ARGUS
ARGUS 41
PLUS
1
ARGUS 41
PLUS
Manual
Version: 1.30/EN
ARGUS
2
ARGUS 41
PLUS
by intec Gesellschaft für Informationstechnik mbH
D-58507 Lüdenscheid, Germany, 2008
Alle Rechte, auch der Übersetzung, vorberhalten. Kein Teil des Werkes darf in irgendeiner
Form (Druck, Fotokopie, Mikrofilm oder einem anderen Verfahren) ohne schriftliche
Genehmigung reproduziert, vervielfältigt oder verbreitet werden.
All rights are reserved. No portion of this document may be reproduced, duplicated or
distributed in any form (print, copies, microfilm or on any other media) without intec’s
written permission.
Version: 1.30/EN
ARGUS
ARGUS 41
PLUS
3
1 Introduction ........................................................................................ 5
2 Safety Instructions ............................................................................. 7
3 Technical data .................................................................................... 8
4 Operation - a brief guide .................................................................... 9
5 Accu charge ...................................................................................... 12
6 Voltage measurement after being switched on ............................. 13
7 ARGUS Configuration ...................................................................... 15
7.1 Connection of the ARGUS to a PC .................................................. 15
7.2 Configurig the ARGUS ..................................................................... 17
7.2.1 Setting the profile parameters ................................................... 17
7.2.2 Setting the device parameters ................................................... 25
7.2.3 Setting the ADSL mode ............................................................. 29
7.3 Resetting all of the parameters on the ARGUS ............................. 30
7.4 Setting the contrast of the ARGUS display .................................... 30
7.5 Enabling a software option ............................................................. 31
8 Operation on ADSL, ADSL2 and ADSL2+ Accesses ..................... 33
8.1 Determining the ADSL connection parameter ............................... 33
8.2 Ping test ............................................................................................ 36
8.3 Bridge mode (optional) .................................................................... 39
8.4 Displaying the test results on the PC ............................................. 41
9 Appendix ........................................................................................... 45
A) ADSL Acronyms ................................................................................ 45
B) Error message: ADSL connection ..................................................... 47
C) Error message: PPP connection ....................................................... 48
D) General Error Messages ................................................................... 50
E) ARGUS Messages - Script Errors ..................................................... 51
F) ARGUS Messages - Routing Rules .................................................. 51
ARGUS
4
ARGUS 41
PLUS
ARGUS
ARGUS 41
PLUS
5
1 Introduction
The ARGUS 41 PLUS provides the installer or service technician with confirmation that the
desired ADSL service (ADSL , ADSL2 or ADSL2+ ) is properly supplied to the subscriber.
By determining the ADSL connection rates and the signal-to-noise ratio of the ATU-C, the
ARGUS 41 PLUS evaluates the quality of the respective service at the customer's site
within a matter of seconds.
By connecting the battery operated ARGUS 41 PLUS to various points in the network (i.e.
to the endpoint or before or after the splitter), the technician can determine whether the
cabling is adequate for the DSL service or whether it should be further improved.
An overview of some important ARGUS functions:
Voltage measurement
Determines the connection parameters
Displays the connection's most important upstream/downstream parameters.
- maximum ATM bitrates
- fast or interleaved ATM bitrate
- signal-to-noise ratio
- output power
- line attenuation
Reads the error counters
Displays the upstream/downstream ATM cell errors and bit error statistics
- Cyclic Redundancy Check (CRC)
- Forward Error Correction (FEC)
- Header Error Checksum (HEC)
- Loss of Cell Delineation (LCD)
- Loss of Signal (LOS)
- No Cell Delineation (NCD)
Ping test
Bridge mode (optional)
Configuration of the ARGUS 41 PLUS from a PC via a web interface
ARGUS
6
ARGUS 41
PLUS
Should you have any further questions, please contact us:
intec Gesellschaft für Informationstechnik mbH
Rahmedestr. 90
D-58507 denscheid
Tel.: +49 (0) 2351 / 9070-0
Fax: +49 (0) 2351 / 9070-70
www.argus.info
support@argus.info
ARGUS
ARGUS 41
PLUS
7
2 Safety Instructions
ARGUS ISDN-Testers may only be used with the included accessories. Usage of other
accessories may lead to erroneous measurements and could even damage the ARGUS
and the connected installation. The ARGUS is only to be used in accordance with the
instructions in this documentation. Any other usage may result in bodily injury and
destruction of the ARGUS.
Return and environmentally acceptable disposal
The RoHS (EU Directive on the “Restriction of Hazardous Substances”) guidelines, which
restrict the use of certain hazardous substances in electrical and electronic equipment,
apply in eight of the ten categories of the WEEE (EU Directive on “Waste Electrical and
Electronic Equipment”) guidelines.
The ARGUS products fall into category 9 and are thus not subject to the RoHS guidelines.
Nonetheless, we have decided to voluntarily ensure that ARGUS products built since 1
January 2007 satisfy all of the RoHS guidelines.
Since October 2005 in compliance with WEEE (EU Directive on Waste of Electrical and
Electronic Equipment) 2002/96/EU and the German Electrical and Electronic Equipment
Act (ElektroG - Elektro- und Elektronikgerätegesetz), we have begun marking our testers
with the following symbol
( ) (DIN EN 50419) .
In other words, the ARGUS may not be disposed of in the household waste.
Regarding the return of old equipment, please contact our Service department.
- To prevent electrical shocks or damage to the ARGUS, do not connect it to lines
with voltages in excess of 100 V!
- Never attempt a measurement with the case open!
- The ARGUS is not watertight. Protect the ARGUS from exposure to water!
- Before replacing the battery (see page 11 Replacing the accumulators), disconnect
all the test leads and switch the ARGUS off.
- Make certain that the polarity is correct when connecting the accumulators!
- Unplug the power supply from the mains, as soon as the ARGUS is no longer in
use and switched off.
ARGUS
8
ARGUS 41
PLUS
3 Technical data
Dimensions and Weight
Height 229 mm, width 72 mm, depth 35
mm
Weight 350 g (without accumulators and
protective case)
Keypad
21 Keys
LCD display
LCD display with switchable
background lighting
4 lines with 16 characters
Memory
EEPROM constants memory: 16 Kbytes
Flash program memory: 10 Mbyte
SDRAM: 16 Mbyte
Inputs / Outputs
1 RJ45 for ADSL
1 jack for an external power supply
1 RJ-11 for the serial interface (optional)
1 RJ-45 10/100 BaseT Ethernet
Temperature Ranges
Operating Temperature: 0 °C to +50 °C
Storage temperature: -15 °C to +70 °C
Power Supply
NiMH accumulators or
9 V / 560mA plug-in switching power supply
ARGUS
ARGUS 41
PLUS
9
4 Operation - a brief guide
Power key:
-
-
-
-
Switch the ARGUS ON
To start up again after a power down
To switch on the display backlighting
In battery mode to save power, the backlighting will switch off
automatically after 5 seconds.
To switch the ARGUS OFF: this must be pressed somewhat longer
If the ARGUS is connected to its power supply, it will automatically
charge its accumulators when switched off (see Page 12).
Arrow keys:
- Scroll through the displayed ADSL connection parameters
LEDs
Softkeys
Arrow keys
Level key
Power / Display backlighting
Fastener for shoulder strap
Numerical keypad
LCD Display: 4x16 characters
ARGUS
10
ARGUS 41
PLUS
The ARGUS is in largest part operated with the three softkeys.
On the following pages, only the softkey's meaning in the respective context is shown -
enclosed in brackets< >, e.g. < START >.
The softkey <
> serves the same function as the arrow key on the ARGUS keypad.
Connections at bottom:
Level key:
- Display the ADSL connection parameter
Softkeys:
- The function of the 3 softkeys varies with the situation. The current
function of each softkey is shown in the bottom, highlighted line of the
display.
PWR
Connection for the external plug-in power supply.
If the plug-in power supply is connected, the ARGUS
will disconnect the accumulators and, when it is
switched off, the ARGUS will automatically recharge
the accumulators (see Page 12).
Line
Connection for the ADSL network
( PIN assignment 7/8 in Germany
7/8 and 4/5 international)
Ser. (optional)
Serial interface to connect a PC
LAN
10/100 Base-T Ethernet interface to connect the
ARGUS to the Ethernet jack on a PC.
ARGUS
ARGUS 41
PLUS
11
Replacing the accumulators
The battery compartment for the three accumulators (rechargeable batteries) is located on
the back of the case. Unscrew the screws to remove the cover of the case and insert the
accumulators in accordance with the polarity marking. Use only the accumulators included
in the package. The current state of the charge will (if the ARGUS is not connected to the
power supply) be displayed graphically. In the LCD display, a battery symbol will begin to
blink, when there is still approximately (depending on the mode of operation) 15 minutes
reserve. During this period, it is possible that there may be audible interference and in rare
cases even malfunctions.
Power Down
In accu/battery operation, if the ARGUS is idle for 5 minutes, it will automatically
switch to the power-down mode (power-down ). The ARGUS will remain in power-
down mode until the Power-Key is pressed again.
As an alternative, it is possible to operate the ARGUS using the included power supply. If
the ARGUS is connected to the plug-in power supply, it will automatically disconnect the
accumulators and will not enter power-down mode.
Unplug the power supply from the mains, once the ARGUS is switched off and will
no longer be used.
You should always operate the ARGUS with accumulators installed. This will ensure the
uninterrupted operation of the real-time clock.
ARGUS
12
ARGUS 41
PLUS
5 Accu charge
Automatic recharging of the accumulators when switched on (trickle charge)
The ARGUS automatically recharges the
accumulators (also shortened to “accus”), if
the ARGUS is connected to the plug-in
power supply and switched on and the
accumulator voltage is too low (only use the
supplied accumulators).
The ARGUS will display the level of charge
and the voltage while charging the
accumulators.
Automatic recharging of the accumulators when the ARGUS is switched on
The ARGUS automatically recharges the accumulators (also
shortened to “accus”), if the ARGUS is connected to the
plug-in power supply and switched off and the accumulator
voltage is too low (only use the supplied accumulators).
While charging, the ARGUS displays the message Charge
accu”. Otherwise, the ARGUS will switch itself off
automatically as soon as the accumulators are recharged. If
you press and hold the power switch, the ARGUS will switch
off before the accumulators are recharged.
<DISC.> The accumulators will first be
discharged (this can take up to
7 hours) and will then, after a
cool down period of about 30
minutes, be automatically
recharged.
<NO> ARGUS will open the "Voltage"
display (see Page 13) without
recharging the accumulators.
< ABORT> The download is stopped.
<CONTI.> ARGUS will open the "Voltage" display (see
Page 13) without recharging the
accumulators.
Press the
<LOAD> softkey
Level of
charge
of the accus
ARGUS
ARGUS 41
PLUS
13
6 Voltage measurement after being switched on
Using the ADSL 2-wire connection cable, the ARGUS is connected directly to the ADSL
access (either before or after the splitter). In this case, the ARGUS replaces both the
modem and the PC.
Use only the cables included in the package.
Press the Power key to switch the ARGUS on. The ARGUS
will display the DC voltage on the access interface.
Evaluating the voltage measurement when the ARGUS is
connected ahead of the splitter:
It is not possible to evaluate a voltage measurement made
with the ARGUS connected behind the splitter:
0 Volts: No definite assessment possible.
approx.90-96
Volts:
ISDN access
less than or
equal to 63.5
Volts:
POTS access
approx. 63-89
Volts:
No definite assessment possible:
In Germany there are still some older
accesses, which range in voltage from 60
V to 90 V, that can be used for ISDN (U-
interface) or POTS.
0 Volts: ISDN or POTS access
Black 2-wire DSL connection cable
Please note the labels!
Jack
Line
Label
ADSL
Label
ARGUS
ADSL access
ATU-C
ARGUS
14
ARGUS 41
PLUS
7 ARGUS Configuration
ARGUS 41
PLUS
15
7 ARGUS Configuration
7.1 Connection of the ARGUS to a PC
The ARGUS is configured from a PC with the aid of a web interface. This same setup is
also used to display the test results.
PC settings:
First start Windows on your PC. The
following steps describe the
procedure when the PC is running
Windows XP Professional.
In the taskbar, click on the "Start"
button (generally at the lower left-
hand corner of the screeen) and then
click on "Control Panel" to open it.
In the "Control Panel" window,
double-click on "Network
Connections" to open it and then, in
the list, select "LAN Connections".
Next right click on "LAN Connections"
and then select "Properties" from the
now open list. The window at the side
will now open.
Select "Internet Protocol (TCP/IP)".
Click on the "Properties" button.
In the now open window, select
"Obtain an IP address automatically"
and "Obtain DNS server address
automatically".
Click on the "OK" button.
7 ARGUS Configuration
16
ARGUS 41
PLUS
As it comes from the factory, the ARGUS is set to "DHCP server mode" and uses
the IP address 192.168.10.1 on the LAN side.
When the ARGUS is first used, it should not be connected to a network via a router
or switch, since in "DHCP Server mode", the ARGUS will assign IP addresses to
any network devices which are in "DHCP Client mode". Furthermore, it can cause
conflicts with any other device in the network which is also in "DHCP Server mode".
Using a PC and the web interface, the ARGUS LAN settings can be changed as
needed to suit your network infrastructure (see Page 25).
7 ARGUS Configuration
ARGUS 41
PLUS
17
7.2 Configurig the ARGUS
To connect the ARGUS to the PC, plug one end of the included patch cable (twisted pair)
into the "LAN" jack on the ARGUS and the other end into the PC's Ethernet jack. Switch the
ARGUS on and the "LAN" LED will light.
Start any common Internet browser and enter the following address http://192.168.10.1.
The web interface will open.
7.2.1 Setting the profile parameters
The ARGUS saves all of the parameters required for setting up a PPP connection and
running the Ping test in profiles. Using the web interface, you can configure up to 10 user-
defined profiles and send them to the ARGUS. Before a PPP connection can be set up, you
must first select a profile on the ARGUS 41 PLUS (see Page 36).
First click on "Settings"
and then on "Profiles".
Web interface
7 ARGUS Configuration
18
ARGUS 41
PLUS
Send profile to the
ARGUS.
All of the profiles
that are
checkmarked as
available will also
be available for
use on the
ARGUS later.
(see Page 36).
Click on a profile to open it to view or
edit its parameters.
First click on "Line" and then
select, e.g. "ATM" to view the ATM
parameters.
Click on the radio button in front of
the profile that you wish make the
"Active Profile" (default profile).
The "Active Profile" will be
automatically set as "available".
The ARGUS will then suggest the
"Active Profile" as the default
profile for any connection or test.
(see Page 36).
7 ARGUS Configuration
ARGUS 41
PLUS
19
7.2.1.1 Setting the ATM parameters
Save the Profile settings
and send them to the ARGUS.
Enter the virtual channel in the
ATM cells
Default values: VPI = 1, VCI = 32
Set the encapsulation of the
packets to be sent
(LLC or VC-MUX)
In the "Line" submenu,
click on ATM”.
7 ARGUS Configuration
20
ARGUS 41
PLUS
7.2.1.2 Setting the protocol
Save the Profile settings
and send them to the ARGUS.
Select the transfer protocol.
In the "Line" submenu,
click on “Protocol”.
7 ARGUS Configuration
ARGUS 41
PLUS
21
7.2.1.3 Setting the PPP parameters
In the "Line" submenu,
click on “PPP”.
Enter the user name
and password
(assigned by the
network operator)
for the Internet
connection.
Save the Profile settings
and send them to the ARGUS.
21

Hulp nodig? Stel uw vraag in het forum

Spelregels

Misbruik melden

Gebruikershandleiding.com neemt misbruik van zijn services uitermate serieus. U kunt hieronder aangeven waarom deze vraag ongepast is. Wij controleren de vraag en zonodig wordt deze verwijderd.

Product:

Bijvoorbeeld antisemitische inhoud, racistische inhoud, of materiaal dat gewelddadige fysieke handelingen tot gevolg kan hebben.

Bijvoorbeeld een creditcardnummer, een persoonlijk identificatienummer, of een geheim adres. E-mailadressen en volledige namen worden niet als privégegevens beschouwd.

Spelregels forum

Om tot zinvolle vragen te komen hanteren wij de volgende spelregels:

Belangrijk! Als er een antwoord wordt gegeven op uw vraag, dan is het voor de gever van het antwoord nuttig om te weten als u er wel (of niet) mee geholpen bent! Wij vragen u dus ook te reageren op een antwoord.

Belangrijk! Antwoorden worden ook per e-mail naar abonnees gestuurd. Laat uw emailadres achter op deze site, zodat u op de hoogte blijft. U krijgt dan ook andere vragen en antwoorden te zien.

Abonneren

Abonneer u voor het ontvangen van emails voor uw Argus 41 Plus bij:


U ontvangt een email met instructies om u voor één of beide opties in te schrijven.


Ontvang uw handleiding per email

Vul uw emailadres in en ontvang de handleiding van Argus 41 Plus in de taal/talen: Engels als bijlage per email.

De handleiding is 1,49 mb groot.

 

U ontvangt de handleiding per email binnen enkele minuten. Als u geen email heeft ontvangen, dan heeft u waarschijnlijk een verkeerd emailadres ingevuld of is uw mailbox te vol. Daarnaast kan het zijn dat uw internetprovider een maximum heeft aan de grootte per email. Omdat hier een handleiding wordt meegestuurd, kan het voorkomen dat de email groter is dan toegestaan bij uw provider.

Stel vragen via chat aan uw handleiding

Stel uw vraag over deze PDF

Uw handleiding is per email verstuurd. Controleer uw email

Als u niet binnen een kwartier uw email met handleiding ontvangen heeft, kan het zijn dat u een verkeerd emailadres heeft ingevuld of dat uw emailprovider een maximum grootte per email heeft ingesteld die kleiner is dan de grootte van de handleiding.

Er is een email naar u verstuurd om uw inschrijving definitief te maken.

Controleer uw email en volg de aanwijzingen op om uw inschrijving definitief te maken

U heeft geen emailadres opgegeven

Als u de handleiding per email wilt ontvangen, vul dan een geldig emailadres in.

Uw vraag is op deze pagina toegevoegd

Wilt u een email ontvangen bij een antwoord en/of nieuwe vragen? Vul dan hier uw emailadres in.



Info