required may vary from
day to day.
If you feel pain or discomfort:
Remove one or both feet from the foot-pads to
stop the stimulation.
Lower the intensity of the foot-pads using the
(–) Intensity level button on the control panel
before replacing your feet on Revitive.
R
R
Instructions for use
2452_IFU06_18063164.indd 18-19
2452_IFU06_18063164.indd 18-19
19/05/2021 10:22
19/05/2021 10:22
20 21
a Select the area you wish to apply electrical stimulation.
Example placements are shown below:
b Clean and dry the area where you will place the body pads.
c Make sure Revitive is switched o. If the control panel is still
illuminated, press the power button and turn Revitive o.
Thigh
Knee
Lower back
Back of
Shoulder
Upper back
Targeting other areas of the body
Important: the foot pads will not produce a stimulation if the Electrode Body Pads
are plugged in.
Revitive Electrode Body Pads, used to deliver electrical stimulation to the knees,
thighs, shoulders, or back are included with every Revitive.
Electrode body pads are for single person use only. Do NOT share body pads with
others.
There is a chance that serious health problems, and in rare cases death,
may be caused by using the body pads on your head, face, neck or chest.
There is a chance that using the body pads on your head,
face or neck may cause closure of the airway, diiculty in
breathing, or adverse eects on heart rhythm or blood pressure.
The introduction of electrical current into the chest may cause
rhythm disturbances to your heart, which could be lethal.
Use body pads only on parts of your body as directed in this guide.
Refer to the Risks on page 8 before using the body pads.
DO NOT PLACE BODY PADS ON YOUR:
Head/Face Neck
Chest
Instructions for use
2452_IFU06_18063164.indd 20-21
2452_IFU06_18063164.indd 20-21
19/05/2021 10:22
19/05/2021 10:22
22 23
h Press the power button on the Revitive.
You will hear a beep and the display lights will read
30 minutes.
To turn o Revitive at any time during
operation, press the power button on the Revitive .
i Press the (+) Intensity Control, one time, to
start stimulation.
Press the (+) button to increase the level.
Press the (–) button to decrease the level.
j Use an intensity that provides you with strong but
comfortable muscle contraction and note that the
intensity required may vary from day to day.
It is normal to feel tingling or varying sensations in
your muscles.
If you feel pain or discomfort:
Lower the intensity of the Body Pads using the (–)
Intensity Control.
You can also choose to stop therapy before the
end of the 30 minute session to avoid skin irritation.
You can do this by pressing the power button.
R
R
R
R
d Attach the ends of the body pads to the body pad
cords.
Press connectors until no gap is between them.
Leaving a gap may cause discomfort.
e Plug the other end of the body pad cord into the
opening on the side of Revitive
f Remove the plastic liner from the body pads, by
peeling it o carefully. Replace the body pads onto the
plastic liner aer use.
g Press the body pads on the area(s) you wish to apply
electrical stimulation, as shown on page 18, gently
pressing the adhesive side against the skin.
R
R
R
R
R
R
R
R
R
R
Instructions for use
2452_IFU06_18063164.indd 23
2452_IFU06_18063164.indd 23
19/05/2021 10:22
19/05/2021 10:22
24 25
Replacing the Electrode Body Pads
The body pads are reusable up to 20-30 applications. If the body pads start to
lose their adhesiveness, rub a few drops of water into the surface. Replacement
Electrode Body Pads can be obtained via www.revitive.com
Cleaning and storing Revitive
a Press the power button and turn Revitive o.
b Wipe down the foot pads with a so damp cloth.
Do not clean with chemicals.
Do not immerse Revitive in water.
c Store Revitive in a cool, dry and dust-free location.
Store out of direct sunlight.
d Store the body pads in a dry and ventilated location.
Store out of direct sunlight.
You do not have to detach the body pads from the body pad cords to
store them.
Note: In the case of multiple users, ensure to clean the device aer each use.
R
R
k The Time display will start to count down in
minutes as Revitive cycles through its program.
l When Revitive times out, the Time display reads
00 and you will hear three beeps, and the device will
switch itself o automatically.
m Gently peel the body pads from the skin.
Do not pull the body pads o the skin using the
body pad cords as this may damage the cords.
n Replace the body pads onto its plastic liner aer use.
o Unplug the body pad cord from the opening on
Revitive.
You do not need to detach the body pads from the
body pad cords.
R
R
R
R
R
Instructions for use
2452_IFU06_18063164.indd 24-25
2452_IFU06_18063164.indd 24-25
19/05/2021 10:22
19/05/2021 10:22
26 27
Problem Possible Cause Solution
No power or lights
to Revitive when it is
switched on.
Power adapter not switched on
at the electrical outlet or power
adapter not plugged into device
properly.
Check electrical outlet power
is switched on and the power
adapter is plugged into the device
correctly. If still not working –
contact your authorised dealer.
Revitive is not vibrating.
REVITIVE IS NOT DESIGNED TO VIBRATE.
My legs are aching
aer treatment.
You may have the
intensity on
too high a setting and your muscles
are being overworked.
Leave adequate time aer each
treatment to allow the muscles
to recover (just like aer vigorous
exercise!). On your next session
start on a lower setting (where
you can feel the mild electrical
stimulation and it is comfortable)
and reduce the duration until your
muscles have acclimatized to the
stimulation.
For more information including Frequently Asked Questions,
please visit www.revitive.com
FOR BEST RESULTS
FOR 60 DAYS
INTENSITY +
30
30-60 MIN
USE DAILY
Troubleshooting
Problem Possible Cause Solution
Revitive is on (lights
illuminated on the LED
display) but I cannot
feel the electrical
stimulation through the
foot pads.
Not placing both bare feet on the
foot pads at the same time.
Ensure that your feet are bare
and each foot is placed on each
of the foot pads at the same
time. Keep increasing the
intensity up to a maximum of 50
until you feel the stimulation.
Your feet may be dry. Moisturize the soles of your feet
to improve conductivity and
stimulation and try the procedure
again. You may also have to increase
the intensity level.
You may be dehydrated. Drink plenty of water before
and aer using the device. The
device uses your body to create
the electrical circuit. Water is an
excellent conductor of electricity
and if your body is less hydrated
(below 60%) then the stimulation
may be less. Therefore it is
important to always remain well
hydrated.
The intensity level may be on too
low a setting. This is a very safe device. Keep
increasing the intensity level
towards 50 until you feel the
stimulation. You may find that you
have to increase the intensity
level as you get used to the therapy.
The aim is not to get to 50 but to find
a setting that is comfortable for you.
The electrode body pads are
connected. Turn o the device and disconnect
the body pad cords from the device
and try again
If, having tried the solutions above,
you still cannot feel the stimulation:
Test the device by placing one hand
across both footpads at the same
time (The heel end of the footpads
is easiest). With your other hand,
and starting from zero, increase
the intensity level until you can
feel the stimulation. If you can feel
the stimulation through your hand
then the device is working. If on 50
you still cannot feel the stimulation
then please contact your authorised
dealer.
2452_IFU06_18063164.indd 26-27
2452_IFU06_18063164.indd 26-27
19/05/2021 10:22
19/05/2021 10:22
28 29
Net Charge @ 500Ω [0.001]mC
Maximum Phase Charge @ 500 Ω Foot: 18µC Body: 18µC
Maximum Current Density Foot: 0.011 mA/cm² Body: 0.084 mA/cm²
Maximum Power Density @ 500Ω Foot: 0.68 mW/cm² Body: 1.66 mW/cm²
Burst Mode: a. Pulses per burst Foot: 32-328 Body: 32-328
b. Bursts per second Foot: 0.11 - 0.24 Body: 0.1 - 0.2
c. Burst duration Foot: 3.2 - 8.3s Body: 3.2 - 8.3s
d. Duty Cycle Foot: 0.75 – 0.89 Body: 0.75 – 0.89
ON Time Foot: 1.9 - 8.3s Body: 1.9 - 6.3s
OFF Time Foot: 1.0 - 1.50s Body: 1.0s
Disposal: At the end of product life do not throw this product into normal household garbage, but take it to a
collection point for the recycling of electronic equipment.
Technical Specications & Symbols
Contraindications, Warnings and Cautions
Make sure you understand these before using the Revitive
Power
Time Remaining
Intensity Level
Center Positive Polarity
Class II medical electrical equipment double insulated
FCC mark
Certification mark employed on electronic products manufactured or sold in
the United States which certifies that the electromagnetic interference from
Revitive is under limits approved by the Federal Communications Commission
Technical Specications
Name of product Revitive Essential
Model RLV
Weight 1.13kg
Dimensions Ø 355 mm x 77(D) mm
Power Consumption 5W
Power Source:
AC adapter – UL (Underwriters Laboratories) Safety
Mark
Input
Output
100-240V AC, 50/60Hz, 0.18A
5v DC, 1.0A
Output Specifications:
Waveform Monophasic
Shape Square symmetrical with polarity reversal
Maximum Output Voltage (+/- 15%) : @500Ω Foot: 21.5Vp Body: 21.5Vp
@2kΩ Foot: 55Vp Body: 62.5Vp
@10kΩ Foot: 65Vp Body: 85Vp
Maximum Output Current: (+/- 15%) : @ 500Ω Foot: 43mA Body: 43mA
@2kΩ Foot: 27.5mA Body: 31.3mA
@10kΩ Foot: 6.5mA Body: 8.5mA
Pulse Duration (+/- 10%) : Foot: 450 or 970µs (Mode 7) Body 455µs
Frequency (+/- 10%) : Foot: 20 to 44Hz, Body: 20 to 44Hz
2452_IFU06_18063164.indd 28-29
2452_IFU06_18063164.indd 28-29
19/05/2021 10:22
19/05/2021 10:22
30 31
Your international two year guarantee
It is important to retain the retailer’s receipt as proof
of purchase. Staple your receipt to this back cover
for future reference.
Please quote the following information if the
product develops a fault. These numbers can be
found on the base of the product:
Model no: ............................................................
Serial no: .................................................................
All Revitive devices are individually tested before
leaving the factory. In the unlikely event of any
device proving to be faulty within 30 days of
purchase, it should be returned to the place of
purchase for it to be replaced.
If the fault develops aer 30 days and within 24
months of original purchase, you should contact
your local distributor quoting model number and
Serial number on the product, or write to your local
distributor at the address shown.
You will be asked to return the product (in secure,
adequate packaging) to the address shown with a
copy of proof of purchase.
Subject to the exclusions set out below (see
Exclusions), the faulty device will then be replaced
and dispatched usually within 14 working days of
receipt.
If, for any reason, this item is replaced during the
2-year guarantee period, the guarantee on the new
item will be calculated from the original purchase
date. Therefore, it is vital to retain your original
till receipt or invoice to indicate the date of initial
purchase.
To qualify for the 2-year guarantee, the device must
have been used according to the manufacturer’s
instructions supplied.
EXCLUSIONS:
1 Actegy Ltd, manufacturer of Revitive devices,
shall not be liable to replace the goods under the
terms of the guarantee where:
The fault has been caused or is attributable
to accidental use, misuse, negligent use
or used contrary to the manufacturer’s
recommendations or where the fault has
been caused by power surges or damage
caused in transit.
The device has been used on a voltage
supply other than that stated on the product
or used with a power adaptor other than the
one supplied with the product.
Repairs have been attempted by persons
other than our service sta (or authorized
dealer).
The device has been used for hire purposes
or non-domestic use.
• T he device is second hand.
2 Actegy Ltd are not liable to carry out any type of
servicing work, under the guarantee.
3 Accessories such as Electrode pads are not
covered by the guarantee.
4 Batteries and damage from leakage are not
covered by the guarantee.
5 This guarantee does not confer any
rights other than those expressly set
out above and does not cover any claims
for consequential loss or damage. This
guarantee is oered as an additional
benefit and does not aect your
statutory rights as a consumer.
To activate your free 2-year warranty please register your device at:
www.revitive.com/myrevitive
Symbols
Type BF medical electrical equipment
Legal manufacturer of the device
Ingress Protection Rating
Consult instructions for use
Do not dispose into normal household garbage.
Take to a collection point for the recycling of electrical equipment
Device serial number including year (YYYY) and month (MM) of
manufacture can be found on the box and back of unit
Item number
Use-by date
Humidity, temperature and air pressure limit for storage
and transport
Humidity, temperature and air pressure limit for operating
conditions
Aer any exposure to hot or cold temperatures outside the specified
operating range of 50 - 104°F allow the product to re-adjust to the
recommended operating temperatures to ensure continued product
performance.
Energy Eiciency Grade 5
Indoor Use Only
Lead wires comply with the Performance Standard for electrode lead
wires (21 CFR part 898)
Not made with natural rubber latex
Do not disassemble
#YYYYMMXXXXX
SN
REF
YYYYMMDD
-4°F
20%
500 hPa
1060 hPa
50°F
30%
104°F
75%
700 hPa
1060 hPa
158°F
90%
2452_IFU06_18063164.indd 31
2452_IFU06_18063164.indd 31
19/05/2021 10:22
19/05/2021 10:22
32
2452_IFU06_18063164 P.D 05.2021
Get the most out of life
Legal Manufacturer
Actegy Ltd
1 West Point
Western Road
Bracknell, RG12 1HJ
United Kingdom
Tel: +44 (0)800 014 6377
Country Contact & Distributor:
United States
Actegy Health Inc
6400 Shafer Court, Suite 225
Rosemont, IL 60018
Toll free: 1-877-260-3114
®
Copyright © 2021 Actegy Ltd. All rights reserved. Actegy®, Revitive® and Circulation Booster®
are registered trademarks of Actegy Ltd. The Revitive product is a proprietary design and is
protected by applicable design laws.
2452_IFU06_18063164.indd 322452_IFU06_18063164.indd 32 19/05/2021 10:2219/05/2021 10:22
16

Hulp nodig? Stel uw vraag in het forum

Spelregels

Misbruik melden

Gebruikershandleiding.com neemt misbruik van zijn services uitermate serieus. U kunt hieronder aangeven waarom deze vraag ongepast is. Wij controleren de vraag en zonodig wordt deze verwijderd.

Product:

Bijvoorbeeld antisemitische inhoud, racistische inhoud, of materiaal dat gewelddadige fysieke handelingen tot gevolg kan hebben.

Bijvoorbeeld een creditcardnummer, een persoonlijk identificatienummer, of een geheim adres. E-mailadressen en volledige namen worden niet als privégegevens beschouwd.

Spelregels forum

Om tot zinvolle vragen te komen hanteren wij de volgende spelregels:

Belangrijk! Als er een antwoord wordt gegeven op uw vraag, dan is het voor de gever van het antwoord nuttig om te weten als u er wel (of niet) mee geholpen bent! Wij vragen u dus ook te reageren op een antwoord.

Belangrijk! Antwoorden worden ook per e-mail naar abonnees gestuurd. Laat uw emailadres achter op deze site, zodat u op de hoogte blijft. U krijgt dan ook andere vragen en antwoorden te zien.

Abonneren

Abonneer u voor het ontvangen van emails voor uw Revitive Essential bij:


U ontvangt een email met instructies om u voor één of beide opties in te schrijven.


Ontvang uw handleiding per email

Vul uw emailadres in en ontvang de handleiding van Revitive Essential in de taal/talen: Engels als bijlage per email.

De handleiding is 3.17 mb groot.

 

U ontvangt de handleiding per email binnen enkele minuten. Als u geen email heeft ontvangen, dan heeft u waarschijnlijk een verkeerd emailadres ingevuld of is uw mailbox te vol. Daarnaast kan het zijn dat uw internetprovider een maximum heeft aan de grootte per email. Omdat hier een handleiding wordt meegestuurd, kan het voorkomen dat de email groter is dan toegestaan bij uw provider.

Stel vragen via chat aan uw handleiding

Stel uw vraag over deze PDF

Uw handleiding is per email verstuurd. Controleer uw email

Als u niet binnen een kwartier uw email met handleiding ontvangen heeft, kan het zijn dat u een verkeerd emailadres heeft ingevuld of dat uw emailprovider een maximum grootte per email heeft ingesteld die kleiner is dan de grootte van de handleiding.

Er is een email naar u verstuurd om uw inschrijving definitief te maken.

Controleer uw email en volg de aanwijzingen op om uw inschrijving definitief te maken

U heeft geen emailadres opgegeven

Als u de handleiding per email wilt ontvangen, vul dan een geldig emailadres in.

Uw vraag is op deze pagina toegevoegd

Wilt u een email ontvangen bij een antwoord en/of nieuwe vragen? Vul dan hier uw emailadres in.



Info