How to use:
1. Beforeusing,pleasedisinfecttheprobewithalcoholrst.
. Presstheon/offbutton,Bisoundwillbeheardanddisplay„88.88°Cforaboutseconds.
3. Afterthat,thethermometerwilldisplaythelasttemperaturememoryrecordforaboutseconds,
andthendisplayL°C(Lo°C),signal°Cashing,thatmeansthethermometeriswaitingforuse.If
thetemperaturemeasuredisbeloworabovethedisplayrangethethermometerwilldisplayeither
the„L“orthe„H“symbol.Thedisplayrangeforthisthermometeris3.00°Cto43.00°C.
4. Usethethermometertomeasurebodytemperature.Thevalueoftemperaturedisplayedwillarise
graduallyand°Cisashing.
5. Themeasurementprocessusuallytakesabout1min(inoral).Aftermeasurement,°Cwillstop
ashingandthealarmsignalwillsoundforabout5seconds,atthesametime,thethermometer
willdisplaytheuser‘sbodytemperature.
6. This thermometer can automatically shut off within 10 minutes after measurement. It is
recommendedtoturnoffthepowermanuallyrightaftermeasurementinordertosaveenergy.
Take temperature:
Whenyoumeasureyourbodytemperature,pleaseremaintomeasureatthesameplaceforsome
time.
1. Totaketemperatureundertongue,beforemeasurementpleasecloseyourmouthfor5minutes,
theninsertthesensorpartunderthetongue,keepingnestledandcloseyourmouth.LCDreading
isnalreading.
. Totaketemperatureinthearmpit,beforemeasurementpleasecompletelywipethesweatthere
away,thenclosethearmpitforabout5minutesandinsertthesensorpartagainstthecenterof
thearmpitclosely.Putbothleftandrightwaistintersection,makingthethermometerincontact
withtheuser‘sbodyclosely.Add0.5°CtoLCDreadingtoobtainnalreading.
3. Totaketemperatureintherectum,insertsthesensorpartintoanusgently.Minus0.5°CtoLCD
readingtoobtainnalreading.
Cautions:
1. Donotbendordropthethermometer.
. Pleasedisinfecttheprobewithalcohol,donotboiltheprobe.
3. Donotputthethermometerincontactwithchemicalreagentordiluents.
4. Avoidexposingthethermometerunderdirectsunlight.
5. Don‘tdisassemblethethermometerexceptreplacingthebattery.
6. Thisthermometerhasalreadybeencalibratedwhilemaking.Ifuseaccordingtotheoperation
instruction,doesnotneedtocalibrateagain.
Battery replacement:
1. ThebatteryneedstobereplacedbyanewbatterywhentheLCDdisplays or .
. Pleaseopenthebatterycover,takeouttheoldbatterywithatoothpick.
3. Placeanew1.5Vbuttonbattery(LR/SR-41).
4. Replacethebatterycover.
: Degree of protection against electric shock is Type B applied part.
: The disposal of battery and device shall comply with the local environment requirements.
The device comply with the standard of IEC 60601-1:1998+ A1:1991+ A2:1995 and IEC60601-1-2:2001