704181
14
Verklein
Vergroot
Pagina terug
1/15
Pagina verder
Digital Blood Pressure Monitor
Model UA-704
Instruction Manual
Original
Manuel d’instructions
Traduction
Manual de Instrucciones
Traducción
Manuale di Istruzioni
Traduzione
使用手冊
翻譯
1WMPD4000735E
English 1
Contents
Dear Customers
...............................................................................
1
Preliminary Remarks
........................................................................
1
Precautions
.......................................................................................
2
Parts Identification
..........................................................................
3
Symbols .................................................................................. 4
Using The Monitor
...........................................................................
5
Installing / Changing The Battery ........................................ 5
Selecting The Correct Cuff ................................................... 6
Applying The Arm Cuff ........................................................ 7
How To Take Proper Measurements ..................................... 7
Measurement
...............................................................................
7
After Measurement
......................................................................
7
Measurement
....................................................................................
8
Notes For Proper Measurement
..................................................
9
What Is An Irregular Heartbeat?
....................................................
10
About Blood Pressure
....................................................................
10
What Is Blood Pressure?
............................................................
10
What Is Hypertension And How Is It Controlled?
....................
10
Why Measure Blood Pressure At Home?
..................................
11
WHO Blood Pressure Classification
..........................................
11
Blood Pressure Variations
.........................................................
11
Troubleshooting
............................................................................
12
Maintenance
...................................................................................
12
Te
chnical Data
...............................................................................
13
Dear Customers
Congratulations on purchasing a state-of-the-art A&D blood pressure
monitor, one of the most advanced monitors available today. Designed for
ease of use and accuracy, this device will facilitate your daily blood
pressure regimen.
We recommend that you read through this manual carefully before
using the device for the first time.
Preliminary Remarks
This device conforms to the European Directive 93/42 EEC for Medical
Products. This is made evident by the
mark of conformity.
(0123: The reference number to the involved notified body)
The device is designed for use on adults only, not newborns or infants.
Environment for use: The device is for use in the home healthcare
environment.
This device is designed to measure blood pressure and pulse rate of
people for diagnosis.
English 2
Precautions
Precision components are used in the construction of this device.
Extremes in temperature, humidity, direct sunlight, shock or dust should
be avoided.
Clean the device and cuff with a dry, soft cloth or a cloth dampened with
water and a neutral detergent. Never use alcohol, benzene, thinner or
other harsh chemicals to clean the device or cuff.
Avoid tightly folding the cuff or storing the hose tightly twisted for long
periods, as such treatment may shorten the life of the components.
The device and cuff are not water-resistant. Prevent rain, sweat and
water from soiling the device and cuff.
Measurements may be distorted if the device is used close to television,
microwave oven, cellular telephone, X-ray or other devices with strong
electrical fields.
Used equipment, parts and battery are not treated as ordinary household
waste, and must be disposed of according to the applicable local
regulations.
When reusing the device, confirm that the device is clean.
Do not modify the device. It may cause accidents or damage to the
device.
To measure blood pressure, the arm must be squeezed by the cuff hard
enough to temporarily stop blood flow through the artery. This may cause
pain, numbness or a temporary red mark to the arm. This condition will
appear especially when measurement is repeated successively. Any
pain, numbness, or red marks will disappear with time.
Contraindications
The following are precautions for proper use of the device.
Do not apply the cuff on an arm with another medical electrical
equipment attached. The equipment may not function properly.
People who have a severe circulatory deficit in the arm must consult a
doctor before using the device, to avoid medical problems.
Do not self-diagnose the measurement results and start treatment by
yourself. Always consult your doctor for evaluation of the results and
treatment.
Do not apply the cuff on an arm with an unhealed wound.
Do not apply the cuff on an arm receiving an intravenous drip or blood
transfusion. It may cause injury or accidents.
Do not use the device where flammable gases such as anesthetic
gases are present. It may cause an explosion.
Do not use the device in highly concentrated oxygen environments,
such as a high-pressure oxygen chamber or an oxygen tent. It may
cause a fire or explosion.
English 3
Parts Identification
Display Parts
Pulse rate
Exhaust mark. Exhaust air from the cuff
Heart mark. Measurement in progress.
Systolic pressure
Diastolic pressure
I.H.B. indicator
(Irregular heartbeat indicator)
Pressure mark. Pump air into the cuff.
Battery indicator
Start button
Cuff
Exhaust valve
Rubber bulb
Display
Battery
R6P, LR6 or AA
Battery cover
Battery compartment
A
ir hose
Proper fit range
Index mark
A
rtery position mark
English 4
Symbols
Symbols Function / Meaning Recommended Action
Turns the device on and off.
Guide to install battery
Type BF: Device, cuff and tubing are
designed to provide special protection
against electrical shocks.
The indicator that appears while
measurement is in progress. It flashes
when the pulse is detected.
Measurement is in
progress -- remain still.
Irregular Heartbeat indicator. (I.H.B.)
The indicator that appears when an
irregular heartbeat or any excessive
body movement is detected during the
measurement.
Flashes when there is air remaining in
the cuff at power up.
Flashes at the completion of the
measurement until exhaust is complete.
Press the exhaust valve
button to exhaust air from
the cuff.
Flashes when the pressure is not
enough for the measurement.
Pump air into the cuff with
the rubber bulb.
Full Battery
The battery power indicator during
measurement.
low Battery
The battery is low when it blinks.
Replace the battery with
a new one, when the
indicator blinks.
Unstable blood pressure due to
movement during the measurement.
Try the measuring again.
Remain very still during
the measurement.
The systolic and diastolic values are
within 10 mmHg of each other.
The pressure value did not increase
during inflation.
The cuff is not fastened correctly.
The pulse is not detected correctly.
Fasten the cuff correctly,
and try the measurement
again.
EC directive medical device label
EU-representative
Manufacturer
Date of manufacture
WEEE label
Refer to instruction manual/booklet
SN
Serial number
English 5
Using The Monitor
Installing / Changing The Battery
1. Remove the battery cover.
2. Insert a new battery into the
battery compartment as shown,
taking care that the polarities
(+) and (-) are correct.
Use only R6P, LR6, AA or
equivalent battery.
3. Close the battery cover.
CAUTION
Insert the battery as shown in the battery compartment. If not, the
device will not work.
When
(LOW BATTERY mark)
blinks in the display, replace with a new
battery.
(LOW BATTERY mark)
does not appear when the battery is drained.
Battery life varies with the ambient temperature and may be shorter at
low temperatures.
Remove the battery if the device is not to be used for a long time.
The battery may leak and cause a malfunction.
Use the specified battery only. The battery provided with the device is
for testing monitor performance and may have a limited life.
Battery cover
Battery
English 6
Using The Monitor
Selecting The Correct Cuff
Using the correct cuff size is important for an accurate reading. If the cuff is
not the proper size, the reading may yield an incorrect blood pressure
value.
The arm size is printed on each cuff.
The index and proper fit range, on the cuff, tell you if you are applying
the correct cuff. (Refer to "Applying The Arm Cuff" on next page)
If the index points outside of the range, contact your local dealer to
purchase a replacement cuff.
The arm cuff is a consumable. If it becomes worn, purchase a new one.
Arm Size Recommended Cuff Size Catalog Number
32 cm to 45 cm Large adult cuff CUF-D-LA
22 cm to 32 cm Adult cuff CUF-D-A
Arm size: The circumference at the biceps.
Note:
Model UA-704 is not designed for using a small cuff.
Symbols printed on the cuff.
Symbols
Function / Meaning Recommended Action
Artery position mark
Set the
mark on the artery of
the upper arm or in line with
the ring finger on the inside of
the arm.
Index
REF
Catalog number
Proper fit range for the adult cuff.
It's printed on the adult cuff.
Range to use the large adult cuff.
Over range printed on the adult cuff.
Use the large adult cuff
instead of the adult cuff.
Proper fit range for the large adult
cuff.
It's printed on the Large adult cuff.
Under range printed on the adult
cuff.
Range to use the adult cuff.
It's printed on the large adult cuff.
Use the adult cuff instead of
the large adult cuff.
Lot number
Large adult cuff
Adult cuff
Proper fit range
Proper fit range
English 7
Using The Monitor
Applying The Arm Cuff
1. Wrap the cuff around the upper arm,
about 1 to 2 cm above the inside of
the elbow, as shown.
Place the cuff directly against the
skin, as clothing may cause a faint
pulse and result in a measurement
error.
2. Constriction of the upper arm, caused
by rolling up a shirtsleeve, may
prevent accurate readings.
3. Confirm that the index
points
within the proper fit range.
How To Take Proper Measurements
For the most accurate blood
pressure measurement:
Sit comfortably on a chair. Rest your arm on the table. Do not cross
your legs. Keep your feet on the floor and straighten your back.
Relax for about five to ten minutes before measurement.
Place the center of the cuff at the same height as your heart.
Remain still and keep quiet during measurement.
Do not measure immediately after physical exercise or a bath. Rest for
twenty or thirty minutes before taking the measurement.
Try to measure your blood pressure at the same time every day.
Measurement
During measurement, it is normal for the cuff to feel very tight.
Measure your blood pressure according to the section "
Measurement".
After Measurement
Press the exhaust valve to exhaust air from the cuff.
Press the
START button to turn off the power after the measurement.
Remove the cuff and record your data.
Note:
The device has an automatic power shut-off function, which turns off
the power automatically one minute after measurement.
Allow at least three minutes between measurements on the same
person.
Do not roll up shirtsleeve tightly
1 to 2 cm
A
ir hose
Cuff
Index
and Proper fit range
A
rtery position mark
Secure the fabric
fastener firmly
English 8
Measurement
1. Place the cuff on the arm (preferably the left arm).
2. Press the
START button.
When the START button is pressed, all of the
display symbols will display for about one
second.
When the " 0 " is flashing, the device is ready for
measurement. If the
mark is displayed, the
cuff has some air trapped in it. Press the
exhaust valve button until the
mark is
displayed.
3. Pressurize the cuff by squeezing
the rubber bulb.
The figure on the left of the
display shows the current
pressure as you inflate the
cuff.
Stop squeezing the rubber
bulb when the
mark
disappears.
4. When pressurization is
completed, the automatic
exhaust mechanism will
gradually reduce the pressure in
the cuff and the
mark will
display along with the current
pressure reading on the left.
Remain still during the
measurement. When a pulse
is detected, the
mark will
flash in unison with the pulse
of the person being
monitored.
5. When the measurement is
complete, the
mark is
displayed. The systolic pressure
is displayed on the left. The
diastolic is displayed on the
right. The pulse rate alternates
with the systolic and diastolic
displays.
START button
Stop squeezing the
bulb when the
mark disappears.
Detecting the pluse,
the mark is flashed.
Flashing
Systolic pressure
Diastolic pressure
A
lternating display
Pulse rate
Squeeze bulb
Current pressure.
Automatic air release.
Gradually being reduced.
Measurement in
Bulb
To next page
English 9
Measurement
6. Press the exhaust valve button to completely
release air from the cuff.
When a measurement is made with insufficient
pressure, the
mark will be displayed.
Re-pressurize the cuff to a pressure. Stop
squeezing the rubber bulb when the
mark
disappears.
When a measurement is made with an erratic
pulse or in a very noisy environment,
will
display.
7. Turning the power off
Press the
START button to turn the power off.
The device will be in standby mode.
8. Subsequent measurements
If a subsequent measurement is required, turn off the power and turn it
on again. When "0 " is displayed, the device is ready for another
measurement.
Note:
Allow at least three minutes between measurements on the same
person.
9. Automatic power-off function
If the device is left on after a measurement, it will turn off
automatically after about one minute. It may be turned off at any time
by pressing the
START button.
Notes For Proper Measurement
Sit down in a comfortable position. Place the arm to be used for the
measurement on a table or other support so that the center of the cuff
will be at the same height as your heart.
Relax for about five or ten minutes before taking a measurement. If you
are excited or depressed by emotional stress, the measurement will
reflect this stress as a higher (or lower) than normal blood pressure
reading and the pulse reading will usually be faster than normal.
Your blood pressure varies constantly, depending on what you are
doing and what you have eaten. What you drink can have a very strong
and rapid effect on your blood pressure.
Press exhaust valve
START button
Releasing air
from the cuff
From previous page
Bulb
English 10
Measurement
This device bases its measurements on the heartbeat. If you have a
very weak or irregular heartbeat, the device may have difficulty
determining your blood pressure.
Should the device detect a condition that is abnormal, it will stop the
measurement and display an error symbol. Refer to page 4 for the
descriptions of the symbols.
This blood pressure device is intended for use by adults only. Consult
with your physician before using this device on a child. A child should
not use this device unattended.
What Is An Irregular Heartbeat?
Model UA-704 provides a blood pressure and pulse rate measurement
even when an irregular heartbeat occurs. An irregular heartbeat is
defined as a heartbeat that varies by 25% from the average of all
heartbeats during the blood pressure measurement. It is important that
you are relaxed, remain still and do not talk during measurements.
Note:
We recommend contacting your physician if you see this
indicator frequently.
About Blood Pressure
What Is Blood Pressure?
Blood pressure is the force exerted by blood against the walls of the
arteries. Systolic pressure occurs when the heart contracts. Diastolic
pressure occurs when the heart expands. Blood pressure is measured
in millimeters of mercury (mmHg). One's natural blood pressure is
represented by the fundamental pressure, which is measured first thing
in the morning while one is still at rest and before eating.
What Is Hypertension And How Is It Controlled?
Hypertension, an abnormally high arterial blood pressure, if left
unattended, can cause many health problems including stroke and
heart attack. Hypertension can be controlled by altering lifestyle,
avoiding stress, and with medication under a doctor’s supervision.
To prevent hypertension or to keep it under control:
Do not smoke Exercise regularly
Reduce salt and fat intake Have regular physical checkups
Maintain proper weight
English 11
About Blood Pressure
Why Measure Blood Pressure At Home?
Blood pressure measured at a clinic or doctor's office may cause
apprehension and can produce an elevated reading, 25 to 30 mmHg
higher than that measured at home. Home measurement reduces the
effects of outside influences on blood pressure readings, supplements
the doctor's readings and provides a more accurate, complete blood
pressure history.
WHO Blood Pressure Classification
Standards to assess high
blood pressure, without
regard to age, have been
established by the World
Health Organization
(WHO), as shown in the
chart at the right.
Blood Pressure Variations
An individual’s blood pressure varies greatly on a daily and seasonal
basis. It may vary by 30 to 50 mmHg due to various conditions during
the day. In hypertensive individual variations are even more
pronounced. Normally, the blood pressure rises while at work or play
and falls to its lowest levels during
sleep. So, do not be overly
concerned by the results of one
measurement.
Take measurements at the same
time every day using the
procedure described in this manual
to get to know your normal blood
pressure. Regular readings give a
more comprehensive blood
pressure history. Be sure to note
date and time when recording your
blood pressure. Consult your
doctor to interpret your blood
pressure data.
Typical fluctuation within a day
(
Measu
r
ed ever
five minutes
Blood Pressure
S
y
stolic
Diastolic
PM
A
MTime
mmH
g
Slee
p
English 12
Troubleshooting
Problem Possible Reason Recommended Action
Battery is drained. Replace with a new battery. Nothing
appears in the
display, even
when the
power is turned
on.
Battery polarities are
not in the correct
position.
Reinstall the battery with their
negative and positive end
matching those indicated in
the battery compartment.
The cuff is not fastened
properly.
Fasten the cuff correctly.
You moved your arm or
body during the
measurement.
Make sure you remain very
still and quiet during the
measurement.
The cuff position is not
correct.
Sit comfortably and still.
Ensure that the cuff is at the
same level as the heart.
The unit does
not measure.
Readings are
too high or too
low.
If you have a very weak or
irregular heart beat, the
device may have difficulty in
determining your blood
pressure.
The value is different
from that measured at a
clinic or doctor’s office.
See “Why measure blood
pressure at home?”.
Other
Remove the battery. Place it
back properly and try the
measurement again.
Note:
If the actions described above do not solve the problem, contact the
dealer.
Do not attempt to open or repair this product, as any attempt to do so
will make your warranty invalid.
Maintenance
Do not open the device. It uses delicate electrical components and an
intricate air unit that could be damaged. If you cannot fix the problem
using the troubleshooting instructions, request service from your dealer
or from the A&D service group. The A&D service group will provide
technical information, spare parts and units to authorized dealers.
The device was designed and manufactured for a long service life.
However it is generally recommended to have the device inspected
every 2 years, to ensure proper functioning and accuracy. Please
contact either your authorized dealer or A&D for maintenance.
English 13
Technical Data
Type UA-704
Measurement method Oscillometric measurement
Measurement range Pressure: 20 - 280 mmHg
Pulse: 40 - 200 beats / minute
Measurement accuracy Pressure: ±3 mmHg
Pulse : ±5%
Power supply 1 x 1.5V battery (R6P, LR6, or AA)
Number of measurements Approximately 4000 measurements, when
AA Alkaline batteries are used, with pressure
value of 180 mmHg at room temperature of
23 °C.
Classification Internally powered EM equipment
Continuous operation mode
Clinical test According to ANSI / AAMI SP-10 1987
EMC IEC 60601-1-2: 2007
Operating conditions +10 °C to +40 °C / 15%RH to 85 %RH
800 hPa to 1060 hPa
Transport / Storage conditions -10 °C to +60 °C / 15%RH to 85 %RH
Dimensions Approx. 100 [W] x 31 [H] x 51 [D] mm
Weight Approx. 73 g, excluding battery
Applied part Cuff Type BF
Useful life Device: 5 years (when used six times a day)
Cuff: 2 years (when used six times a day)
Accessories sold separately
Arm size: The circumference at the biceps.
Note: Specifications are subject to change without prior notice.
EMC table information is listed on our website:
http://www.aandd.jp/products/manual/medical/emc_en.pdf
Cuff Catalog Number Cuff Size Arm Size
CUF-D-LA Large adult cuff 32 cm to 45 cm
CUF-D-A Adult cuff 22 cm to 32 cm
Asahi 1-243, Kitamoto-shi, Saitama 364-8585, JAPAN
Telephone: [81] (48) 593-1111 Fax: [81] (48) 593-1119
A&D INSTRUMENTS LIMITED
Unit 24/26 Blacklands Way, Abingdon Business Park, Abingdon, Oxfordshire
OX14 1DY United Kingdom
Telephone: [44] (1235) 550420 Fax: [44] (1235) 550485
A&D ENGINEERING, INC.
1756 Automation Parkway, San Jose, California 95131, U.S.A.
Telephone: [1] (408) 263-5333 Fax: [1] (408)263-0119
A&D AUSTRALASIA PTY LTD
32 Dew Street, Thebarton, South Australia 5031, AUSTRALIA
Telephone: [61] (8) 8301-8100 Fax: [61] (8) 8352-7409
14

Hulp nodig? Stel uw vraag in het forum

Spelregels

Misbruik melden

Gebruikershandleiding.com neemt misbruik van zijn services uitermate serieus. U kunt hieronder aangeven waarom deze vraag ongepast is. Wij controleren de vraag en zonodig wordt deze verwijderd.

Product:

Bijvoorbeeld antisemitische inhoud, racistische inhoud, of materiaal dat gewelddadige fysieke handelingen tot gevolg kan hebben.

Bijvoorbeeld een creditcardnummer, een persoonlijk identificatienummer, of een geheim adres. E-mailadressen en volledige namen worden niet als privégegevens beschouwd.

Spelregels forum

Om tot zinvolle vragen te komen hanteren wij de volgende spelregels:

Belangrijk! Als er een antwoord wordt gegeven op uw vraag, dan is het voor de gever van het antwoord nuttig om te weten als u er wel (of niet) mee geholpen bent! Wij vragen u dus ook te reageren op een antwoord.

Belangrijk! Antwoorden worden ook per e-mail naar abonnees gestuurd. Laat uw emailadres achter op deze site, zodat u op de hoogte blijft. U krijgt dan ook andere vragen en antwoorden te zien.

Abonneren

Abonneer u voor het ontvangen van emails voor uw AnD UA-704 bij:


U ontvangt een email met instructies om u voor één of beide opties in te schrijven.


Ontvang uw handleiding per email

Vul uw emailadres in en ontvang de handleiding van AnD UA-704 in de taal/talen: Engels als bijlage per email.

De handleiding is 0,36 mb groot.

 

U ontvangt de handleiding per email binnen enkele minuten. Als u geen email heeft ontvangen, dan heeft u waarschijnlijk een verkeerd emailadres ingevuld of is uw mailbox te vol. Daarnaast kan het zijn dat uw internetprovider een maximum heeft aan de grootte per email. Omdat hier een handleiding wordt meegestuurd, kan het voorkomen dat de email groter is dan toegestaan bij uw provider.

Stel vragen via chat aan uw handleiding

Stel uw vraag over deze PDF

Uw handleiding is per email verstuurd. Controleer uw email

Als u niet binnen een kwartier uw email met handleiding ontvangen heeft, kan het zijn dat u een verkeerd emailadres heeft ingevuld of dat uw emailprovider een maximum grootte per email heeft ingesteld die kleiner is dan de grootte van de handleiding.

Er is een email naar u verstuurd om uw inschrijving definitief te maken.

Controleer uw email en volg de aanwijzingen op om uw inschrijving definitief te maken

U heeft geen emailadres opgegeven

Als u de handleiding per email wilt ontvangen, vul dan een geldig emailadres in.

Uw vraag is op deze pagina toegevoegd

Wilt u een email ontvangen bij een antwoord en/of nieuwe vragen? Vul dan hier uw emailadres in.



Info