568658
3
Verklein
Vergroot
Pagina terug
1/4
Pagina verder
8.1 Function button
The function button on the pulse oximeter has three functions in total:
Switch-on function
: When the pulse oximeter is switched o you can hold down the function button to switch it on.
Display function
: To select your desired display format (vertical format, horizontal format), press the function button during operation. There are
four dierent display formats to choose from.
Settings menu function:
To enter the settings menu, you must first set the display format to a horizontal format. To call up the settings menu,
press and hold down the function button during operation. You can set the following parameters in the settings menu: display brightness, alarm set-
tings, activate/deactivate data transfer, data storage and upload data to a PC.
Note
To be able to set the following functions (sections 8.2 to 8.6), you must select a horizontal format as the display format (see 8.1 Function button /
Display function). If the display format is set to a vertical format and you would like to use the following functions, an error message appears on the
display (“Please change direction first!”).
8.2 Display brightness
To set the display brightness, switch on the pulse oximeter and press and hold down the function button. In the settings menu, the menu item
“Brightness” is selected.
There are four dierent brightness levels to choose from. To set the desired display brightness, press and hold down the function button until the
desired level of brightness has been reached.
To exit the settings menu, use the function button to select the “Exit” menu item and confirm by pressing and holding down the function button.
8.3 Alarm settings
Switch on the pulse oximeter and press and hold down the function button. The settings menu appears on the display.
In the settings menu, use the function button to select the “Alarm” menu item and confirm by pressing and holding down the function button.
Use the function button to select the desired parameter and set the desired value by pressing and holding down the function button.
Y
ou can set the following parameters in the alarm menu:
“Dir”
Here you can set whether the setting value runs up or down when setting the alarm limits in the alarm menu. It is nec-
essary to change the setting direction if you would like to move the limits up or down.
“SPO2 ALM HI”
Here you can set an upper limit for oxygen saturation. If, during a measurement, the set limit is exceeded, the satura-
tion value appears yellow and a signal sounds (if the alarm is activated).
“SPO2 ALM LO”
Here you can set a lower limit for oxygen saturation. If, during a measurement, the set limit is undercut, the saturation
value appears yellow and a signal sounds (if the alarm is activated).
“PR ALM HI”
Here you can set an upper limit for the pulse rate. If, during a measurement, the set limit is exceeded, the pulse rate
appears yellow and a signal sounds (if the alarm is activated).
“PR ALM LO”
Here you can set a lower limit for the pulse rate. If, during a measurement, the set limit is undercut, the pulse rate ap-
pears yellow and a signal sounds (if the alarm is activated).
Alarm”
Here you can activate (“on”) or deactivate (“o”) the alarm. If you have activated the alarm and the set upper or lower
limit is exceeded or undercut, a signal sounds.
“Pulse Sound”
Here you can activate (“on”) or deactivate (“o”) the pulse tone. If you have activated the pulse tone, a signal sounds at
every beat during the measurement.
To exit the alarm menu, use the function button to select the “Exit” menu item and confirm by pressing and holding down the function button.
8.4 Displaying measurement values in real time on a PC (“USB”)
Note
You can create a printout at any time during measurement
data recording by using the print function. The printout shows
the trend of the current measurement up to that point.
Note
You can view the measurement data of up to four devices at
once in the software program. In the software program, you
can enlarge the view of the individual measurement data by
selecting the corresponding measurement position number.
Measurement posi-
tion number & view
Trend display time Report
Current
wave form
of the
pulse rate
Trend
Connect the pulse oximeter to your PC using the supplied USB data cable.
Start the “SpO₂-Viewer” program on your PC. If you have not yet installed the program, follow the instructions in section 7.2 (Installing the “SpO₂-
Viewer/Manager” software).
Switch on the pulse oximeter and press and hold down the function button. The settings menu appears on the display.
Select the “USB” menu item using the function button.
Press and hold down the function button. The word “on” is displayed in the “USB” menu item. Insert one finger into the finger opening of the pulse
oximeter. After a few seconds, the measurement data is displayed on your PC in real time.
If you disconnect the data cable from the computer or close the program, an entry mask appears on the PC screen. Here you can enter personal
details and store the current measurement data. As standard, the measurement data is stored under C:/Programs/SpO2/Data.
If you no longer want to display the measurement data on your PC, press and hold down the function button. The word “o” is displayed in the
“USB” menu item.
To exit the settings menu, use the function button to select the “Exit” menu item and confirm by pressing and holding down the function button.
8.5 Recording measurement data
With the pulse oximeter PO 80, you can record your measurement data over a period of up to 24 hours. If required, the measurement data can be
stored on your computer or printed out as a report.
To record the measurement data, follow these steps:
Switch on the pulse oximeter and press and hold down the function button. The settings menu appears on the display.
Select the “Record” menu item using the function button.
Press and hold down the function button. The word “Time” will appear on the display. Here you can enter the current time of day so that, later on in
the PC software program, you are able to read o the time you started your recording.
Time;
hours
Time;
minutes
Start record-
ing
Do not start
recording
To enter the current time of day, press and hold down the function button and confirm your entry by pressing the function button briefly.
Note
If you start a new recording, the previous recording is automatically overwritten and cannot be recalled. The maximum recording duration is 24
hours.
To start the recording, use the function button to select the “Y” on the display and confirm by pressing and holding down the function button. The
pulse oximeter starts recording. If you do not want to start recording, select “N” and confirm by pressing and holding down the function button.
Note
If you have started a recording and return to the display, a red recording symbol appears (REC •). To save battery power, the device automatically
switches to energy-saving mode 30 seconds after the recording starts. If the pulse tone is activated, it is automatically deactivated.
If you press the function button briefly when in energy-saving mode, “Recording” appears on the display for 2 seconds.
If you press and hold down the function button when in energy-saving mode, the normal display appears.
“Memory is full” appears on the display when the pulse oximeter memory is full.
8.6 Uploading measurement data to a computer
Note
You cannot transfer measurement data to the computer while a measurement data recording is underway.
You can upload the measurement data that you previously stored on the pulse oximeter to your computer.
To upload your stored measurement data to your computer, follow these steps:
Connect the pulse oximeter to your PC using the supplied USB data cable.
Start the “SpO₂-Manager” program on your PC.
Options
glass symbole
(New Session)
Trend display time
Report
Click the magnifying glass symbol in lower section of the program. The “New Session” window opens.
Please enter your user data. To confirm the user data, click “ok”. The program now displays that the pulse oximeter is connected to the computer
and that the program is ready to receive the measurement data (“Device connected : wating for data ...”).
Switch on the pulse oximeter.
Press and hold down the function button on the pulse oximeter. The settings menu appears.
Select the “Upload” menu item by pressing the function button briefly.
Briefly press the function button. The word “on” is displayed in the “Upload” menu item. Your measurement data are now transferred to the com-
puter. It is not possible to cancel the transfer of the measurement data manually.
When the transfer of the measurement data is complete, the “Exit” menu item is highlighted automatically. To exit the settings menu, press and hold
down the function button.
As soon as your measurement data has been transferred to the computer, your measurement data appears in the software program.
After the measurement data has been transferred successfully to the PC, you can also print out the data in the software program, if required.
To do so, follow these steps:
Click the print symbol in the bottom right-hand corner of the program. The “Report” window opens.
Note
When printing the measurement data, you can choose between four dierent display
variants (Summary Report, Full Study Report, OxiMetry Report, Strip Chart Report).
Together with the printout, the user is oered an automatic appraisal of the recorded
measurement data. This printout serves exclusively to support the diagnostic findings
of your attending doctor. The printout is not a substitute for a medical certificate and is
not exhaustive!
8.7 Calling up stored measurement data
To call up previously stored measurement data in the software program, follow these steps:
Start the “SpO₂-Manager” program on your PC.
Click the magnifying glass symbol in the lower section of the program. Select “Review” in the upper section of the window and click “Find File” in
the lower section.
Select the file you stored (as standard, the files are located under C:/Programs/SpO2/Data) and click “Open” followed by “Ok”.
Your stored measurement data is now displayed in the software program.
9. Evaluating measurement results
WARNING
The following table for evaluating your measurements does NOT
apply to people with certain pre-existing conditions (e.g. asthma,
heart failure, respiratory diseases) or whilst staying at altitudes
above 1500 metres. If you have a pre-existing condition, always
consult your doctor to evaluate your measurements.
Decline in oxygen saturation depending on altitude
Note
The following table informs you of the eects of various altitudes on oxygen
saturation value and its impact on the human body. The following table does
NOT apply to people with certain pre-existing conditions (e.g. asthma, heart
failure, respiratory diseases etc.). People with pre-existing conditions can show
signs of illness (e.g. hypoxia) at lower altitudes.
SpO₂ (oxygen satura-
tion) measurement
in %
Classification/measures to be
taken
Altitude
Expected SpO₂
value (oxygen satu-
ration) in %
Impact on human body
99-94 Normal range 1500-2500 m > 90 No altitude sickness (normally)
94-90
Decreased range:
visit to the doctor recommended
2500-3500 m ~90
Altitude sickness, acclimatisa-
tion recommended
< 90
Critical range:
seek medical attention urgently
3500-5800 m <90
Very frequent altitude sickness,
acclimatisation absolutely
essential
5800-7500 m <80
Severe hypoxia, only limited
length of stay possible
7500-8850 m <70 Immediate, acute danger to life
Source: Hackett PH, Roach RC: High-Altitude Medicine. In: Auerbach PS (ed):
Wilderness Medicine, 3rd edition; Mosby, St.Louis, MO 1995; 1-37.
10. Maintenance/cleaning
IMPORTANT:
D
o not use high-pressure sterilisation on the pulse oximeter!
Under no circumstances should you hold the pulse oximeter under water, as this can cause liquid to enter and damage the pulse oximeter.
Clean the housing and the interior rubber surface with a soft cloth dampened with medical alcohol after each use.
11. Storage
IMPORTANT:
Store the pulse oximeter in a dry place (relative humidity ≤95 %). If the humidity is too high it may shorten the service life of the pulse oximeter or dam-
age it. Store the pulse oximeter in a place where the ambient temperature is between -40°C and 60°C.
12. Disposal
Batteries must not be disposed of with household waste. As a consumer, you are required by law to recycle used batteries. You can recycle your
old batteries and rechargeable batteries at public collection points in your community, or wherever batteries of the relevant type are sold.
Note
The codes below are printed on batteries containing harmful substances:
Pb = Battery contains lead,
Cd = Battery contains cadmium,
Hg = Battery contains mercury.
For environmental reasons, do not dispose of the devices including the batteries in the household waste at the end of their service life. Dispose of
the unit at a suitable local collection or recycling point. Observe the local regulations for material disposal.
Please dispose of the device in accordance with EC Directive – WEEE (Waste Electrical and Electronic Equipment). If you have any queries,
please contact the appropriate local authorities.
13. What if there are problems?
Problem Possible cause Solution
The pulse oximeter is not displaying meas-
urement values
The pulse oximeter has run out of battery. Charge the battery via the USB connection.
The pulse oximeter is displaying measure-
ment interruptions or high measurement
value jumps
Insucient circulation in the measurement
finger
Observe the warnings and safety notes in
section 5
Measurement finger is too large or too small
Fingertip must have the following measure-
ments: Width between 10 and 22 mm
Thickness between 5 and 15 mm
Finger, hand or body is moving
Keep your finger, hand and body still during
the measurement.
Cardiac arrhythmia Seek medical attention
14. Technical Data
Model no. PO 80
Measurement method Non-invasive measurement of arterial oxygen saturation of haemoglobin and pulse rate in finger
Measurement range SpO₂ 0 100%,
Pulse 0 254 beats /minute
Accuracy SpO₂ 70 100%, ± 2%,
Pulse 30-250 bpm, ± 2 beats /minute
Dimensions L 57 mm x W 32 mm x H 30 mm
Weight Approx. 42 g
Sensor to measure SpO₂ Red light (wave length 660 nm); infra-red (wave length 905 nm); silicon receiver diode
Permissible operating conditions +10 °C to +40 °C, ≤75 % relative humidity, 700 –1060 hPa ambient pressure
Permissible storage conditions -40 °C to +60 °C, ≤95 % relative humidity, 500 –1060 hPa ambient pressure
Power supply
Integrated, rechargeable lithium battery, 500mAh / 3.7 V
Classification IP22, application part, type BF
System requirements for software Supported operating systems: Windows XP, Windows Vista and Windows 7
Technical information is subject to change without notification to allow for updates.
This device complies with European Standard EN60601-1-2 and is subject to particular precautions with regard to electromagnetic compatibility.
Please note that portable and mobile HF communication systems may interfere with this unit. More details can be requested from the stated Cus-
tomer Service address or found at the end of the instructions for use.
This device complies with the EU Directive 93/42/EC concerning medical devices, the Medizinproduktegesetz (German Medical Devices Act) and
the DIN EN ISO 9919 standard (Medical electrical equipment – Particular requirements for the basic safety and essential performance of pulse
oximeter equipment for medical use)
Guidance and manufacture’s declaration-electromagnetic emission
for all EQUIPMENT and SYSTEMS
Guidance and manufacture’s declaration –electromagnetic emission
The PO80 Pulse Oximeter is intended for use in the electromagnetic environment specified below. The customer or the user of the PO80 Pulse
Oximeter should assure that it is used in such an environment.
Emission test compliance Electromagnetic environment-guidance
RF emissions
CISPR 11
Group 1 The PO80 Pulse Oximeter uses RF energy only for its internal func-
tion. Therefore, its RF emissions are very low and are not likely to
cause any interference in nearby electronic equipment.
RF emissions
CISPR 11
Class B The PO80 Pulse Oximeter is suitable for use in all establishments,
including domestic establishments and those directly connected to
the public low-voltage power supply network that supplies buildings
used for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Not applicable
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Not applicable
Guidance and manufacture’s declaration-electromagnetic immunity
for all EQUIPMENT and SYSTEMS
Guidance and manufacture’s declaration-electromagnetic immunity
The PO80 Pulse Oximeter is intended for use in the electromagnetic environment specified below. The customer or the user of PO80 Pulse
Oximeter should assure that it is used in such an environment.
Immunity test IEC60601 test level Compliance level Electromagnetic environ-
ment-guidance
Electrostatic discharge (ESD) IEC 61000-4-2 ±6kV contact
±8kV air
±6kV contact
±6kV air
Floors should be wood, con-
crete or ceramic tile. If floor are
covered with synthetic material,
the relative humidity should be
at least 30%. The manufacturer
may recommend the ESD pre-
cautionary procedures to user.
Power frequency (50 / 60Hz) magnetic field
IEC 61000-4-8
3 A/m 3 A/m Power frequency magnetic fields
should be at levels characte-
ristic of a typical location in a
typical commercial or hospital
environment.
3

Hulp nodig? Stel uw vraag in het forum

Spelregels
1

Forum

Beurer-PO-80

Zoeken resetten

  • Heb een beurre PO 80.
    Bij gebruik van het apparaat knippert op het display een R met rode punt. Wat betekend dit. Gesteld op 28-1-2023 om 14:25

    Reageer op deze vraag Misbruik melden

Misbruik melden

Gebruikershandleiding.com neemt misbruik van zijn services uitermate serieus. U kunt hieronder aangeven waarom deze vraag ongepast is. Wij controleren de vraag en zonodig wordt deze verwijderd.

Product:

Bijvoorbeeld antisemitische inhoud, racistische inhoud, of materiaal dat gewelddadige fysieke handelingen tot gevolg kan hebben.

Bijvoorbeeld een creditcardnummer, een persoonlijk identificatienummer, of een geheim adres. E-mailadressen en volledige namen worden niet als privégegevens beschouwd.

Spelregels forum

Om tot zinvolle vragen te komen hanteren wij de volgende spelregels:

Belangrijk! Als er een antwoord wordt gegeven op uw vraag, dan is het voor de gever van het antwoord nuttig om te weten als u er wel (of niet) mee geholpen bent! Wij vragen u dus ook te reageren op een antwoord.

Belangrijk! Antwoorden worden ook per e-mail naar abonnees gestuurd. Laat uw emailadres achter op deze site, zodat u op de hoogte blijft. U krijgt dan ook andere vragen en antwoorden te zien.

Abonneren

Abonneer u voor het ontvangen van emails voor uw Beurer PO 80 bij:


U ontvangt een email met instructies om u voor één of beide opties in te schrijven.


Ontvang uw handleiding per email

Vul uw emailadres in en ontvang de handleiding van Beurer PO 80 in de taal/talen: Duits, Engels als bijlage per email.

De handleiding is 4,02 mb groot.

 

U ontvangt de handleiding per email binnen enkele minuten. Als u geen email heeft ontvangen, dan heeft u waarschijnlijk een verkeerd emailadres ingevuld of is uw mailbox te vol. Daarnaast kan het zijn dat uw internetprovider een maximum heeft aan de grootte per email. Omdat hier een handleiding wordt meegestuurd, kan het voorkomen dat de email groter is dan toegestaan bij uw provider.

Stel vragen via chat aan uw handleiding

Stel uw vraag over deze PDF

loading

Andere handleiding(en) van Beurer PO 80

Beurer PO 80 Gebruiksaanwijzing - Nederlands - 32 pagina's

Beurer PO 80 Gebruiksaanwijzing - Français - 32 pagina's

Beurer PO 80 Gebruiksaanwijzing - Italiano - 32 pagina's

Beurer PO 80 Gebruiksaanwijzing - Espanõl - 32 pagina's


Uw handleiding is per email verstuurd. Controleer uw email

Als u niet binnen een kwartier uw email met handleiding ontvangen heeft, kan het zijn dat u een verkeerd emailadres heeft ingevuld of dat uw emailprovider een maximum grootte per email heeft ingesteld die kleiner is dan de grootte van de handleiding.

Er is een email naar u verstuurd om uw inschrijving definitief te maken.

Controleer uw email en volg de aanwijzingen op om uw inschrijving definitief te maken

U heeft geen emailadres opgegeven

Als u de handleiding per email wilt ontvangen, vul dan een geldig emailadres in.

Uw vraag is op deze pagina toegevoegd

Wilt u een email ontvangen bij een antwoord en/of nieuwe vragen? Vul dan hier uw emailadres in.



Info