See picture 7-4.
Picture 7 Picture 7-1 Picture 7-2 Picture 7-3 Picture 7-4
c. Press “MEM” button again, LCD will display the average value of all the
results which is measured from 18 o’clock to 20 o’clock in last 7 days in
the current user memory zone. See picture 7-5. If no result stored from 18
o’clock to 20 o’clock in last 7 days, LCD will display “0” for blood pressure
and pulse rate. See picture 7-6.
Picture 7-5 Picture 7-6 Picture 7-7 Picture 7-8 Picture 7-9
d. Press “MEM” button again, the most recent result will be displayed with date
and time stamp. See picture 7-7. Irregular heartbeat symbol (if any) and blood
pressure classication indicator will blink at the same time. If the monitor has
no result stored in the current user memory zone, the LCD will display “0” for
blood pressure and pulse rate. See picture 7-8.
e. Press “MEM” button again to review the next result. See picture 7-9. In this
way, repeatedly pressing the “MEM” button displays the respective results
measured previously.
f. When reviewing the results, the monitor will turn off automatically after 1 min-
ute of no operation. You can also press the “START/STOP” button to turn off
the monitor manually.
Note: When the monitor displaying the measurement, the classication color
indicator can be shown different color according to the systolic pressure and
diastolic pressure. Refer to the “ASSESSING HIGH BLOOD PRESSURE FOR
ADULTS” section.
8. Deleting measurements from the memory
When any result is displaying, keeping on pressing but-
ton “MEM” for three seconds, all results will be deleted.
Press the button “START/STOP”, the monitor will turn off.
9. Assessing high blood pressure for adults
The following guidelines for assessing high blood pressure (without regard to
age or gender) have been established by the World Health Organization (WHO).
Please note that other factors (e.g. diabetes, obesity, smoking, etc.) need to be
taken into consideration. Consult with your physician for accurate assessment,
and never change your treatment by yourself.
Classication of blood pressure for adults
MAINTENANCE
1. Do not drop this monitor or subject it to strong impact.
2. Avoid high temperature and solarization. Do not immerse the monitor in
water as this will result in damage to the monitor.
Note: It is not intended to provide a basis of any type of rush toward emergency
conditions/diagnosis based on the color scheme and that the color scheme is
meant only to discriminate between the different levels of blood pressure.
10. Technical alarm description
The monitor will show ‘HI’ or ‘Lo’ as technical alarm on LCD with no delay if
the determined blood pressure (systolic or diastolic) is outside the rated range
specied in part SPECIFICACIONS. In this case, you should consult a physician
or check if your operation violated the instructions.
The technical alarm condition (outside the rated range) is preset in the factory
and cannot be adjusted or inactivated. This alarm condition is assigned as low
priority according to IEC 60601-1-8.
The technical alarm is non-latching and need no reset. The signal displayed on
LCD will disappear automatically after about 8 seconds.
12. Troubleshooting (2)
ELECTROMAGNETIC COMPATIBILITY INFORMATION
Table 1
Emission
12 It is recommended the cuff should be disinfected 2 times every week if need-
ed (For example, in hospital or clinic). Wipe the inner side (the side contacting
the skin) of the cuff with a soft cloth squeezed after being moistened with
Ethyl alcohol (75-90%), then dry the cuff by airing.
EXPLANATION OF SYMBOLS ON UNIT
Finally the blood pressure and pulse rate will be calculated and displayed
on the LCD screen. The blood pressure classication indicator and Irregular
heartbeat symbol (if any) will blink on the screen. The result will be automati-
cally stored in the monitor.
d. After measurement, the monitor will turn off automatically after 1 minute of no
operation.
e. During measurement, you can press the “START/STOP” button to turn off the
monitor manually.
Note: Please consult a health care professional for interpretation of pressure
measurements.
7. Displaying stored results
a.
In StandBy Mode, press “MEM” button, the monitor will display sign of current
group. The amount of results in current user memory zone will be displayed. See
picture 7. Press “START/STOP” button to switch group,press “MEM” to conrm
current group. Then LCD will display the average value of all results in the current
user memory zone. See picture 7-1. If no result stored in the current user memory
zone, LCD will display “0” for blood pressure and pulse rate. See picture 7-2.
b. Press “MEM” button, LCD will display the average value of all the results
which is measured from 5 o’clock to 9 o’clock in last 7 days in the current
user memory zone. See picture 7-3. If no result stored from 5 o’clock to 9
o’clock in last 7 days, LCD will display “0” for blood pressure and pulse rate.
M32901-GB-Rev-2-02.20
21
22
23
24
20
19
27
28
29
30
26
25
33
34
35
36
32
31
32901 / KD-5923
PROBLEM POSSIBLE CAUSE SOLUTION
LCD shows
battery symbol
Low Battery Change the batteries
LCD shows “Er 0” Pressure system is unstable
before measurement
Don’t move and try
again
LCD shows “Er 1” Fail to detect systolic pressure
LCD shows “Er 2” Fail to detect diastolic
pressure
LCD shows “Er 3” Pneumatic system blocked or
cuff is too tight during ination
Apply the cuff correctly
and try again.
If the monitor is still ab-
normal, please contact
the local distributor or
the factory.
LCD shows “Er 4” Pneumatic system leakage
or cuff is too loose during
ination
LCD shows “Er 5” Cuff pressure above
300mmHg
Measure again after ve
minutes. If the monitor
is still abnormal, please
contact the local
distributor or the factory
LCD shows “Er 6” More than 3 minutes with cuff
pressure above 15 mmHg
LCD shows “Er 7” Inner memory error
LCD shows “Er 8” Device parameter checking
error
LCD shows “Er A” Pressure sensor parameter
error
No response when
you press button or
load battery
Incorrect operation or strong
electromagnetic interference
Take out batteries for
ve minutes, and then
reinstall all batteries
Table 2
Enclosure Port
Table 3
Proximity elds from RF wireless communications equipment
11. Troubleshooting (1)
PROBLEM POSSIBLE CAUSE SOLUTION
LCD Display shows
abnormal result
The cuff position was
not correct or it was not
properly tightened
Apply the cuff correctly
and try again
Body posture was not
correct during testing
Review the “BODY POSTURE
DURING MEASUREMENT”
sections of the instructions
and re-test
Speaking, arm or body
movement, angry, excited
or nervous during testing
Re-test when calm and
without speaking or moving
during the test
Irregular heartbeat
(arrhythmia)
It is inappropriate for people
with serious arrhythmia
to use this Electronic
Sphygmomanometer
Disposal: The product must not be disposed of along with other domestic
waste. The users must dispose of this equipment by bringing it to a specic
recycling point for electric and electronic equipment.
GIMA WARRANTY TERMS
The Gima 12-month standard B2B warranty applies.
WHO/ISH Denitions and classication of blood pressure levels
80 85 90 100 110
120
130
140
160
180
Severe Hypertension
Moderate Hypertension
Mild Hypertension
High-normal BP
Normal BP
Diastolic
(mmHg)
Systolic
(mmHg)
BLOOD PRESSURE
CLASSIFICATION
SBP
mmHg
DBP
mmHg
Color
Indicator
Optimal
<120 <80 green
Normal
120-129 80-84 green
High-Normal
130-139 85-89 green
Grade 1 Hypertension 140-159 90-99 yellow
Grade 2 Hypertension 160-179 100-109 orange
Grade 3 Hypertension ≥180 ≥110 red
Follow instructions for use
Caution: read instructions (warnings) carefully
Type BF applied part
WEEE disposal
0197
Medical Device complies with Directive
93/42/EEC
Manufacturer
Date of manufacture
Authorized representative in the European
community
Serial number
IP20 Covering Protection rate
Keep in a cool, dry place
Keep away from sunlight
Product code
Lot number
ANDON HEALTH CO., LTD.
No. 3 JinPing Street, Ya An Road, Nankai District,
Tianjin 300190, China
Made in China
iHealthLabs Europe SAS
36 Rue de Ponthieu,
75008, Paris, France
0197
IP20
Phenomenon Compliance Electromagnetic
environment
RF emissions CISPR 11 Group 1,
Class B
Home healthcare
environment
Harmonic distortion IEC 61000-3-2
Class A
Home healthcare
environment
Voltage uctuations
and icker
IEC 61000-3-3
Compliance
Home healthcare
environment
Phenomenon Basic EMC
standard
Immunity test levels
Home Healthcare
Environment
Electrostatic Discharge IEC 61000-4-2 ±8 kV contact
±2kV, ±4kV, ±8kV, ±15kV air
Radiated RF EM eld IEC 61000-4-3 10V/m
80MHz-2.7GHz
80% AM at 1kHz
Proximity elds from RF
wireless communications
equipment
IEC 61000-4-3 Refer to table 3
Rated power frequency
magnetic elds
IEC 61000-4-8 30A/m
50Hz or 60Hz
Test frequency
(MHz)
Band (MHz) Immunity test levels
Professional healthcare facility
environment
385 380-390 Pulse modulation 18Hz, 27V/m
450 430-470 FM, ±5kHz deviation, 1kHz sine,
28V/m
710 704-787 Pulse modulation 217Hz, 9V/m
745
780
810 800-960 Pulse modulation 18Hz, 28V/m
870
930
1720 1700-1990 Pulse modulation 217Hz, 28V/m
1845
1970
2450 2400-2570 Pulse modulation 217Hz, 28V/m
5240 5100-5800 Pulse modulation 217Hz, 9V/m
5500
5785
3. If this monitor is stored near freezing, allow it to acclimate to room tempera-
ture before use.
4. Do not attempt to disassemble this monitor.
5. It is recommended the performance should be checked every 2 years or after
repair. Please contact the service center.
6. Clean the monitor with a dry, soft cloth or a soft cloth squeezed well after
moistened with water, diluted disinfectant alcohol, or diluted detergent.
7. No component can be maintained by user in the monitor. The circuit dia-
grams, component part lists, descriptions, calibration instructions, or other
information which will assist the user’s appropriately qualied technical per-
sonnel to repair those parts of equipment which are designated repairably
can be supplied.
8. The monitor can maintain the safety and performance characteristics for a
minimum of 10,000 measurements or three years, and the cuff integrity is
maintained after 1,000 open–close cycles of the closure.
9. The monitor requires 6 hours to warm from the minimum storage tempera-
ture between uses until the monitor is ready for its INTENDED USE when the
ambient temperature is 20°C.
10. The monitor requires 6 hours to cool from the maximum storage temperature
between uses until the monitor is ready for its INTENDED USE when the
ambient temperature is 20°C.
11. Not servicing/maintenance while the monitor is in use.
2

Hulp nodig? Stel uw vraag in het forum

Spelregels

Misbruik melden

Gebruikershandleiding.com neemt misbruik van zijn services uitermate serieus. U kunt hieronder aangeven waarom deze vraag ongepast is. Wij controleren de vraag en zonodig wordt deze verwijderd.

Product:

Bijvoorbeeld antisemitische inhoud, racistische inhoud, of materiaal dat gewelddadige fysieke handelingen tot gevolg kan hebben.

Bijvoorbeeld een creditcardnummer, een persoonlijk identificatienummer, of een geheim adres. E-mailadressen en volledige namen worden niet als privégegevens beschouwd.

Spelregels forum

Om tot zinvolle vragen te komen hanteren wij de volgende spelregels:

Belangrijk! Als er een antwoord wordt gegeven op uw vraag, dan is het voor de gever van het antwoord nuttig om te weten als u er wel (of niet) mee geholpen bent! Wij vragen u dus ook te reageren op een antwoord.

Belangrijk! Antwoorden worden ook per e-mail naar abonnees gestuurd. Laat uw emailadres achter op deze site, zodat u op de hoogte blijft. U krijgt dan ook andere vragen en antwoorden te zien.

Abonneren

Abonneer u voor het ontvangen van emails voor uw Gima KD-5923 Andon bij:


U ontvangt een email met instructies om u voor één of beide opties in te schrijven.


Ontvang uw handleiding per email

Vul uw emailadres in en ontvang de handleiding van Gima KD-5923 Andon in de taal/talen: Engels als bijlage per email.

De handleiding is 0.66 mb groot.

 

U ontvangt de handleiding per email binnen enkele minuten. Als u geen email heeft ontvangen, dan heeft u waarschijnlijk een verkeerd emailadres ingevuld of is uw mailbox te vol. Daarnaast kan het zijn dat uw internetprovider een maximum heeft aan de grootte per email. Omdat hier een handleiding wordt meegestuurd, kan het voorkomen dat de email groter is dan toegestaan bij uw provider.

Stel vragen via chat aan uw handleiding

Stel uw vraag over deze PDF

loading

Uw handleiding is per email verstuurd. Controleer uw email

Als u niet binnen een kwartier uw email met handleiding ontvangen heeft, kan het zijn dat u een verkeerd emailadres heeft ingevuld of dat uw emailprovider een maximum grootte per email heeft ingesteld die kleiner is dan de grootte van de handleiding.

Er is een email naar u verstuurd om uw inschrijving definitief te maken.

Controleer uw email en volg de aanwijzingen op om uw inschrijving definitief te maken

U heeft geen emailadres opgegeven

Als u de handleiding per email wilt ontvangen, vul dan een geldig emailadres in.

Uw vraag is op deze pagina toegevoegd

Wilt u een email ontvangen bij een antwoord en/of nieuwe vragen? Vul dan hier uw emailadres in.



Info