Technical data
Name: COMP AIR Pro
Model: NE-C900 (NE-C900-E)
Type: Compressor Nebulizer
Rating: AC 220-240V, 50Hz
Power Consumption: 220VA
Operating Temperature/
Humidity:
+10°C to +40°C / 30% to 85%RH
Storage and Transport
Temperature/Humidity/
Air Pressure:
-20°C to +60°C / 10% to 95%RH /
700 - 1060hPa
Weight: Approx. 2.2kg (compressor only)
Dimensions: Approx. 175 (W) × 110 (H) ×215 (D) mm
(compressor only)
Contents:
Compressor, Nebulizer Kit, Air Tube (PVC,
200cm), Mouthpiece, Nosepiece, Adult Mask
(PVC), Child Mask (PVC), 5 pcs Spare Air
Filters, Instruction Manual, Warranty Card.
Classification: Class ll equipment, Type B applied part
= Class ll
equipment
= Type B
applied part
Read the instruction
manual carefully
= Power off
= Power on
Technical data for the OMRON NE-C900 Compressor with the Nebulizer Kit:
Particle Size: **MMAD Approx. 3m
MMAD = Mass Median Aerodynamic Diameter
Medication Tank Capacity: 7ml maximum
Appropriate Medication
Quantities:
2ml minimum - 7ml maximum
Sound: *Approx. 55dB
Nebulization Rate: *Approx. 0.4ml/min (by weight loss)
Airflow/Compressor
Output:
*7L/min @100kPa
Aerosol Output: **0.57ml (2ml, 1%NaF)
Aerosol Output Rate: **0.08ml/min (2ml, 1%NaF)
Result of cascade impactor measurements for particle size**
*
Measured by OMRON HEALTHCARE Co., Ltd.
**
Independently measured by Prof. Dr. Hiroshi Takano, Department of Chemical
Engineering and Materials Science, Faculty of Science and Engineering, Doshisha
University, Kyoto, Japan, according to EN 13544-1:2007+A1:2009.
Notes:
Subject to technical modification without prior notice.
This OMRON product is produced under the strict quality system of OMRON
HEALTHCARE Co., Ltd., Japan.
The device may not work if the temperature and voltage conditions are different to
those defined in the specifications.
This device fulfils the provisions of the EC directive 93/42/EEC (Medical Device
Directive) and the European Standard EN13544-1:2007+A1:2009, Respiratory
therapy equipment - Part1: Nebulizing systems and their components.
Performance may vary with drugs such as suspensions or high viscosity. See drug
supplier’s data sheet for further details.
See website of OMRON HEALTHCARE EUROPE to update technical information.
URL: www.omron-healthcare.com
Removing condensation from the air tube
If there is moisture or liquid remaining in the air tube, be sure to follow the
procedure below to remove any moisture from within the air tube.
1) Make sure that the air tube is connected to the air connector on the
compressor.
2) Remove the air tube from the nebulizer kit.
3) Turn on the compressor and pump air through the air tube to expel the
moisture.
Changing the air filter
If the air filter has changed colour, or has been used on average for more
than 60 days, replace it with a new one.
1. Pull the air filter cover off the compressor.
2. Change the air filter.
Remove the old air filter with a sharp object,
such as a toothpick, and insert the new air
filter.
3. Put the air filter cover back in place.
Notes:
Use only OMRON air filters designed for this device. Do not operate without a filter.
Do not attempt to wash or clean the filter. If the air filter becomes wet, replace it. Damp
air filters can cause blockages.
There is no front or back orientation for the air filters.
Check that air filters are clean and free of dust before inserting them.
To prevent the air filter cover from becoming blocked, wash the cover regularly. Do not
boil the air filter cover. Remove the air filter cover, wash it, and make sure it is dry and
reattached correctly.
Cleaning and daily disinfecting
Cleaning
Clean the parts after each use to remove residual medication. This will avoid
inefficient nebulization or infection.
Firstly, make sure that the power plug unplugged from the power outlet.
Nebulizer Kit (Inhalation top, Medication tank), Mask (PVC),
Mouthpiece, Nosepiece
Wash them in warm water and mild detergent. Rinse them thoroughly with
clean hot tap water and allow to air dry in a clean place.
Note: Nebulizer kit should be replaced after using for one year.
Compressor, Air Tube
Wipe clean with a soft cloth moistened with water or mild detergent.
Disinfecting
Always disinfect the parts after the last treatment of the day. If the parts are
heavily stained, replace them with new ones.
Use one of the following methods:
To select the method, refer to the table below.
A. Use a commercially available disinfectant. Follow the instructions
provided by the disinfectant manufacturer.
1) Submerge the parts in the disinfectant for the specified period.
2) Rinse them with clean hot tap water and allow to air dry in a clean place.
Note: Never clean with benzene, thinner or a flammable chemical.
B. Parts may be boiled between 15 to 20 minutes.
After boiling, carefully remove the parts, shake
off excess water and allow to air dry in a clean
environment.
Note: Do not boil Air tube, Mask (PVC), Air filter and Air filter
cover.
Use the table below as a guideline to select a method for disinfecting.
O: applicable ×: not applicable
Parts
Mouthpiece
Nosepiece
Nebulizer Kit
Air Filter
Air Filter Cover
Adult Mask (PVC)
Child Mask (PVC)
Infant Mask (PVC)
(optional)
Air Tube
(PVC, 200cm)
Adult Mask (SEBS)
(optional)
Child Mask (SEBS)
(optional)
Materials
PP
PP
PP
Polyester
ABS
Mask: PVC
(Phthalate free)
Band: Rubber
(non-latex)
PVC
(Phthalate free)
Mask: SEBS
Band: Rubber
(non-latex)
Adapter: PP
Boiling O O O × × × × O
Alcohol
OOO×O O × O
Disinfecting
ethanol
Sodium
hypochlorite
OOO×O O × O
Milton*
(0.1%)
Quaternary
ammonium
OOO×O O × O
Osvan*
(0.1%)
Chlorhexidine
OOO×O O × O
Hibitane*
(0.5%)
Amphoteric
Surfactant
OOO×O O × O
Tego*
(0.2%)
*
an example of commercially available disinfectant.
How to use
1. Make sure that the power switch is in the off (
) position.
2. Plug the power plug into a power outlet.
3. Remove the inhalation top from the medication
tank.
4. Add the correct amount of prescribed medication
to the medication tank.
5. Put the inhalation top back inside the medication
tank.
6. Attach the mask, mouthpiece or nosepiece to the nebulizer kit tightly.
7. Attach the air tube.
Twist the air tube plug slightly and push it firmly into the air tube connector.
8. Hold the nebulizer kit as
indicated on the right.
Follow the instructions of
your doctor or respiratory
therapist.
Caution:
Do not tilt the nebulizer kit at an
angle of greater than 45 degrees
in all directions. Medication may
flow into the mouth or it may result
in not nebulizing effectively.
9. Press the power switch to the on (
) position. As the compressor
starts, nebulization begins and aerosol is generated.
Inhale the medication.
10. When treatment is complete, turn the power off and unplug the
compressor from the power outlet.
Troubleshooting
Check the following if your unit should fail during operation.
Problem Cause Remedy
Nothing happens when the
power switch is pressed.
Is the power plug plugged into
a power outlet?
Check that the plug is inserted
in a power outlet. Unplug then
reinsert the plug if necessary.
No nebulization or low
nebulization rate, when the
power is on.
Is there medication in the
medication tank?
Add the correct amount of
medication to the medication tank.
Is there too much/little
medication in the medication
tank?
Is the nebulizer kit assembled
correctly?
Assemble the nebulizer kit
correctly.
Is the nozzle blocked? Make sure that the nozzle is free
of blockages.
Is the nebulizer kit tilted at a
sharp angle?
Make sure that the nebulizer kit is
not tilted at an angle of more than
45 degrees.
Is the air tube connected
correctly?
Make sure that the air tube
is correctly connected to the
compressor and nebulizer kit.
Is the air tube folded or
damaged?
Make sure that the air tube does
not contain kinks.
Is the air tube blocked? Make sure that the air tube is free
of blockages.
Is the air filter dirty? Replace the air filter with a new
one.
The device is abnormally
loud.
Is the air filter cover attached
correctly?
Attach the air filter cover correctly.
The device is very hot. Is the compressor covered?
Do not cover the compressor with
any type of cover during use.
Note:
If the suggested remedy does not solve the problem, do not try to repair the device
- no parts of the unit are user serviceable. Return the unit to an authorized
OMRON retail outlet or distributor.
1
2
1
2
Clicks
Right angle
45°45°
100
90
80
70
60
50
40
30
20
10
0
1 10 1000.1
Cumulative % particle mass of sodium fluoride under size
Cumulative Undersize%
Particle Size Dp (m)
Individual tests
Mean
2

Hulp nodig? Stel uw vraag in het forum

Spelregels

Misbruik melden

Gebruikershandleiding.com neemt misbruik van zijn services uitermate serieus. U kunt hieronder aangeven waarom deze vraag ongepast is. Wij controleren de vraag en zonodig wordt deze verwijderd.

Product:

Bijvoorbeeld antisemitische inhoud, racistische inhoud, of materiaal dat gewelddadige fysieke handelingen tot gevolg kan hebben.

Bijvoorbeeld een creditcardnummer, een persoonlijk identificatienummer, of een geheim adres. E-mailadressen en volledige namen worden niet als privégegevens beschouwd.

Spelregels forum

Om tot zinvolle vragen te komen hanteren wij de volgende spelregels:

Belangrijk! Als er een antwoord wordt gegeven op uw vraag, dan is het voor de gever van het antwoord nuttig om te weten als u er wel (of niet) mee geholpen bent! Wij vragen u dus ook te reageren op een antwoord.

Belangrijk! Antwoorden worden ook per e-mail naar abonnees gestuurd. Laat uw emailadres achter op deze site, zodat u op de hoogte blijft. U krijgt dan ook andere vragen en antwoorden te zien.

Abonneren

Abonneer u voor het ontvangen van emails voor uw Omron CompAIR Pro - C900 bij:


U ontvangt een email met instructies om u voor één of beide opties in te schrijven.


Ontvang uw handleiding per email

Vul uw emailadres in en ontvang de handleiding van Omron CompAIR Pro - C900 in de taal/talen: Engels als bijlage per email.

De handleiding is 1,96 mb groot.

 

U ontvangt de handleiding per email binnen enkele minuten. Als u geen email heeft ontvangen, dan heeft u waarschijnlijk een verkeerd emailadres ingevuld of is uw mailbox te vol. Daarnaast kan het zijn dat uw internetprovider een maximum heeft aan de grootte per email. Omdat hier een handleiding wordt meegestuurd, kan het voorkomen dat de email groter is dan toegestaan bij uw provider.

Stel vragen via chat aan uw handleiding

Stel uw vraag over deze PDF

loading

Uw handleiding is per email verstuurd. Controleer uw email

Als u niet binnen een kwartier uw email met handleiding ontvangen heeft, kan het zijn dat u een verkeerd emailadres heeft ingevuld of dat uw emailprovider een maximum grootte per email heeft ingesteld die kleiner is dan de grootte van de handleiding.

Er is een email naar u verstuurd om uw inschrijving definitief te maken.

Controleer uw email en volg de aanwijzingen op om uw inschrijving definitief te maken

U heeft geen emailadres opgegeven

Als u de handleiding per email wilt ontvangen, vul dan een geldig emailadres in.

Uw vraag is op deze pagina toegevoegd

Wilt u een email ontvangen bij een antwoord en/of nieuwe vragen? Vul dan hier uw emailadres in.



Info